Rosemount 00390-7101-0001 Low Flow Panel
Purchase options
$767.62
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Low flow pH panel for high-purity water: holds a 3900 sensor at a constant 2 GPH head so a general-purpose electrode can measure down to boiler-water conductivity.
Key features
- Low flow panel PN 00390-7101-0001 that holds a Rosemount 3900 general-purpose pH sensor and meters high-purity sample past it at a fixed rate
- General-purpose pH sensors are designed for water above about 50 uS/cm; high-pressure boiler water can run at 5 to 10 uS/cm, where the liquid junction is slow and the reading takes a long time to come down
- Constant head is geometry, not a valve: an inner drain holds the water column, and a 0.4 in height difference between the two chambers sets the pressure and the 2 GPH flow
- Regulated low flow steadies the reference junction and minimises KCl leaching, which is what prolongs sensor life
- Uses more than two-thirds less water than other side-stream solutions
- Keeps one pH sensor type across the plant: no specialised sensor for one application, no extra training, repair by sensor replacement
Key specifications
- Accuracy
- The panel does not add an accuracy specification of its own; it removes the flow sensitivity that degrades the sensor's reading in low-conductivity water. Rated flow 2 GPH through the sensor cell
- Output signal
- No electrical output. The panel is the hydraulic mount that carries a Rosemount 3900 pH sensor, whose signal goes to the transmitter or analyzer of the loop
- Diagnostics
- The mechanism is a weir and a controlled level, so there is no pressure regulator to adjust and no orifice to plug. Flow cannot drift without the water level changing visibly, which is what separates this from a flow that quietly goes wrong
- Pressure range
- Not applicable as a process pressure rating - the panel is an atmospheric sample conditioner. The pressure the sensor sees is set by a fixed 0.4 in (about 10 mm) height difference between the two chambers, which is what makes the flow constant
- Variants
- Panel PN 00390-7101-0001 accepts the Rosemount 3900 general-purpose pH sensor. Two transparent chambers with white push-fit fittings and tubing, an inner drain that fixes the water column, and an orange O-ring seal at the base
- Application
- High-purity water and low-conductivity sample points: boiler and feedwater chemistry in high-pressure steam systems, demineralised water, and any point below the general-purpose sensor's conductivity floor that still needs a pH trend rather than a grab sample
A panel built so a general-purpose sensor can work outside its stated envelope
The Rosemount 00390-7101-0001 low flow panel holds a 3900 general-purpose pH sensor and meters the sample past it at a fixed rate for high-purity water. Its purpose is not to add a feature to the sensor — it is to make a standard sensor usable in a water chemistry the sensor was never designed for.
The measurement the sensor is honestly not rated for
A general-purpose pH sensor is designed to work best where the water conducts: the technical literature puts that floor around 50 µS/cm. Boiler water in a high-pressure steam system can run at 5 to 10 µS/cm — a tenth of the floor or less. In that water there are almost no ions to carry the circuit between the glass electrode and the reference electrode, and the liquid junction potential becomes slow and unstable. The symptom is not a wrong number; it is a reading that takes a long time to come down to the true value and never quite settles.
The obvious answer is a specialised sensor built for pure water. The technical note is candid about why that answer is expensive: a specialised sensor works in one application only, so it becomes another part number in the stockroom, another training item, and another thing that has to be right at 2 a.m. Changing the hydraulics instead keeps one sensor type across the plant.
Constant head is geometry, not a valve
The panel's flow control is worth understanding because it has no moving parts to drift. Two chambers sit side by side, and an inner drain holds the water column in the first at a fixed level. Because the level is fixed, the head pressure above the outlet is fixed, and the flow through the sensor cell is fixed with it — 2 GPH as the constant operating value. The pressure the sensor actually sees is held by maintaining a height difference of 0.4 in (about 10 mm) between the two chambers.
That is the whole mechanism: a weir and a controlled level. There is no pressure regulator to adjust and no orifice to plug, which matters because a flow that drifts changes the measurement in a way nobody notices until the calibration is questioned. A flow set by the height of an overflow cannot drift without the water level changing visibly.
Why low flow is the measurement, not a water-saving trick
Two things improve when the sample moves slowly and steadily. The first is the junction: a stable, gentle flow keeps the reference junction in a steady state instead of being swept by a fast stream, which is what shortens the come-down time in low-conductivity water. The second is electrolyte life — flow regulation minimises leaching of the KCl fill solution, and that is what prolongs the sensor's working life. A faster side stream does not just waste water; it drains the reference electrolyte faster and moves the replacement date closer.
The water saving is real and follows from the same design: the panel uses more than two-thirds less water than other side-stream solutions. On a continuous sample line that runs every hour of the year, that difference is the operating cost of the measurement.
Standardisation is the quiet benefit
Because the panel carries the ordinary 3900 sensor, the plant keeps one pH sensor type, one spare, one calibration routine and one training scope. The sensor installs the way a normal pH sensor installs, and repair is replacement of the sensor rather than service of a special assembly. On a site with several functional groups, that is the difference between a stockroom that can answer a question and one that cannot.
Where it fits
High-purity water and low-conductivity sample points: boiler and feedwater chemistry in high-pressure steam systems, demineralised water, and any point where the conductivity is below the general-purpose sensor's design floor but the plant still needs a pH trend rather than a laboratory grab sample. Mount the panel where the overflow can be seen, because the visible water level is the only indication that the constant head — and therefore the flow rate — is what it should be.
Manufacturer reference: Rosemount technical note on using the 3900 pH sensor with the low flow panel to measure high purity water for the conductivity envelope, the constant-head mechanism, flow rate and test results.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | The panel does not add an accuracy specification of its own; it removes the flow sensitivity that degrades the sensor's reading in low-conductivity water. Rated flow 2 GPH through the sensor cell |
| Output signal | No electrical output. The panel is the hydraulic mount that carries a Rosemount 3900 pH sensor, whose signal goes to the transmitter or analyzer of the loop |
| Diagnostics | The mechanism is a weir and a controlled level, so there is no pressure regulator to adjust and no orifice to plug. Flow cannot drift without the water level changing visibly, which is what separates this from a flow that quietly goes wrong |
| Pressure range | Not applicable as a process pressure rating - the panel is an atmospheric sample conditioner. The pressure the sensor sees is set by a fixed 0.4 in (about 10 mm) height difference between the two chambers, which is what makes the flow constant |
| Variants | Panel PN 00390-7101-0001 accepts the Rosemount 3900 general-purpose pH sensor. Two transparent chambers with white push-fit fittings and tubing, an inner drain that fixes the water column, and an orange O-ring seal at the base |
| Application | High-purity water and low-conductivity sample points: boiler and feedwater chemistry in high-pressure steam systems, demineralised water, and any point below the general-purpose sensor's conductivity floor that still needs a pH trend rather than a grab sample |
| Mounting | Bench or wall position where the overflow is visible. The visible water level is the only indication that the constant head, and therefore the flow rate, is what the design intends |
| Operating temperature (°C) | Set by the sample temperature and the wetted materials of the panel; the assembly is not a heated or cooled device |