Overview
Retractable contacting conductivity sensor with titanium electrodes that can be withdrawn through a 1-1/4 in full-port ball valve without shutting down the line.
Key features
- Retractable contacting conductivity sensor, withdrawn through a 1-1/4 in full-port ball valve
- Titanium electrodes, glass-filled PEEK insulator and 316 stainless steel body
- Cell constants of 0.01, 0.1 or 1.0/cm, factory measured
- 32 to 212 °F (0 to 100 °C), 200 psig, vacuum to 1.6 in. Hg
- Pt 1000 RTD standard
- 402 with 10 ft integral cable, or 402VP with Variopol VP6 quick disconnect
Key specifications
- Measurement principle
- Retractable contacting (electrolytic) conductivity - titanium electrodes, glass-filled PEEK insulator and 316 stainless steel tube
- Process temperature
- 32 to 212 °F (0 to 100 °C)
- Pressure rating
- 200 psig (1481 kPa abs); suitable for vacuum down to 1.6 in. Hg
- Process connection
- Sensor is inserted through a 1-1/4 in NPT full-port ball valve; the retraction assembly (316 stainless steel valve kit) is ordered separately
- Wetted materials
- Titanium electrodes, glass-filled PEEK insulator, 316 stainless steel body, EPDM O-ring and Neoprene washer; 15 micro-in. (0.4 um) Ra finish
- Measurement range
- Clean, non-corrosive liquids with conductivity below 20,000 uS/cm
A conductivity sensor you can pull out of a live line
The 402 is the retractable version of the Endurance contacting sensor: titanium electrodes and a PEEK insulator on a long stainless tube that slides through a 1¼ in full-port ball valve. Close the valve, withdraw the sensor, clean or verify it, push it back — without shutting the line down or draining it.
What retractable buys you
Cleaning is the maintenance task that actually keeps a conductivity reading honest, and on a fixed sensor it means a shutdown or a bypass. On the 402 it is a two-minute job at the valve. That changes the maintenance plan: instead of calibrating around a fouled cell, the cell gets cleaned and checked against a standard on a schedule, and the reading stays traceable between verifications. In pharmaceutical and food service, where the sensor has to come out for cleaning validation anyway, the retraction assembly is not a convenience — it is how the instrument stays in compliance.
Installing it
The valve kit is a separate item (316 stainless, 1¼ in full port), welded or threaded into the line; the sensor passes through it on its packing gland. Set the insertion depth so the electrode pair sits in the flow and not in the branch — the retraction tube has enough travel that it is easy to leave the tip short of the stream and read the wall instead. The packing gland is what seals the sensor while it moves, so it is the part to check if a drip appears at the valve when the sensor is withdrawn.
Range and limits
Same electrical family as the fixed 400: cell constants of 0.01, 0.1 or 1.0/cm, factory measured, for clean non-corrosive liquids below 20,000 µS/cm. The retractable body is rated to 212 °F (100 °C) and 200 psig — lower than the fixed sensor's high-temperature option, because the EPDM O-ring and Neoprene washer in the gland set that limit rather than the metal. Pt 1000 RTD is standard.
Two versions
The 402 has 10 ft of integral cable; the 402VP ends in a Variopol VP6 connector so the cable separates at the sensor head — useful on a retractable, where the sensor moves and the cable should not. An integral junction box (option -60) is available for longer runs.
Where it fits
High purity water, pharmaceutical and food lines, CIP circuits and any conductivity point where the sensor has to be removed for cleaning without stopping production.
Manufacturer reference: Rosemount 402 and 402VP Contacting Conductivity Sensors Product Data Sheet, 00813-0100-3402 Rev AB.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | Retractable contacting (electrolytic) conductivity - titanium electrodes, glass-filled PEEK insulator and 316 stainless steel tube |
| Process temperature | 32 to 212 °F (0 to 100 °C) |
| Pressure rating | 200 psig (1481 kPa abs); suitable for vacuum down to 1.6 in. Hg |
| Process connection | Sensor is inserted through a 1-1/4 in NPT full-port ball valve; the retraction assembly (316 stainless steel valve kit) is ordered separately |
| Wetted materials | Titanium electrodes, glass-filled PEEK insulator, 316 stainless steel body, EPDM O-ring and Neoprene washer; 15 micro-in. (0.4 um) Ra finish |
| Measurement range | Clean, non-corrosive liquids with conductivity below 20,000 uS/cm |
| Cable connection | 10 ft (3 m) integral cable on the 402; Variopol VP6 quick disconnect on the 402VP, with the mating VP cable ordered separately |
| Cell constant | 0.01, 0.1 or 1.0/cm, factory measured so no initial calibration is needed |