Rosemount 105P Liquid Analysis Standard Panel

Part number105PManufacturerEmerson

Purchase options

From$737.77

Availability
Call for availability Originalx
Select option

Configurable — choose options to build the exact part number.

Independent supplier

Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.

Overview

Pre-engineered liquid analysis panel with one transmitter and up to two sensors: pH, ORP, DO, chlorine, fluoride or ammonia on one repeatable layout.

Key features

  • Pre-engineered liquid analysis panel carrying one transmitter (1058 or 1056) and up to two sensors in a repeatable layout, for pH, ORP, dissolved oxygen, chlorine, fluoride and ammonia
  • The specification is a set of limits the sample must already meet: 3 to 65 psig, 0 to 50 C, 3 to 80 gal/hr flow, greater than 50 microsiemens per centimetre conductivity, and a drain that runs to open atmosphere
  • The upper flow limit has a stated reason: above 80 gal/hr the overflow tube backs up, because the panel's flow control is a gravity overflow arrangement - the result is water on the panel, not a bad reading
  • The drain must go to open atmosphere: the flow cells are open-channel by design, so a closed pressurised drain or a submerged sump connection is a civil works problem discovered on commissioning day
  • The conductivity figure belongs to the panel, not the sensor: a 396A pH sensor needs 100 microsiemens per centimetre, so a sample at 70 is inside the panel's window and outside the sensor's, and the tighter limit governs
  • Two flow cells in series measure the same water at the same moment, which is what makes a pH and chlorine pair or a pH and ORP pair a control decision rather than two unrelated numbers

Key specifications

Accuracy
Not a measuring device - the accuracy belongs to the sensor and transmitter selected. What the panel fixes is the hydraulic envelope: 3 to 80 gal/hr flow (11 to 303 L/hr), 0 to 50 C, and greater than 50 microsiemens per centimetre conductivity at 25 C
Output signal
Set by the transmitter: Rosemount 1058 or 1056, one per panel, with sensor cables included and two flow cells arranged in series on one sample stream
Diagnostics
Two limits are plumbing limits rather than sensor limits and are the ones that stop a project after delivery. Above 80 gal/hr the overflow tube backs up, because flow control is a gravity overflow arrangement - the result is water on the panel, not a bad reading. And the drain must go to open atmosphere: the flow cells are open-channel by design, so a pressurised or submerged drain connection is a civil works problem discovered on commissioning day. Note also that the panel's 50 microsiemens per centimetre floor is not the sensor's: a 396A pH sensor needs 100, and the tighter limit governs
Pressure range
Sample pressure 3 to 65 psig (122 to 549 kPa abs). The panel is not a pressure vessel: a sample above 65 psig needs a regulator upstream, which is the conditioning the pre-engineered panel deliberately does not include
Application
Small to mid-sized panel projects for water quality monitoring: drinking water and municipal treatment, power generation cooling water and condensate, chemical and refining process water, pulp and paper, and mining process water. The value is standardisation - the same layout on every panel, so the technician who commissioned the first can service the tenth and spares are common
Mounting
Wall-mounted backplate assembly weighing 6 lb (2.7 kg), with a 1/4 in OD tubing compression process connection that can be replaced with a barbed fitting for soft tubing, and a 3/4 in barbed drain
See all specifications