Non-intrusive pipe clamp RTD for surface temperature measurement: -50 to 300 C, pipe 1/2 in. to 60 in., spring-loaded element.

Key features

  • Non-intrusive pipe clamp RTD sensor: measures pipe surface temperature without welding or penetrating the line
  • Range -58 to 572 F (-50 to 300 C); fits pipe from 1/2 in. to 60 in. (22 to 1524 mm)
  • Spring-loaded element pressed against the wall, so contact pressure holds as the pipe expands and contracts
  • Single element 4-wire or dual element 3-wire RTD with silver tip
  • Nipple-union extension in stainless steel with 1/2 in. NPT instrument connection, 80 mm or 150 mm, or sensor-only without extension
  • Measures surface and ambient temperature, so a low-velocity gas line needs the wall-to-process difference considered

Key specifications

Accuracy
Range -58 to 572 F (-50 to 300 C); the element measures pipe wall temperature and infers the process, so a low velocity gas or partly filled line has to be assessed on wall-to-process difference, not on sensor accuracy alone
Output signal
RTD resistance to a transmitter or direct to the control system; single element 4-wire (P1) or dual element 3-wire (P2), both with a silver tip
Diagnostics
Contact pressure is the accuracy mechanism: a loose clamp leaves an air gap that adds thermal resistance, drifting the reading toward ambient and slowing the response, so the spring is what keeps the measurement repeatable
Pressure range
Not applicable to the sensor itself - it clamps to the outside of the pipe, so the line's pressure rating is unaffected and no process connection is required
Variants
Element P1 single 4-wire or P2 dual 3-wire; extension J nipple-union stainless steel with 1/2 in. NPT instrument connection at 3.1 in. (80 mm) or 5.9 in. (150 mm), or N for sensor only without extension
Application
Non-intrusive surface temperature measurement on existing pipe: retrofits, small bore lines where a thermowell would restrict flow, sanitary or non-penetrating installations, and temporary surveys along a line