Rosemount 060-761855 Calibration Kit
No product images are published for this product yet.
Purchase options
$350.19
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Premixed 4000 NTU formazin turbidity calibration kit (060-761855): the primary standard supplied ready to use, with lower points made by dilution.
Key features
- Premixed 4000 NTU formazin turbidity calibration kit, part number 060-761855
- Primary standard supplied ready to use, with no in-house polymerisation step
- Lower calibration points made by dilution with turbidity-free water
- Solution composition 0.5% hydrazine sulfate, 5% hexamethylenetetramine, 94.5% water
- Settles and ages like any formazin: mix before use, discard spent standard
- Replacement solution 905-761854 available separately when only the standard runs out
Key specifications
- Diagnostics
- Formazin is a reacting suspension that settles and ages, so the bottle is mixed before use and spent standard is discarded rather than returned; the SDS carries an EPCRA Section 313 note for hydrazine sulfate
- Variants
- Lower calibration points (2000, 1000, 200 and 20 NTU) are prepared by dilution with turbidity-free water; the 905-761854 replacement solution covers topping up an existing kit
- Application
- Calibration and verification of Rosemount turbidity instruments in drinking water, wastewater and industrial water treatment where calibration runs on a schedule
- Kit contents
- Premixed 4000 NTU formazin turbidity standard, part number 060-761855, supplied ready to use so no in-house polymerisation step is required
- Material of construction
- Aqueous solution of 0.5% hydrazine sulfate and 5% hexamethylenetetramine in 94.5% water, as documented in the SDS for the 905-761854 standard
Formazin without the mixing, for labs that calibrate on a schedule
The 060-761855 is the premixed 4000 NTU formazin calibration kit — the same primary standard as the 905-761854 replacement solution, supplied as a kit rather than as a bottle to top up an existing set. It exists for one reason: making formazin from hydrazine sulfate and hexamethylenetetramine in-house takes time, temperature control and a settling period, and a kit removes that work from the calibration schedule.
What premixed changes, and what it does not
Premixed means the polymerisation has already happened in a controlled batch, so the standard arrives at 4000 NTU and is used rather than made. What it does not change is the physics: the suspension still settles and still ages, so the bottle is mixed before use and an old kit reads low for the same reason an old bottle does. The kit is a convenience in preparation, not a way around handling.
The dilution arithmetic is still yours
A 4000 NTU standard covers the whole calibration curve only if the lower points are diluted from it with turbidity-free water: 2000, 1000, 200 and 20 NTU, each made at the time of calibration rather than stored. Diluting with water that carries 0.5 NTU puts a 2.5% error into a 20 NTU point before the instrument is connected — the most common way a good kit produces a bad calibration.
What is in the solution
The standard is water with 0.5% hydrazine sulfate and 5% hexamethylenetetramine, which is what the SDS for 905-761854 documents. It carries an EPCRA Section 313 note for hydrazine sulfate, so disposal follows the site's chemical procedure rather than the drain, and the safety data sheet is the document to keep with the kit.
Where it fits
Rosemount turbidity instruments in drinking water, wastewater and industrial water treatment where calibration is on a schedule and the standard is part of the regulatory evidence chain. Buy the kit when the set needs replacing as a whole; buy the 905-761854 replacement solution when only the standard has run out.
Handling
Store sealed and out of direct sunlight, and let the kit reach room temperature before opening so condensation does not form on the optics of the sample cell. Never return used standard to its bottle, and keep the SDS with the kit rather than in a folder somewhere else.
Manufacturer reference: Material Safety Data Sheet for 905-761854, 4000 NTU Formazin Turbidity Standard, Rosemount Analytical, revised 01/26/11, which documents the solution supplied in this kit.
Specifications
| Attribute | Value |
|---|---|
| Diagnostics | Formazin is a reacting suspension that settles and ages, so the bottle is mixed before use and spent standard is discarded rather than returned; the SDS carries an EPCRA Section 313 note for hydrazine sulfate |
| Variants | Lower calibration points (2000, 1000, 200 and 20 NTU) are prepared by dilution with turbidity-free water; the 905-761854 replacement solution covers topping up an existing kit |
| Application | Calibration and verification of Rosemount turbidity instruments in drinking water, wastewater and industrial water treatment where calibration runs on a schedule |
| Kit contents | Premixed 4000 NTU formazin turbidity standard, part number 060-761855, supplied ready to use so no in-house polymerisation step is required |
| Material of construction | Aqueous solution of 0.5% hydrazine sulfate and 5% hexamethylenetetramine in 94.5% water, as documented in the SDS for the 905-761854 standard |