Skip to main content
Sold and shipped by Sproh

Rosemount 214C Thermocouple

Part number214C TCManufacturerEmerson

Purchase options

From$53.08

Availability
Call for availability Originalx
Select option

Configurable — choose options to build the exact part number.

Independent supplier

Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.

Overview

Type J, K or T thermocouple in a 1/4 in. sheath: Type K to 1200 C, grounded or ungrounded junctions, Class 1 or Special tolerances.

Key features

  • Type J, K or T thermocouple in the same 1/4 in. sheath platform as the 214C RTD
  • Ranges: Type J -40 to 760 C, Type K -40 to 1200 C, Type T -196 to 370 C - each type has its own span
  • Accuracy codes: T1 Class 1 and T2 Class 2 per IEC 60584, or SP Special and ST Standard per ASTM E230
  • Junction options SG single grounded, SU single ungrounded, DG dual grounded unisolated, DU dual ungrounded isolated
  • 321 SST sheath, with Alloy 600 for Type K up to 1200 C (2192 F); spring loaded, adjustable and welded mounting styles
  • Insertion lengths 0 to 78.5 inches in 1/4 in. increments; AR1 aluminium head or head with display cover

Key specifications

Accuracy
T1 Class 1 or T2 Class 2 per IEC 60584, and SP Special Tolerances or ST Standard Tolerances per ASTM E230; the two standards do not define identical limits, so the type and the standard have to be matched before comparing
Output signal
Thermocouple millivolt output to a transmitter or directly to the control system; junction configuration SG, SU, DG or DU determines whether the circuit is grounded to the sheath
Diagnostics
Paired with a 114C thermowell on the H plus U equals L rule, so the well and the sensor are ordered to the same insertion length; ungrounded isolated junctions are the usual choice where a ground loop is possible
Pressure range
Set by the thermowell and process connection selected with the sensor; the sheath is 1/4 in. (6 mm) nominal diameter in 321 SST, or Alloy 600 for Type K
Variants
Types J, K and T; accuracy T1, T2, SP or ST; single or dual elements, grounded or ungrounded, isolated or unisolated; 321 SST or Alloy 600 sheath (Alloy 600 for Type K only, to 1200 C)
Application
Furnace and flue gas temperature, high temperature process lines and cold service to -196 C on Type T, where an RTD runs out of range or the response time matters more than long term stability
See all specifications