Rosemount 397 TUpH pH Sensor
Purchase options
From$285.41
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
TUpH pH sensor for high-solids streams: 0 to 14 pH, 0 to 100 C, 0 to 100 psig, polypropylene or Tefzel body, Quik-Loc mounting.
Key features
- TUpH pH sensor for aqueous streams with high suspended solids, model 397
- Coating-resistant design: large area reference junction, helical reference pathway, SILCORE triple seal barrier
- Percent linearity 99% from pH 2 to 12, 97% from 12 to 13, and 92 to 94% at the extremes
- 0 to 14 pH, 0 to 100 C (32 to 212 F), 0 to 100 psig, minimum conductivity 75 uS/cm
- Polypropylene or Tefzel body with an Accu-Glass hemispherical bulb behind a protective cage
- Mounts in the Quik-Loc kit (sold separately) with a 1 in. MNPT coupler; 15 ft integral cable, remote preamplifier
Key specifications
- Accuracy
- Percent linearity 94% at pH 0 to 2, 99% at pH 2 to 12, 97% at pH 12 to 13 and 92% at pH 13 to 14
- Diagnostics
- TUpH reference design: large area reference junction for ionic diffusion, helical reference pathway that slows poisoning ions, SILCORE triple seal barrier and a titanium solution ground
- Pressure range
- 0 to 100 psig (0 to 6.9 barg) with the hemispherical bulb
- Measurement range
- 0 to 14 pH with a GPHT hemispherical Accu-Glass bulb; minimum conductivity 75 uS/cm nominal
- Sensor connection
- No process thread of its own: the Quik-Loc kit supplies a 1 in. (25 mm) MNPT connection using a PEEK adapter (23753-01) and a 316 stainless Dixon coupler with an EP gasket (9160447); 15 ft integral cable with remote preamplifier (no integral preamplifier)
- Application
- pH measurement in aqueous solutions with high levels of suspended solids: slurries, lime and caustic lines, pulp stock, aeration basins and other coating-prone streams
Built for the stream that coats everything else
The Rosemount 397 is a TUpH pH sensor for aqueous streams with high suspended solids — slurries, lime and caustic lines, pulp stock, aeration basins — where an ordinary glass electrode disappears under a coating within days. The design answer is a reference junction with a large surface area and a helical path, plus a glass bulb that sits behind a protective plastic cage so the moving solids hit the cage rather than the glass.
What TUpH actually means
Three features do the work. The reference junction is large-area, so a given amount of fouling still leaves ionic contact. The reference pathway is helical, which lengthens the diffusion path so poisoning ions reach the reference slowly instead of immediately. And the whole reference is behind a SILCORE triple-seal barrier with a titanium solution ground, which is what keeps a wet, dirty environment from becoming an electrical path through the sensor body. The visible result is a sensor that keeps reading in a stream where a conventional one has already lost its reference.
The accuracy that matters
Emerson publishes percent linearity by pH band rather than as a single number, and the shape is instructive: 99% between pH 2 and 12, 97% from 12 to 13, and 92 to 94% at the extremes below pH 2 and above pH 13. A sensor quoted at one flat figure across the whole range is usually quoting a mid-range number; this table says where the glass actually behaves, which is what a control loop should be designed around.
Mechanical envelope
0 to 14 pH, 0 to 100 °C (32 to 212 °F), 0 to 100 psig on the hemispherical bulb, and a minimum conductivity of 75 µS/cm. The body is polypropylene or Tefzel, both chosen for chemical resistance rather than pressure, with 15 ft of integral cable and no preamplifier — for long runs Emerson specifies a remote preamplifier rather than an integral one, which is the correct arrangement in a wet, washdown area anyway.
Quik-Loc, and a warning worth repeating
The sensor has no process thread of its own. It mounts in the Quik-Loc kit — a PEEK adapter and a 316 stainless Dixon coupler with a 1 in. MNPT connection and locking arms — which lets the sensor be pulled for cleaning without twisting the cable, and gives visual confirmation that it is locked. Emerson explicitly does not recommend the Quik-Loc kit for hazardous, corrosive or strongly oxidizing chemicals because of spray risk during removal. That is a limitation to know before specifying it on a line where the process is nasty.
Where it fits
Wastewater aeration and digestion, pulp and paper stock lines, lime slurry and scrubber loops, and food or beverage streams with high solids where a flat or caged sensor would blind over.
Manufacturer reference: Rosemount 397 TUpH pH Sensor Product Data Sheet, MS-00813-0100-3397 Rev AA, December 2024.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Percent linearity 94% at pH 0 to 2, 99% at pH 2 to 12, 97% at pH 12 to 13 and 92% at pH 13 to 14 |
| Diagnostics | TUpH reference design: large area reference junction for ionic diffusion, helical reference pathway that slows poisoning ions, SILCORE triple seal barrier and a titanium solution ground |
| Pressure range | 0 to 100 psig (0 to 6.9 barg) with the hemispherical bulb |
| Measurement range | 0 to 14 pH with a GPHT hemispherical Accu-Glass bulb; minimum conductivity 75 uS/cm nominal |
| Sensor connection | No process thread of its own: the Quik-Loc kit supplies a 1 in. (25 mm) MNPT connection using a PEEK adapter (23753-01) and a 316 stainless Dixon coupler with an EP gasket (9160447); 15 ft integral cable with remote preamplifier (no integral preamplifier) |
| Application | pH measurement in aqueous solutions with high levels of suspended solids: slurries, lime and caustic lines, pulp stock, aeration basins and other coating-prone streams |
| Mounting | Inserted through the Quik-Loc kit, which allows removal without twisting the cable and gives visual confirmation of the locked position |
| Material of construction | Polypropylene, ethylene-propylene (EP) and glass wetted materials; Tefzel body option; weight 0.45 kg (1 lb) |
| Operating temperature (°C) | 0 to 100 C (32 to 212 F) |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer