Rosemount 1199 SCW Hygienic Seal
Purchase options
From$500.98
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
In-line sanitary Tri-Clamp diaphragm seal to 3-A Standard 74, 1-1/2 to 4 in, with clamp-dependent pressure rating to 1,500 psi and finish to 10 micro-inch Ra.
Key features
- In-line sanitary diaphragm seal conforming to 3-A Standard 74, clamped between two Tri-Clamp ferrules in a pipeline
- Sizes 1-1/2, 2, 2-1/2, 3 and 4 in; the SSW tank spud seal is the flush-on-a-vessel alternative, not an interchangeable part
- The pressure rating is a CLAMP property: option 2 supplies a high-pressure Ladish clamp rated 1,500 psi at 1-1/2 in and 1,000 psi at the larger sizes, derated 20 percent at 250 F
- The 1-1/2 in size carries the HIGHEST rating because a smaller clamp diameter seats harder - bigger is not stronger here
- Wetted finish from 32 micro-inch standard down to 20, 15 and 10 micro-inch Ra with electropolishing available
- Option P is a non-sanitary fill fluid and breaks 3-A conformance even when every metal part is correct; gasket and clamp are normally customer-supplied
Key specifications
- Accuracy
- The seal transmits pressure through a fill fluid; its contribution is the diaphragm area and the fill-fluid temperature effect rather than an accuracy figure of its own
- Output signal
- Not applicable - a passive pressure boundary that mounts to a Rosemount transmitter directly or by capillary
- Diagnostics
- 3-A Standard 74 conformance covers the whole assembly: option P, a non-sanitary fill fluid, breaks it even with perfect metalwork because the fluid sits behind the diaphragm. The 1-1/2 in size has the highest clamp rating because a smaller clamp diameter seats harder - bigger is not stronger on a Tri-Clamp joint
- Pressure range
- Set by the clamp, not by the seal. A high-pressure Ladish clamp (option 2) rates 1,500 psi at 70 F in 1-1/2 in and 1,000 psi in the larger sizes, derated to 1,200 and 800 psi at 250 F - 20 percent lost to temperature
- Variants
- Sizes 30 (1-1/2 in), 50 (2 in), 60 (2-1/2 in), 70 (3 in) and 90 (4 in). Diaphragm and housing in 316L SST (WNr 1.4435), Hastelloy C-276 with a 316L housing, or C-276 for both. Finish options H 20 micro-inch, G 15 micro-inch and D 10 micro-inch Ra, plus electropolishing. Option 2 high-pressure Ladish clamp with Buna N gasket; option 3 Buna N gasket alone
- Application
- Line pressure and differential measurement on hygienic process piping: product lines, CIP supply and return, filter and separator differential, and filling-line pressure. On a vessel wall the SSW tank spud version is the correct part instead
The clamp is the pressure rating, and the smallest size is the strongest
The Rosemount 1199 SCW is the in-line sanitary diaphragm seal: a Tri-Clamp body that clamps between two ferrules in a pipeline, with the diaphragm facing the flow. It conforms to 3-A Standard 74, comes in 1½, 2, 2½, 3 and 4 inch sizes, and its wetted finish runs from a standard 32 µin Ra down to 10 µin (0.25 µm) — one step finer than the tank spud version.
In-line, where the tank spud is flush
These two sanitary seals solve opposite geometry problems and it is worth being clear which is which. The SSW tank spud mounts on a vessel wall with the diaphragm flush to the inside surface, which is what a level measurement needs. The SCW mounts in a pipe, clamped between two Tri-Clamp ferrules, which is what a line pressure or a filter differential needs. Both are sanitary for the same reason — a flat, cleanable face with no threads or recesses — but they are not interchangeable, and a specification that names "a 1199 sanitary seal" without the style is not yet a specification.
Where the pressure rating actually comes from
On a Tri-Clamp joint the clamp holds the two ferrules together and the gasket seals between them, so the assembly's pressure limit is a clamp property, not a seal property. The ordering table reflects that with option 2, a high-pressure Ladish clamp with a Buna N gasket, and the ratings it delivers are worth reading carefully:
| Size | Rating at 70 °F | Rating at 250 °F |
|---|---|---|
| 1½ in | 1,500 psi | 1,200 psi |
| 2 in | 1,000 psi | 800 psi |
| 2½ in | 1,000 psi | 800 psi |
| 3 in | 1,000 psi | 800 psi |
| 4 in | 1,000 psi | 800 psi |
The 1½ inch size carries the highest rating — 1,500 psi against 1,000 for everything larger — which runs against the usual assumption that bigger connections hold more pressure. It follows from the geometry: a smaller clamp diameter puts the same clamp force around a shorter circumference, so the joint seats harder. And every size derates at temperature: 250 °F costs 20 percent of the rating. Both facts belong in the specification, because the clamp a customer already has in the spares bin may not be the one that meets the line's design pressure.
Gasket and clamp are usually yours
The typical installation figure shows the gasket, clamp and mating connection as customer-supplied — the seal provides the diaphragm face, not the joint hardware. Option 2 is the exception: it ships the high-pressure Ladish clamp with a Buna N gasket, and option 3 supplies a Buna N gasket on its own. For a hygienic line the gasket is not a consumable detail: it is the one elastomer in the wetted path, and it has to be sanitary-rated as well as the right size.
The finish ladder goes further here
Options are H at 20 µin (0.5 µm), G at 15 µin (0.375 µm) and D at 10 µin (0.25 µm) against a standard 32 µin, plus option 6, electropolishing. The 10 µin option exists for the surfaces where biofilm retention is the governing risk rather than a secondary concern, and it is worth remembering what the number is for: a caustic wash removes soil from a smooth surface but does not reliably remove a film that has established itself in a scratch pattern, so roughness is a cleanability specification.
The same trap as the spud version
Option P is a non-sanitary fill fluid and does not conform to 3-A Standard 74. The reasoning is identical to the tank spud seal: the metal can be perfect, and the fluid behind the diaphragm still sits inside the hygienic boundary, a thin barrier away from the product. On the in-line version the temptation is slightly higher, because a line pressure seal is often ordered for a utility or CIP circuit where nobody thinks of it as product contact — and then the same seal part number gets used on a product line.
Materials and where it fits
Diaphragm options are 316L SST (WNr 1.4435) with a 316L upper housing, Hastelloy C-276 with a 316L housing, or C-276 for both. The SCW is for line pressure and differential measurement on hygienic process piping: product lines, CIP supply and return, filter and separator differential, and filling-line pressure. Where the measurement is on a vessel wall, that is the SSW; where it is in the pipe, this is the one.
Manufacturer reference: Rosemount 1199 Sanitary Seals Product Data Sheet 00813-0100-4016, Tri-Clamp (SCW) section, ordering information and high-pressure clamp ratings.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | The seal transmits pressure through a fill fluid; its contribution is the diaphragm area and the fill-fluid temperature effect rather than an accuracy figure of its own |
| Output signal | Not applicable - a passive pressure boundary that mounts to a Rosemount transmitter directly or by capillary |
| Diagnostics | 3-A Standard 74 conformance covers the whole assembly: option P, a non-sanitary fill fluid, breaks it even with perfect metalwork because the fluid sits behind the diaphragm. The 1-1/2 in size has the highest clamp rating because a smaller clamp diameter seats harder - bigger is not stronger on a Tri-Clamp joint |
| Pressure range | Set by the clamp, not by the seal. A high-pressure Ladish clamp (option 2) rates 1,500 psi at 70 F in 1-1/2 in and 1,000 psi in the larger sizes, derated to 1,200 and 800 psi at 250 F - 20 percent lost to temperature |
| Variants | Sizes 30 (1-1/2 in), 50 (2 in), 60 (2-1/2 in), 70 (3 in) and 90 (4 in). Diaphragm and housing in 316L SST (WNr 1.4435), Hastelloy C-276 with a 316L housing, or C-276 for both. Finish options H 20 micro-inch, G 15 micro-inch and D 10 micro-inch Ra, plus electropolishing. Option 2 high-pressure Ladish clamp with Buna N gasket; option 3 Buna N gasket alone |
| Application | Line pressure and differential measurement on hygienic process piping: product lines, CIP supply and return, filter and separator differential, and filling-line pressure. On a vessel wall the SSW tank spud version is the correct part instead |
| Mounting | Clamped between two Tri-Clamp ferrules in the pipeline. The gasket, clamp and mating connection are customer-supplied in the standard configuration, so the joint hardware has to be specified with the seal rather than assumed |
| Operating temperature (°C) | Fill-fluid dependent. The clamp ratings are quoted at 70 F and 250 F, and option P specifies a non-sanitary fill fluid |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer