Rosemount 105P Liquid Analysis Standard Panel
Purchase options
From$737.77
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Pre-engineered liquid analysis panel with one transmitter and up to two sensors: pH, ORP, DO, chlorine, fluoride or ammonia on one repeatable layout.
Key features
- Pre-engineered liquid analysis panel carrying one transmitter (1058 or 1056) and up to two sensors in a repeatable layout, for pH, ORP, dissolved oxygen, chlorine, fluoride and ammonia
- The specification is a set of limits the sample must already meet: 3 to 65 psig, 0 to 50 C, 3 to 80 gal/hr flow, greater than 50 microsiemens per centimetre conductivity, and a drain that runs to open atmosphere
- The upper flow limit has a stated reason: above 80 gal/hr the overflow tube backs up, because the panel's flow control is a gravity overflow arrangement - the result is water on the panel, not a bad reading
- The drain must go to open atmosphere: the flow cells are open-channel by design, so a closed pressurised drain or a submerged sump connection is a civil works problem discovered on commissioning day
- The conductivity figure belongs to the panel, not the sensor: a 396A pH sensor needs 100 microsiemens per centimetre, so a sample at 70 is inside the panel's window and outside the sensor's, and the tighter limit governs
- Two flow cells in series measure the same water at the same moment, which is what makes a pH and chlorine pair or a pH and ORP pair a control decision rather than two unrelated numbers
Key specifications
- Accuracy
- Not a measuring device - the accuracy belongs to the sensor and transmitter selected. What the panel fixes is the hydraulic envelope: 3 to 80 gal/hr flow (11 to 303 L/hr), 0 to 50 C, and greater than 50 microsiemens per centimetre conductivity at 25 C
- Output signal
- Set by the transmitter: Rosemount 1058 or 1056, one per panel, with sensor cables included and two flow cells arranged in series on one sample stream
- Diagnostics
- Two limits are plumbing limits rather than sensor limits and are the ones that stop a project after delivery. Above 80 gal/hr the overflow tube backs up, because flow control is a gravity overflow arrangement - the result is water on the panel, not a bad reading. And the drain must go to open atmosphere: the flow cells are open-channel by design, so a pressurised or submerged drain connection is a civil works problem discovered on commissioning day. Note also that the panel's 50 microsiemens per centimetre floor is not the sensor's: a 396A pH sensor needs 100, and the tighter limit governs
- Pressure range
- Sample pressure 3 to 65 psig (122 to 549 kPa abs). The panel is not a pressure vessel: a sample above 65 psig needs a regulator upstream, which is the conditioning the pre-engineered panel deliberately does not include
- Application
- Small to mid-sized panel projects for water quality monitoring: drinking water and municipal treatment, power generation cooling water and condensate, chemical and refining process water, pulp and paper, and mining process water. The value is standardisation - the same layout on every panel, so the technician who commissioned the first can service the tenth and spares are common
- Mounting
- Wall-mounted backplate assembly weighing 6 lb (2.7 kg), with a 1/4 in OD tubing compression process connection that can be replaced with a barbed fitting for soft tubing, and a 3/4 in barbed drain
The panel's specification is a list of things your sample must already be
The Rosemount 105P is a pre-engineered liquid analysis panel: a backplate carrying a transmitter, flow cells, sample tubing and flow control, factory assembled so that a pH, ORP, dissolved oxygen, chlorine, fluoride or ammonia measurement can be specified as one part number instead of a plumbing project. It holds one transmitter (1058 or 1056) and up to two sensors — commonly a pH and a chlorine, or a pH and an ORP, measuring the same sample stream in series.
What makes it useful is not a clever component. It is that a panel is a repeated design: the engineering that would otherwise be done once per site — flow cell planning, sample handling, wiring and plumbing — is already done, and the layout is the same every time, which is what makes commissioning and maintenance predictable.
What makes it a specification rather than an accessory is the list of limits it imposes on the sample, because those limits are plumbing limits, not sensor limits, and they are the ones that stop a project after the panel has arrived.
Five numbers the sample has to satisfy
| Limit | Value |
|---|---|
| Sample pressure | 3 to 65 psig (122 to 549 kPa abs) |
| Sample temperature | 32 to 122 °F (0 to 50 °C) |
| Flow | 3 to 80 gal/hr (11 to 303 L/hr) |
| Sample conductivity | greater than 50 µS/cm at 25 °C |
| Drain | must drain to open atmosphere |
Two of these deserve a sentence each because they are the ones that get argued about after installation.
The upper flow limit has a stated reason. At more than 80 gal/hr the datasheet says the overflow tube backs up — the panel's flow control is a gravity overflow arrangement, and above its capacity the sample cannot leave as fast as it arrives. The consequence is not a bad reading, it is water on the panel and in the cabinet. A sample line with a generous supply pressure and no restriction will do exactly this, which is why flow control belongs in the commissioning, not in the site's discretion.
The drain must go to open atmosphere. A panel that drains into a pressurised line, a submerged sump connection or a common header under vacuum will not behave: the flow cells are open-channel by design and the drain is a ¾ in barbed fitting intended to run freely into a funnel or a trench. Sites with a closed drain system routinely discover this on the day of commissioning, and it is a civil works problem rather than an instrument one.
The conductivity number belongs to the panel, not to the sensor
Fifty microsiemens per centimetre is the panel's minimum, and the sensor on it may have a higher one. A Rosemount 396A anti-coating pH sensor, for example, specifies a minimum of 100 µS/cm because its reference needs ions for a stable liquid junction.
The tighter of the two governs. So a sample at 70 µS/cm is inside the panel's window and outside that sensor's — the panel is satisfied, the measurement is not, and nothing about the installation looks wrong. When specifying, check the sensor's limit against the water first and treat the panel's figure as the floor it is.
What the two sensors are for
Two flow cells in series on one sample stream is the layout's real value: the water passes one sensor and then the other, so the two readings describe the same water at the same moment. That matters when the two measurements are being compared — pH and ORP together for disinfection, pH and chlorine for dosing control, pH and fluoride for treatment compliance.
Two sensors in two different sample lines would give two numbers taken at different times from different water, and the control decision that follows is subtly wrong in a way that is very hard to diagnose. That is the argument for measuring in series rather than in parallel, and it is why the panel exists in this shape.
What is on the plate
The wetted parts are acrylic, polycarbonate, polyester, Kynar, nylon and silicone, and the panel weighs 6 lb (2.7 kg) — it is a wall-mounted assembly, not a skid. The process connection is ¼ in OD tubing compression and can be replaced with a barbed fitting for soft tubing; the drain is a ¾ in barbed fitting.
Those materials set the chemical envelope as much as the sensor does. Kynar and acrylic are fine for the water and wastewater duties the panel is intended for — drinking water, power station cooling and condensate, chemical and refining process water, pulp and paper, mining process water — and are not a universal answer for aggressive chemical streams.
Where it fits
Small to mid-sized projects where a panel is needed and nobody wants to design one: a municipal waterworks adding chlorine and pH monitoring, a power station watching condensate, a plant monitoring its own process water. The value is the standardisation — the same layout on every panel means the technician who commissioned the first one can service the tenth, and spares are common.
It is the wrong answer where the sample needs conditioning the standard panel does not do — cooling and pressure reduction for a hot high-pressure tap, or a sample below the conductivity floor — and in those cases the engineering that the pre-engineered panel was meant to save has to be done anyway, but before the order rather than after.
Manufacturer reference: Rosemount 105P Panel Systems data sheet MS-00807-0100-9834 for sample limits, compatible sensors and transmitters, materials, connections and panel dimensions, with the 105P quick start guide for installation.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Not a measuring device - the accuracy belongs to the sensor and transmitter selected. What the panel fixes is the hydraulic envelope: 3 to 80 gal/hr flow (11 to 303 L/hr), 0 to 50 C, and greater than 50 microsiemens per centimetre conductivity at 25 C |
| Output signal | Set by the transmitter: Rosemount 1058 or 1056, one per panel, with sensor cables included and two flow cells arranged in series on one sample stream |
| Diagnostics | Two limits are plumbing limits rather than sensor limits and are the ones that stop a project after delivery. Above 80 gal/hr the overflow tube backs up, because flow control is a gravity overflow arrangement - the result is water on the panel, not a bad reading. And the drain must go to open atmosphere: the flow cells are open-channel by design, so a pressurised or submerged drain connection is a civil works problem discovered on commissioning day. Note also that the panel's 50 microsiemens per centimetre floor is not the sensor's: a 396A pH sensor needs 100, and the tighter limit governs |
| Pressure range | Sample pressure 3 to 65 psig (122 to 549 kPa abs). The panel is not a pressure vessel: a sample above 65 psig needs a regulator upstream, which is the conditioning the pre-engineered panel deliberately does not include |
| Application | Small to mid-sized panel projects for water quality monitoring: drinking water and municipal treatment, power generation cooling water and condensate, chemical and refining process water, pulp and paper, and mining process water. The value is standardisation - the same layout on every panel, so the technician who commissioned the first can service the tenth and spares are common |
| Mounting | Wall-mounted backplate assembly weighing 6 lb (2.7 kg), with a 1/4 in OD tubing compression process connection that can be replaced with a barbed fitting for soft tubing, and a 3/4 in barbed drain |
| Operating temperature (°C) | Sample temperature 32 to 122 F (0 to 50 C). A hotter sample needs cooling before the panel, which again is the conditioning the panel does not do |
| Measuring range | Up to two sensors per panel, compatible with the 3900VP pH/ORP, 490A dissolved oxygen, 499ACL chlorine, 499ADO/TrDO dissolved oxygen and 396 series pH/ORP sensors. Two measurements on one sample stream means the two readings describe the same water at the same moment, which is what makes a pH and chlorine pair or a pH and ORP pair a control decision rather than two unrelated numbers |
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.