Overview
Retractable TUpH pH/ORP sensor removed through a ball valve for service without stopping the process: pH 0 to 14 or ORP +/-1500 mV.
Key features
- TUpH retractable pH/ORP sensor, models 398R and 398RVP, withdrawn through a ball valve for service
- pH 0 to 14 (hemispherical or flat AccuGLASS bulb) or ORP -1500 to +1500 mV with a platinum electrode
- 250 psig at 100 C in service, but retraction or insertion limited to 64 psig (code 21) or 35 psig (code 25)
- Titanium tube, Tefzel, PEEK internals, glass and Viton or Kalrez seals
- Permeable Tefzel reference with a secondary helical pathway and SILCORE contaminant barrier
- Ball valve kit 1-1/4 in. or 1-1/2 in. sold separately; 15 ft cable standard, 9.5 in. option for a sensor head junction box
Key specifications
- Accuracy
- Percent linearity on the hemispherical bulb 94% at pH 0 to 2, 99% at pH 2 to 12, 97% at pH 12 to 13 and 92% at pH 13 to 14; the flat bulb gives 93%, 98% and 95% for the first three bands
- Diagnostics
- TUpH reference: permeable Tefzel junction with a secondary helical pathway behind a SILCORE contaminant barrier, designed for streams that coat, foul and poison conventional electrodes
- Pressure range
- 250 psig at 100 C in service; maximum pressure at retraction or insertion is 64 psig (524 kPa abs) with code 21 and 35 psig (343 kPa abs) with code 25
- Measurement range
- pH 0 to 14 on the GPHT hemispherical or GPLR flat AccuGLASS bulb, or ORP -1500 to +1500 mV with a platinum electrode
- Sensor connection
- No process thread of its own: mounts through a ball valve kit (1-1/4 in. or 1-1/2 in., sold separately) or a 1 in. NPT process connector; 15 ft integral cable standard with a 9.5 in. option for a sensor head junction box fitted by compression fitting; RVP version has a Variopol connector
- Application
- Retractable pH or ORP measurement in processes that cannot be shut down for sensor service: sour water strippers, chlorine dioxide pulp bleaching towers, lime slurry, waste treatment and solvent-bearing streams
The sensor you can pull out without stopping the process
The Rosemount 398R and 398RVP are the retractable versions of the 398 TUpH pH/ORP sensor. The measurement end is the same titanium tube with a Tefzel-shrouded AccuGLASS bulb and the TUpH reference; what changes is that the whole sensor travels through a ball valve, so it can be withdrawn for calibration or cleaning while the line stays live.
What retractable actually buys
On a sour water stripper or a pulp bleaching tower, taking the line down to clean a pH electrode is not an option, and a sensor left in place until the next shutdown is a sensor nobody trusts by then. The 398R removes that trade: the sensor is withdrawn through the valve, serviced on the bench, and inserted again — with the valve closed while it is out.
The pressure limit is the whole story
Maximum process pressure is 250 psig at 100 °C, but that figure applies to the sensor in service, not to the sensor moving. At retraction or insertion the limit drops to 64 psig with code 21 and 35 psig with code 25 — measured against the pressure the packing sees as the sensor slides through. A line running at 100 psig can therefore use a 398R, but not pull it while running. That distinction decides whether a retractable sensor is the right answer at all, and it is the specification most often read as a formality.
Measurement, unchanged
pH 0 to 14 on the hemispherical or flat AccuGLASS bulb, or ORP from -1500 to +1500 mV with a platinum electrode. Percent linearity is 94, 99, 97 and 92 across the four pH bands on the hemispherical bulb, and 93, 98 and 95 on the flat bulb for the first three. Minimum conductivity is 75 µS/cm nominal.
Construction
Titanium tube, Tefzel, glass, PEEK internals and a choice of Viton or Kalrez seals, with platinum on the ORP version. The reference is permeable Tefzel with a secondary helical pathway behind a SILCORE contaminant barrier — the combination that lets the sensor sit in a stream that coats and poisons other electrodes without losing its reference.
Mounting and cable
The sensor has no process thread: it mounts through the ball valve kit (1-1/4 in. or 1-1/2 in., sold separately) or a 1 inch NPT process connector. Cable is 15 ft integral as standard, with a 9.5 in. option for the 398R when a sensor head junction box is fitted to the tube through a compression fitting — the arrangement used where the transmitter sits at the sensor rather than in the rack. The RVP version replaces the cable with a Variopol plug.
Where it fits
Sour water strippers, chlorine dioxide bleaching towers, lime slurry and waste treatment lines, and process streams with organic solvents — anywhere the electrode fouls fast enough to need cleaning on a schedule but the line cannot be shut down to do it.
Manufacturer reference: Rosemount Models 398 and 398R pH/ORP Sensors Product Data Sheet, PDS 71-398 Rev F, and 00813-0100-3398 for the current ordering information.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Percent linearity on the hemispherical bulb 94% at pH 0 to 2, 99% at pH 2 to 12, 97% at pH 12 to 13 and 92% at pH 13 to 14; the flat bulb gives 93%, 98% and 95% for the first three bands |
| Diagnostics | TUpH reference: permeable Tefzel junction with a secondary helical pathway behind a SILCORE contaminant barrier, designed for streams that coat, foul and poison conventional electrodes |
| Pressure range | 250 psig at 100 C in service; maximum pressure at retraction or insertion is 64 psig (524 kPa abs) with code 21 and 35 psig (343 kPa abs) with code 25 |
| Measurement range | pH 0 to 14 on the GPHT hemispherical or GPLR flat AccuGLASS bulb, or ORP -1500 to +1500 mV with a platinum electrode |
| Sensor connection | No process thread of its own: mounts through a ball valve kit (1-1/4 in. or 1-1/2 in., sold separately) or a 1 in. NPT process connector; 15 ft integral cable standard with a 9.5 in. option for a sensor head junction box fitted by compression fitting; RVP version has a Variopol connector |
| Application | Retractable pH or ORP measurement in processes that cannot be shut down for sensor service: sour water strippers, chlorine dioxide pulp bleaching towers, lime slurry, waste treatment and solvent-bearing streams |
| Mounting | Retractable insertion through the ball valve so the sensor can be withdrawn for calibration and maintenance without a process shutdown; minimum conductivity 75 uS/cm nominal |
| Material of construction | Titanium tube, Tefzel, glass, PEEK internal construction, platinum on the ORP version, with a choice of Viton or Kalrez seals; 398R weighs 0.9 kg (2 lb) with code 21 |
| Operating temperature (°C) | 0 to 100 C (32 to 212 F) at up to 250 psig (1825 kPa abs) in service |