Rosemount 1208A Level and Flow Transmitter
Purchase options
From$353.03
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Compact 80 GHz radar for level and open channel flow, +/-2 mm and up to 15 m (10 m in oil-based media), in a submersible PVDF housing with IO-Link.
Key features
- Compact 80 GHz FMCW radar for continuous level and for volume flow in open channels, measuring +/-2 mm with a maximum range of 49 ft (15 m)
- The measuring range falls to 33 ft (10 m) for oil-based media with a dielectric constant below 10, so a 15 m radar is the wrong specification for a fuel tank
- Level is measured; open channel flow is calculated from level using the channel structure's head-flow curve, which makes the channel the limiting error, not the radar
- Accuracy is quoted excluding installation-dependent offset, so nozzle geometry and tank internals are where a real installation loses accuracy
- PVDF housing rated for outdoor conditions and submersible, with a stainless steel M12 connection: built for sumps, wet wells and open channels
- Three-wire IO-Link device with 4-20 mA and digital switch outputs, and Bluetooth models for configuration without opening the enclosure
Key specifications
- Accuracy
- +/-0.08 in (+/-2 mm) under reference conditions per IEC 60770-1, excluding installation-dependent offset. Repeatability +/-1 mm, ambient temperature effect +/-1 mm per 10 K
- Output signal
- Three-wire IO-Link device providing 4-20 mA and digital switch outputs, with process data, events and parameters over IO-Link to a master. Models with Bluetooth add local wireless configuration and maintenance
- Diagnostics
- The 15 m range is a claim about a surface that reflects well: water at a dielectric constant near 80 reflects almost completely while a hydrocarbon near 2 returns a few percent. The datasheet then reduces the range to 10 m and warns that heavy turbulence, foam and condensation with poor reflection can reduce it further
- Pressure range
- Not a pressure instrument: the process connection is a 1-1/2 in NPT or G thread with brackets, adapters and flanges available for non-pressurized applications, including PE100 plastic flanges
- Variants
- Measuring range 49 ft (15 m) maximum, limited to 33 ft (10 m) for oil-based media with a dielectric constant below 10. Process connections NPT 1-1/2 in or ISO 228/1-G1-1/2 in, with stainless steel and PE100 flange options and mounting brackets for pipe, ceiling or wall
- Application
- Continuous level in small tanks, outdoor sumps, ponds and wet wells, and volume flow in open channels for water and wastewater, where a compact non-contacting instrument replaces a pressure or float device
The footnote decides whether this instrument fits: 49 ft, or 33 ft in oil
The Rosemount 1208A is a compact 80 GHz FMCW radar transmitter for continuous level, and for volume flow in open channels. It measures ±2 mm under reference conditions with a maximum range of 49 ft (15 m) — and the datasheet's own footnote takes a third of that away for a whole class of products.
Dielectric constant is the real range specification
The measuring range is limited to 33 ft (10 m) for oil-based media with a dielectric constant below 10. That is not a tolerance; it is a different instrument capability, and it follows from how radar works. A radar measures the reflection from the surface, and the strength of that reflection depends on how much the surface differs electrically from what is above it. Water, at a dielectric constant around 80, is close to a mirror. A hydrocarbon at 2 reflects a few percent of the energy — enough to measure, over a shorter distance.
So “15 metre radar” is a correct description of the product and a wrong specification for a diesel tank, a fuel oil sump or a solvent vessel. The check belongs in the selection conversation, not in the commissioning report: what is the dielectric constant of the medium, and at what distance does it have to work? The footnote also warns that heavy turbulence, foam and condensation combined with a poorly reflecting product can reduce the range further — which is the honest way of saying that the published number assumes a cooperating surface.
Flow is arithmetic on top of level
The instrument is sold as a level and flow transmitter, and it is worth being precise about which of those is a measurement. Level is measured. Volume flow in an open channel is calculated from the level using the head–flow relationship of the channel structure — a flume or a weir — so the flow value is only as good as the geometry the instrument has been told about.
That has two practical consequences. First, the flume or weir has to be the one whose curve is being applied; a structure that has silted up, been modified, or is not the type assumed will produce a flow number that looks perfectly reasonable and is wrong. Second, the level measurement is the part that deserves the accuracy attention — at ±2 mm the level is far more precise than most channel structures are, which means the channel, not the radar, is the limiting error in a flow installation.
±2 mm, and what the caveat excludes
The accuracy figure is ±0.08 in (±2 mm) per IEC 60770-1, excluding installation-dependent offset — and that exclusion is where real installations lose accuracy. A radar mounted in a nozzle that is too narrow, aimed at a wall, or sitting in a tank with an agitator sees reflections the reference conditions do not include, and the resulting offset belongs to the installation rather than the instrument.
The supporting numbers describe a device that is quick and stable: repeatability ±1 mm, ambient temperature effect ±1 mm per 10 K, at least one update per second (typically five), and a maximum level rate of 200 mm/s — the fastest surface movement the instrument will follow. A tank that fills faster than that is a tank where the reading lags rather than lies, which is the better failure but still a specification to respect.
Built for wet places
The housing is PVDF — plastic, not painted aluminium — described as compact and robust, able to withstand outdoor conditions and submersible. The M12 housing and connector are stainless steel. For sumps, wet wells, ponds and open channels that is the right material argument: the instrument lives where it gets hosed, splashed and occasionally submerged, and a plastic housing with a sealed M12 connection is a simpler answer than a metal enclosure with a gland.
Process and ambient temperature both run −40 to 176 °F (−40 to 80 °C), and the process connection is a 1½ in NPT or G thread with a choice of brackets, adapters and flanges — including PE100 plastic flanges for non-pressurized applications, which suit water and wastewater work where a stainless flange is unnecessary and a dissimilar-metal joint is a corrosion site.
Integration: analog, digital, and a phone
The transmitter is a three-wire IO-Link device that provides 4–20 mA and digital switch outputs, with parameters and events travelling over IO-Link to a master. Models with Bluetooth add local wireless access for configuration and maintenance — which on a wet well means the technician can read and configure the instrument without opening an enclosure that is standing in a damp pit.
Where it fits
Level in small tanks and outdoor sumps and ponds, and volume flow in open channels for water and wastewater — the applications where a compact, submersible, maintenance-free non-contacting instrument beats a pressure or float device. Check the dielectric constant before specifying a range, confirm the channel structure matches the flow curve being applied, and mount it where the beam has a clear path to the surface rather than into a nozzle wall.
Manufacturer reference: Rosemount 1208A Level and Flow Transmitter Product Data Sheet 00813-0100-7062 for accuracy, measuring range and its dielectric-constant limitation, temperature limits, materials and IO-Link integration.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | +/-0.08 in (+/-2 mm) under reference conditions per IEC 60770-1, excluding installation-dependent offset. Repeatability +/-1 mm, ambient temperature effect +/-1 mm per 10 K |
| Output signal | Three-wire IO-Link device providing 4-20 mA and digital switch outputs, with process data, events and parameters over IO-Link to a master. Models with Bluetooth add local wireless configuration and maintenance |
| Diagnostics | The 15 m range is a claim about a surface that reflects well: water at a dielectric constant near 80 reflects almost completely while a hydrocarbon near 2 returns a few percent. The datasheet then reduces the range to 10 m and warns that heavy turbulence, foam and condensation with poor reflection can reduce it further |
| Pressure range | Not a pressure instrument: the process connection is a 1-1/2 in NPT or G thread with brackets, adapters and flanges available for non-pressurized applications, including PE100 plastic flanges |
| Variants | Measuring range 49 ft (15 m) maximum, limited to 33 ft (10 m) for oil-based media with a dielectric constant below 10. Process connections NPT 1-1/2 in or ISO 228/1-G1-1/2 in, with stainless steel and PE100 flange options and mounting brackets for pipe, ceiling or wall |
| Application | Continuous level in small tanks, outdoor sumps, ponds and wet wells, and volume flow in open channels for water and wastewater, where a compact non-contacting instrument replaces a pressure or float device |
| Mounting | PVDF transmitter housing described as compact and robust, able to withstand outdoor conditions and submersible, with a stainless steel M12 housing and polyamide contact body. Mounted by thread, flange or bracket depending on the structure |
| Operating temperature (°C) | Process and ambient both -40 to 176 F (-40 to 80 C), with a maximum level rate of 200 mm/s and at least one measurement update per second, typically five |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer