Type J, K or T thermocouple in a 1/4 in. sheath: Type K to 1200 C, grounded or ungrounded junctions, Class 1 or Special tolerances.
Key features
- Type J, K or T thermocouple in the same 1/4 in. sheath platform as the 214C RTD
- Ranges: Type J -40 to 760 C, Type K -40 to 1200 C, Type T -196 to 370 C - each type has its own span
- Accuracy codes: T1 Class 1 and T2 Class 2 per IEC 60584, or SP Special and ST Standard per ASTM E230
- Junction options SG single grounded, SU single ungrounded, DG dual grounded unisolated, DU dual ungrounded isolated
- 321 SST sheath, with Alloy 600 for Type K up to 1200 C (2192 F); spring loaded, adjustable and welded mounting styles
- Insertion lengths 0 to 78.5 inches in 1/4 in. increments; AR1 aluminium head or head with display cover
Key specifications
- Accuracy
- T1 Class 1 or T2 Class 2 per IEC 60584, and SP Special Tolerances or ST Standard Tolerances per ASTM E230; the two standards do not define identical limits, so the type and the standard have to be matched before comparing
- Output signal
- Thermocouple millivolt output to a transmitter or directly to the control system; junction configuration SG, SU, DG or DU determines whether the circuit is grounded to the sheath
- Diagnostics
- Paired with a 114C thermowell on the H plus U equals L rule, so the well and the sensor are ordered to the same insertion length; ungrounded isolated junctions are the usual choice where a ground loop is possible
- Pressure range
- Set by the thermowell and process connection selected with the sensor; the sheath is 1/4 in. (6 mm) nominal diameter in 321 SST, or Alloy 600 for Type K
- Variants
- Types J, K and T; accuracy T1, T2, SP or ST; single or dual elements, grounded or ungrounded, isolated or unisolated; 321 SST or Alloy 600 sheath (Alloy 600 for Type K only, to 1200 C)
- Application
- Furnace and flue gas temperature, high temperature process lines and cold service to -196 C on Type T, where an RTD runs out of range or the response time matters more than long term stability
Type J, K and T in the same sheath platform as the 214C RTD
The Rosemount 214C thermocouple is the other half of the 214C sensor family: the same 1/4 inch nominal diameter sheath, the same connection heads and mounting styles, with a Type J, K or T thermocouple element instead of a Pt100. It is the choice where the temperature goes beyond what an RTD survives, where the response has to be faster than a wire-wound element allows, or where the point simply does not justify an RTD's cost.
Three types, three different ranges — and one of them is a trap
Type J covers -40 to 760 °C (-40 to 1400 °F). Type K covers -40 to 1200 °C (-40 to 2192 °F) and is the general-purpose high-temperature choice. Type T covers -196 to 370 °C (-321 to 698 °F) and is the low-temperature one. The trap is assuming the family's headline range — the 214C literature quotes -196 to 1200 °C — applies to whichever type is ordered: each type has its own span, and a Type T ordered for a 500 °C point is outside its range entirely.
Tolerance: two standards, not one ladder
The ordering table offers four accuracy codes and they are not on a single scale. T1 is Class 1 and T2 is Class 2, both per IEC 60584. SP is Special Tolerances and ST is Standard Tolerances, both per ASTM E230. Class 1 and Special are not identical limits in every type, so a specification written as "Class 1 or better" and a quoted SP part are two different claims; the type and the standard have to be matched before the comparison is meaningful.
Junction configuration decides noise and isolation
Four options: SG single grounded, SU single ungrounded, DG dual grounded unisolated, and DU dual ungrounded isolated. A grounded junction gives the fastest response because the hot junction is in contact with the sheath, at the cost of an electrical path to the process — which matters where the thermocouple shares a circuit with a transmitter or where a ground loop is possible. An ungrounded junction is slower and isolated, which is the usual answer for transmitters and for any installation with stray currents in the pipe.
Sheath and mounting
The standard sheath is 321 stainless steel; Alloy 600 is available for Type K only and raises the sheath limit to 1200 °C (2192 °F), which is what makes a Type K usable at the top of its own range. Mounting styles include a spring-loaded adapter, an adjustable spring-loaded fitting and a welded adapter; connection heads include the aluminium AR1 and a head with a display cover. Insertion lengths run from 0 to 78.5 inches in ¼ inch increments.
Where it fits
Furnace and flue gas temperatures, high-temperature process lines, and cold service down to -196 °C on Type T — anywhere an RTD runs out of range or where the response time matters more than the long-term stability an RTD provides. It pairs with the 114C thermowell on the same H plus U equals L rule, so the well and the sensor are ordered to the same insertion length.
Manufacturer reference: Rosemount 214C Temperature Sensors Product Data Sheet, thermocouple ordering information and type codes.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | T1 Class 1 or T2 Class 2 per IEC 60584, and SP Special Tolerances or ST Standard Tolerances per ASTM E230; the two standards do not define identical limits, so the type and the standard have to be matched before comparing |
| Output signal | Thermocouple millivolt output to a transmitter or directly to the control system; junction configuration SG, SU, DG or DU determines whether the circuit is grounded to the sheath |
| Diagnostics | Paired with a 114C thermowell on the H plus U equals L rule, so the well and the sensor are ordered to the same insertion length; ungrounded isolated junctions are the usual choice where a ground loop is possible |
| Pressure range | Set by the thermowell and process connection selected with the sensor; the sheath is 1/4 in. (6 mm) nominal diameter in 321 SST, or Alloy 600 for Type K |
| Variants | Types J, K and T; accuracy T1, T2, SP or ST; single or dual elements, grounded or ungrounded, isolated or unisolated; 321 SST or Alloy 600 sheath (Alloy 600 for Type K only, to 1200 C) |
| Application | Furnace and flue gas temperature, high temperature process lines and cold service to -196 C on Type T, where an RTD runs out of range or the response time matters more than long term stability |
| Mounting | Spring loaded adapter, adjustable spring loaded fitting or welded adapter; AR1 aluminium connection head or a head with display cover; insertion lengths 0 to 78.5 in. in 1/4 in. increments |
| Operating temperature (°C) | Type J -40 to 760 C (-40 to 1400 F); Type K -40 to 1200 C (-40 to 2192 F); Type T -196 to 370 C (-321 to 698 F). Each type has its own span, so the family headline range does not apply to every order |