Overview
DIN rail temperature transmitter with HART 7 and 4-20 mA output, universal RTD/thermocouple input, Callendar-Van Dusen sensor matching and 0.05% annual stability.
Key features
- Rail-mount (DIN rail) temperature transmitter, 23.5 mm wide, HART 7 and 4-20 mA
- Single sensor input with universal ranges: RTD, thermocouple, mV and ohms
- Transmitter-sensor matching with Callendar-Van Dusen constants
- Stability better than +/-0.05% of span per year; supply effect under 0.005% of span/V
- 7.5 to 48 Vdc ordinary location, 7.5 to 30 Vdc hazardous approval, 2.5 kVAC isolation
- -50 to +85 C ambient, IP20, NAMUR NE21/NE43/NE44/NE89 and NE107 diagnostics
Key specifications
- Accuracy
- Long-term stability better than +/-0.05% of span per year, +/-0.18% over five years; effect of supply voltage variation under 0.005% of span per volt; response time 70 ms; signal/noise ratio above 60 dB
- Output signal
- 4-20 mA with HART 7 protocol; 18-bit output signal dynamics on a 24-bit digitised input
- Power supply
- 7.5 to 48 Vdc in ordinary locations, 7.5 to 30 Vdc with hazardous-area approval; maximum internal dissipation 850 mW per channel; isolation test 2.5 kVAC (55 VAC operation, 42 VAC hazardous)
- Diagnostics
- NAMUR NE107 compliant diagnostic information, programmable damping 0 to 60 s, and write protection by jumper or software
- Certifications
- Hazardous-area approval versions available; NAMUR NE21, NE43, NE44, NE89 and NE107; vibration tested to IEC 60068-2-6 (2 to 25 Hz +/-1.6 mm, 25 to 100 Hz +/-4 g)
- Measurement range
- Single sensor universal input: RTD, thermocouple, millivolt and ohm; transmitter-sensor matching with Callendar-Van Dusen constants
Temperature conversion in the cabinet, on DIN rail
The Rosemount 248R is the rail-mount version of the 248 transmitter. Instead of travelling to the sensor head, it sits in the marshalling cabinet on a DIN rail, 23.5 mm wide, and takes the RTD or thermocouple leads from the field. It sends 4-20 mA with HART 7 on the loop.
When rail mount is the right choice
The head-mount version puts electronics where the sensor is — outdoors, inside a housing, exposed to whatever the process area offers. The rail-mount version concentrates the electronics in a controlled environment where a technician can reach them, and it wins when the sensors are simple and many, when the heads are already full, or when a cabinet is where the maintenance team actually works. IP20 says it plainly: this is equipment for inside an enclosure, not for the weather.
Accuracy and signal handling
Long-term stability is better than ±0.05% of span per year, or ±0.18% over five years, with a supply-voltage effect under ±0.005% of span per volt. Response time is 70 ms, damping is programmable from 0 to 60 seconds, and the input is digitised at 24 bits with an 18-bit output signal — the numbers that let a tight span hold up in a control loop rather than merely on a data sheet.
Sensor matching
The transmitter accepts RTD, thermocouple, millivolt and ohm inputs on a single sensor channel, and supports transmitter-sensor matching using Callendar-Van Dusen constants. That last point is where the accuracy actually comes from in practice: an individual Pt100 is not an ideal Pt100, and entering its own constants removes the sensor's error from the loop.
The electrical envelope
7.5 to 48 Vdc in ordinary locations, 7.5 to 30 Vdc with a hazardous-area approval, drawing at most 850 mW per channel, with 2.5 kVAC isolation on test. Start-up is under 2.75 seconds, with a warm-up of five minutes to full specification. Terminals take 0.13 to 2.08 mm² (AWG 26 to 14) at 0.5 Nm, and write protection is available by jumper or in software — worth setting before a loop is handed over.
Diagnostics worth wiring
Diagnostic information follows NAMUR NE107, alongside NE21 for EMC, NE43 for signal-level breakdown, NE44 for cycle-time monitoring and NE89 for digital temperature transmitters. In practice that means the HART master is told what kind of fault it is, not merely that one exists.
Manufacturer reference: Rosemount 248 Rail Mount Temperature Transmitter with RK Option and HART 7 Protocol Product Data Sheet, 00813-0200-4825 Rev AA, April 2022.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Long-term stability better than +/-0.05% of span per year, +/-0.18% over five years; effect of supply voltage variation under 0.005% of span per volt; response time 70 ms; signal/noise ratio above 60 dB |
| Output signal | 4-20 mA with HART 7 protocol; 18-bit output signal dynamics on a 24-bit digitised input |
| Power supply | 7.5 to 48 Vdc in ordinary locations, 7.5 to 30 Vdc with hazardous-area approval; maximum internal dissipation 850 mW per channel; isolation test 2.5 kVAC (55 VAC operation, 42 VAC hazardous) |
| Diagnostics | NAMUR NE107 compliant diagnostic information, programmable damping 0 to 60 s, and write protection by jumper or software |
| Certifications | Hazardous-area approval versions available; NAMUR NE21, NE43, NE44, NE89 and NE107; vibration tested to IEC 60068-2-6 (2 to 25 Hz +/-1.6 mm, 25 to 100 Hz +/-4 g) |
| Measurement range | Single sensor universal input: RTD, thermocouple, millivolt and ohm; transmitter-sensor matching with Callendar-Van Dusen constants |
| Application | Rail-mount transmitter for single-point temperature measurement, installed on DIN rail inside a cabinet or marshalling panel |
| Mounting | DIN rail mount, 23.5 mm wide module with polarity protection on all inputs and outputs |
| Dimensions | 109 x 23.5 x 104 mm (H x W x D), 150 g single input; terminals accept 0.13 to 2.08 mm² (AWG 26 to 14) at 0.5 Nm torque |
| Operating temperature (°C) | Ambient -50 to +85 °C and storage -50 to +85 °C; humidity below 99% RH non-condensing; protection degree IP20; calibration temperature 23 to 25 °C |
Documents