Rosemount 1058 Dual Channel Transmitter
Purchase options
From$1,006.40
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Dual channel liquid analysis transmitter for pH, conductivity, oxygen, chlorine, ozone and turbidity, digital or analog per channel.
Key features
- Dual channel liquid analysis transmitter: two independent channels in one instrument, each configurable for a digital Modbus RS-485 sensor or an analog sensor
- Supported measurements across the two channels: pH, ORP, ISE, contacting and toroidal conductivity, dissolved oxygen, chlorine, ozone, turbidity and flow or current input
- Independent channels allow a brownfield move: put a digital sensor on one channel and keep the existing analog probe on the other until its replacement is due
- Contacting conductivity for clean low conductivity water and toroidal for high conductivity, fouling or aggressive media - the technology is chosen in the model code and needs its matching signal board
- Ready to connect digital sensors configure in under ten seconds because the sensor carries its own identity and calibration
- Predictive pH sensor health diagnostics report when a pH sensor is nearing the end of its life, turning a reactive callout into a planned probe change
Key specifications
- Accuracy
- Determined by the sensors and their calibration. Digital sensors carry their own calibration and configure in under ten seconds; analog channels must be told the parameter and calibrated to the probe
- Output signal
- Four current outputs, two per channel, as shown on the transmitter's own display. Digital channels use Modbus RS-485; analog channels need the signal board matching the parameter, such as the separate contacting and toroidal conductivity boards
- Diagnostics
- Contacting conductivity suits clean low conductivity water with the best resolution at the low end; toroidal suits high conductivity, fouling, coating and aggressive media with no electrodes exposed. Choosing the wrong one is a fouling problem or a resolution problem, never a defect
- Pressure range
- Set by the sensor and its mounting, not by the transmitter. The enclosure is an IP66 / Type 4X engineered polymer housing with universal mounting options
- Variants
- Channel signal codes D for digital Modbus RS-485, and analog codes P pH/ORP/ISE, C contacting conductivity, T toroidal conductivity, X dissolved oxygen, L chlorine, Z ozone, N turbidity and F flow or current input, chosen independently per channel
- Application
- Water and wastewater treatment, potable water, cooling and boiler chemistry, and process liquid analysis where two parameters are measured at one point or where a single instrument is preferred over two
Two channels, and each one decides its own technology
The Rosemount 1058 is a dual channel liquid analysis transmitter: two independent measurement channels in one instrument, each independently configurable for a digital or an analog sensor. Between them they cover pH, ORP, ISE, conductivity, dissolved oxygen, chlorine, ozone, turbidity and a flow or current input. The enclosure is IP66 / Type 4X engineered polymer with universal mounting, the interface is a large colour LCD with an alphanumeric keypad, and digital sensors use Ready to connect technology — connect and configure in under ten seconds.
Why "independently configurable" is the useful part
Each channel is ordered with its own signal code: D for a digital Modbus RS-485 sensor, and then an analog code for the parameter the sensor measures — P for pH, ORP or ISE, C for contacting conductivity, T for toroidal conductivity, X for dissolved oxygen, L for chlorine, Z for ozone, N for turbidity, F for flow or current.
Because the two channels are configured separately, the instrument does not force a plant to move everything at once. A site with an old analog pH probe can put a digital sensor on channel one and keep the existing probe on channel two, then replace the second when its budget or its maintenance cycle allows. The same pattern suits a loop that genuinely needs two different technologies — pH and conductivity are the common pair, and the instrument's own display shows exactly that: a pH value on one channel, a conductivity value on the other, with a temperature for each and four current outputs.
Contacting or toroidal: the water decides
Both conductivity technologies are offered, and they are not interchangeable preferences. They behave differently in the same service:
| Technology | Where it belongs |
|---|---|
| Contacting | Clean, low conductivity water. Electrodes in direct contact with the liquid, with the best resolution at the low end and the most accurate reading in pure and demineralised water |
| Toroidal | High conductivity, fouling, coating, or aggressive media. An inductive loop with no electrodes exposed to the liquid, so there is nothing to foul and no polarisation limit at the top of the range |
A contacting sensor is the right choice for a demineraliser outlet at a few microsiemens. Put the same sensor in a brine line, a slurry or a wastewater stream and it fouls, polarises and drifts. A toroidal sensor in that service is untroubled — and the same toroidal sensor on a pure water line has resolution that is too coarse to be useful. The value worth watching on the display is the one that tells you which case you have: a reading in the tens of millisiemens is toroidal territory, and a reading in fractions of a microsiemen is contacting territory.
Note the ordering consequence. The analog channel needs the signal board that matches its parameter — there are separate kits for contacting and toroidal conductivity, for example — so the technology is chosen in the model code, not switchable in the field.
A digital sensor arrives knowing what it is
Ready to connect is the practical difference between the two channel types. A digital sensor carries its own identity and calibration, so plugging it in gives the transmitter the parameter, the range and the calibration without anyone typing a model code; that is where the sub-ten-second claim comes from. An analog channel, by contrast, has to be told what is on the end of the cable and calibrated to it.
That is also why the two channel types age differently. A digital sensor can be moved between transmitters without losing anything; an analog channel's calibration belongs to the pair, and replacing the probe means repeating the calibration. On a site with several analyzers and one maintenance team, that difference is a scheduling consideration rather than a technical one.
The diagnostic that matters most
The transmitter includes predictive pH sensor health diagnostics, which report when a pH sensor is nearing the end of its life rather than when it has already failed. For the measurement that is most often the reason an analyzer exists, that converts a reactive callout — the reading is wrong, find out why — into a planned probe replacement during a normal visit.
Where it fits
Water and wastewater treatment, potable water, cooling and boiler chemistry, and process liquid analysis where two parameters are measured at one point or where a single instrument is preferred over two. Choose the channel codes from the sensors actually available at the site rather than from what would be ideal, pair the conductivity technology to the water, and use the digital channels where the maintenance team wants probe changes to be a swap rather than a calibration exercise.
Manufacturer reference: Rosemount 1058 Dual Channel Transmitter Product Data Sheet MS-00813-0100-1058 for channel signal codes, supported sensors, enclosure ratings and diagnostics.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Determined by the sensors and their calibration. Digital sensors carry their own calibration and configure in under ten seconds; analog channels must be told the parameter and calibrated to the probe |
| Output signal | Four current outputs, two per channel, as shown on the transmitter's own display. Digital channels use Modbus RS-485; analog channels need the signal board matching the parameter, such as the separate contacting and toroidal conductivity boards |
| Diagnostics | Contacting conductivity suits clean low conductivity water with the best resolution at the low end; toroidal suits high conductivity, fouling, coating and aggressive media with no electrodes exposed. Choosing the wrong one is a fouling problem or a resolution problem, never a defect |
| Pressure range | Set by the sensor and its mounting, not by the transmitter. The enclosure is an IP66 / Type 4X engineered polymer housing with universal mounting options |
| Variants | Channel signal codes D for digital Modbus RS-485, and analog codes P pH/ORP/ISE, C contacting conductivity, T toroidal conductivity, X dissolved oxygen, L chlorine, Z ozone, N turbidity and F flow or current input, chosen independently per channel |
| Application | Water and wastewater treatment, potable water, cooling and boiler chemistry, and process liquid analysis where two parameters are measured at one point or where a single instrument is preferred over two |
| Mounting | Panel or surface mounting with universal mounting options in an IP66 / Type 4X polymer enclosure. Large colour LCD with alphanumeric keypad for configuration and calibration |
| Operating temperature (°C) | Set by the sensors and the installation. The transmitter provides predictive pH sensor health diagnostics that report when a pH sensor is nearing the end of its life |