Rosemount 398 / 398VP TUpH pH and ORP Sensor
Purchase options
From$409.91
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
TUpH pH/ORP sensor with a large area reference junction: pH 0 to 14 or ORP +/-1500 mV, titanium body, 0 to 100 C, 0 to 250 psig.
Key features
- TUpH pH/ORP sensor, models 398 and 398VP, with a large area reference junction and helical path
- pH 0 to 14 or ORP -1500 to +1500 mV with a platinum electrode (model code 12)
- Titanium body, Tefzel shroud, glass bulb and Kalrez, Viton or EPDM seal options
- Percent linearity 99% at pH 2 to 12, 97% at 12 to 13 and 92% at 13 to 14 on the hemispherical bulb
- 0 to 100 C, 0 to 250 psig hemispherical or 0 to 100 psig flat bulb, minimum conductivity 75 uS/cm
- No process thread: needs the 1 in. NPT connector 23166-00 or 23166-01; 398 has 15 ft cable, 398VP has a Variopol plug
Key specifications
- Accuracy
- Percent linearity on the hemispherical bulb: 94% at pH 0 to 2, 99% at pH 2 to 12, 97% at pH 12 to 13 and 92% at pH 13 to 14; the flat bulb gives 93%, 98% and 95% for the first three bands
- Diagnostics
- TUpH reference design: large area reference junction for ionic diffusion and a helical reference pathway that slows the effect of poisoning ions
- Pressure range
- 0 to 250 psig (100 to 1825 kPa abs) with the hemispherical bulb; 0 to 100 psig (100 to 790 kPa abs) with the flat bulb
- Measurement range
- pH 0 to 14, or ORP -1500 to +1500 mV on the platinum electrode version (model code 12)
- Sensor connection
- No process thread of its own: requires a 1 in. NPT process connector 23166-00 or 23166-01 (sold separately); 398 has 15 ft integral cable without preamplifier (4 ft and 10 ft options), 398VP has a Variopol connector
- Application
- pH or ORP measurement in dirty, coating and poisoning-prone streams: wastewater, chemical process, pulp and paper, and dosing control loops such as chlorination, cyanide destruction and chromate reduction
One sensor for pH and ORP, built for the dirty end of the plant
The Rosemount 398 and 398VP are TUpH pH/ORP sensors: a titanium body with a Tefzel shroud around the glass bulb, a large-area reference junction, and the option of ORP instead of pH in the same housing. They are specified where the stream coats, blinds or poisons — the reason the reference design is the product, not the accessory.
pH and ORP in one platform
The pH version measures 0 to 14 pH; the ORP version measures -1500 to +1500 mV with a platinum electrode. Both share the body, the reference and the mounting, which matters practically: a plant running both measurements can stock one mounting arrangement and one set of connectors instead of two families of hardware. Model code 12 selects ORP.
Why the reference is the specification
The TUpH design uses a large area junction so fouling does not immediately break ionic contact between the reference and the process, and a helical internal path so poisoning ions must diffuse a long way before reaching the reference element. On a stream carrying sulphides, cyanides or heavy metals — the classic reference poisons — that path is what separates months of service from weeks.
What you get in percent
Emerson publishes linearity by pH band: 94% from 0 to 2, 99% from 2 to 12, 97% from 12 to 13 and 92% from 13 to 14 on the hemispherical bulb. The flat bulb gives 93%, 98% and 95% in the first three bands and is not offered for the top band. Using the band that matches the process is more honest than quoting one mid-range figure.
Pressure, temperature and the parts that are not included
Operating range is 0 to 100 °C, with 250 psig on the hemispherical bulb and 100 psig on the flat bulb. Wetted materials are titanium, Tefzel, glass and a choice of Kalrez, Viton or EPDM seals, plus platinum for the ORP electrode. The sensor has no process thread of its own: it mounts through a 1 inch NPT process connector, 23166-00 or 23166-01, sold separately — and on the retractable versions a ball valve kit is what allows insertion and removal under pressure.
Versions
The 398 terminates in 15 ft of integral cable without a preamplifier (4 ft and 10 ft are options), and the 398VP replaces the cable with a Variopol connector so the sensor can be disconnected without pulling cable through conduit. The R and RVP versions are the retractable variants for insertion through a valve.
Where it fits
Wastewater aeration and collection, chemical process streams with high solids, pulp and paper, and any loop where ORP controls dosing — chlorination, cyanide destruction, chromate reduction — and the electrode must survive the same stream the pH sensor does.
Manufacturer reference: Rosemount 398/398VP/398R/398RVP TUpH pH/ORP Sensors Product Data Sheet, 00813-0100-3398, November 2019.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Percent linearity on the hemispherical bulb: 94% at pH 0 to 2, 99% at pH 2 to 12, 97% at pH 12 to 13 and 92% at pH 13 to 14; the flat bulb gives 93%, 98% and 95% for the first three bands |
| Diagnostics | TUpH reference design: large area reference junction for ionic diffusion and a helical reference pathway that slows the effect of poisoning ions |
| Pressure range | 0 to 250 psig (100 to 1825 kPa abs) with the hemispherical bulb; 0 to 100 psig (100 to 790 kPa abs) with the flat bulb |
| Measurement range | pH 0 to 14, or ORP -1500 to +1500 mV on the platinum electrode version (model code 12) |
| Sensor connection | No process thread of its own: requires a 1 in. NPT process connector 23166-00 or 23166-01 (sold separately); 398 has 15 ft integral cable without preamplifier (4 ft and 10 ft options), 398VP has a Variopol connector |
| Application | pH or ORP measurement in dirty, coating and poisoning-prone streams: wastewater, chemical process, pulp and paper, and dosing control loops such as chlorination, cyanide destruction and chromate reduction |
| Mounting | Insertion or submersion through the process connector; the R and RVP retractable versions use a ball valve kit (1-1/4 in. or 1-1/2 in.) for insertion and removal under pressure; minimum conductivity 75 uS/cm nominal |
| Material of construction | Titanium body, Tefzel, glass, and a choice of Kalrez, Viton or EPDM seals; platinum electrode for ORP |
| Operating temperature (°C) | 0 to 100 C (32 to 212 F) |