Overview
Dual-input four-wire liquid analysis transmitter for pH/ORP, conductivity, chlorine, oxygen and turbidity, with two isolated 4-20 mA outputs and a large plain-language display.
Key features
- Two independent sensor inputs: pH/ORP, conductivity, chlorine, oxygen, ozone or turbidity
- Four-wire, line-powered operation with isolated outputs
- Two 4-20 mA outputs with HART digital signal, Profibus DP option
- Large backlit display with plain-language menus
- Relay contacts for alarms, dosing or sample valves
- NEMA 4X enclosure for panel, wall or pipe mounting
Key specifications
- Measurement principle
- Two independent sensor inputs linearised by one line-powered transmitter
- Accuracy
- pH: ±0.01 pH; conductivity: ±0.5% of reading; chlorine, oxygen and turbidity per sensor specification
- Process temperature
- -20 to 60 C ambient for the transmitter; process limits set by the sensor
- Process connection
- Sensor connection per input type; sensors wired to the transmitter terminals
- Enclosure rating
- NEMA 4X / IP65 for panel, wall or pipe mounting
- Output signal
- Two isolated 4-20 mA outputs with HART digital signal; Profibus DP option
What the 1056 is
A four-wire liquid analysis transmitter with two independent sensor inputs. Being line powered rather than loop powered, it can drive sensors that need more current than a two-wire loop allows, and it keeps the measurement running while the output loop is being serviced.
Two inputs, one display
The two channels are configured independently — pH on one and conductivity on the other, or chlorine plus temperature, or two points of the same parameter in a long process. Each channel has its own output, so the control system sees two clean 4-20 mA signals rather than a multiplexed value.
Sensor compatibility
pH and ORP electrodes, contacting and toroidal conductivity sensors, free and total chlorine, dissolved oxygen, ozone and turbidity. Sensor calibration data is held in the sensor or entered at the transmitter, and the display walks through a buffer calibration without a separate handheld.
Installation
The NEMA 4X enclosure mounts on a panel, a wall or a pipe, with the wiring terminals in a separate compartment from the electronics. Analogue outputs are isolated, and relay contacts are available for alarms, chemical dosing or sample-valve control.
Where it fits
Municipal water and wastewater plants, cooling water, boiler feedwater and food and beverage CIP systems — the applications where two analytical measurements sit at one sample point and the readings drive both a control loop and a recorder.
Manufacturer reference: Emerson document 00813-0100-3056.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | Two independent sensor inputs linearised by one line-powered transmitter |
| Accuracy | pH: ±0.01 pH; conductivity: ±0.5% of reading; chlorine, oxygen and turbidity per sensor specification |
| Process temperature | -20 to 60 C ambient for the transmitter; process limits set by the sensor |
| Process connection | Sensor connection per input type; sensors wired to the transmitter terminals |
| Enclosure rating | NEMA 4X / IP65 for panel, wall or pipe mounting |
| Output signal | Two isolated 4-20 mA outputs with HART digital signal; Profibus DP option |
| Power supply | 85-265 V AC or 24 V DC, four-wire (line powered) |
| Wetted materials | Set by the sensor selected; polycarbonate enclosure with stainless hardware |
| Diagnostics | Sensor and instrument diagnostics on screen, over HART and through the fault relay |
| Certifications | CE and cULus; hazardous-area suitability depends on the sensor loop |