Rosemount Model 1057 Multi-Parameter Analyzer
Purchase options
From$829.62
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Three-channel liquid analysis transmitter that accepts pH/ORP, conductivity, chlorine, oxygen or turbidity sensors and reports them over dual 4-20 mA outputs with HART.
Key features
- Three input channels: pH/ORP, conductivity, chlorine, oxygen, ozone or turbidity
- Dual 4-20 mA outputs with HART digital signal
- Profibus DP option for plant networks
- Backlit display showing measurement, temperature and diagnostics
- Menu-driven setup from the front panel
- Panel, wall or pipe mounting in a NEMA 4X enclosure
Key specifications
- Measurement principle
- Three independent sensor inputs converted and linearised by one transmitter
- Accuracy
- pH: ±0.01 pH; conductivity: ±0.5% of reading; chlorine and oxygen per sensor specification
- Process temperature
- -20 to 60 C ambient for the transmitter; process limits set by the sensor
- Process connection
- Sensor connection per input type; sensors wired to the transmitter terminals
- Enclosure rating
- NEMA 4X / IP65 enclosure for panel, wall or pipe mounting
- Output signal
- Two 4-20 mA outputs with HART digital signal; Profibus DP option
What the 1057 is
A three-channel liquid analysis transmitter. Each channel takes a different sensor — pH or ORP, conductivity, chlorine, dissolved oxygen, ozone or turbidity — so one instrument can watch a complete water process instead of one parameter per box.
Why three inputs matter
Water treatment is rarely about a single number. Free chlorine depends on pH, conductivity shifts with the chemicals dosed upstream, and a turbidity reading from the same sample point tells you whether a filter is passing solids. Putting those on one transmitter means one display, one wiring run to the control room and one set of alarm relays.
Operation
Setup is menu-driven from the front panel in plain language, with a backlit display that shows the measurement, the temperature and the diagnostic state at the same time. The enclosure mounts on a panel, a wall or a pipe, and the electronics are separated from the wiring terminals so a sensor change does not disturb the loop wiring.
Outputs and integration
Two 4-20 mA outputs carry the primary measurements with a HART digital signal superimposed, and a Profibus DP option is available where the plant runs a fieldbus network rather than analogue loops. Alarm and wash contacts are provided for relay-driven sample systems.
Where it fits
Municipal drinking water and wastewater plants, cooling towers, boiler feedwater, and industrial effluent monitoring — anywhere two or three analytical parameters are measured at one point and the readings have to reach a control system.
Manufacturer reference: Emerson document 00813-0100-3157.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | Three independent sensor inputs converted and linearised by one transmitter |
| Accuracy | pH: ±0.01 pH; conductivity: ±0.5% of reading; chlorine and oxygen per sensor specification |
| Process temperature | -20 to 60 C ambient for the transmitter; process limits set by the sensor |
| Process connection | Sensor connection per input type; sensors wired to the transmitter terminals |
| Enclosure rating | NEMA 4X / IP65 enclosure for panel, wall or pipe mounting |
| Output signal | Two 4-20 mA outputs with HART digital signal; Profibus DP option |
| Power supply | 85-265 V AC or 24 V DC depending on the model |
| Wetted materials | Set by the sensor selected; transmitter enclosure is polycarbonate with stainless hardware |
| Diagnostics | Sensor and instrument diagnostics on screen and over HART, with fault relay |
| Certifications | CE and cULus; hazardous-area approval depends on the sensor loop and installation |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer