Rosemount 403 / 403VP Endurance Conductivity Sensor
Purchase options
From$475.60
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Sanitary flange contacting conductivity sensor for water for injection and purified water, with titanium electrodes, 0.01/0.1/1.0 per cm cell constants and steam sterilisation to 275 F.
Key features
- Sanitary flange contacting conductivity sensor for WFI and purified water
- 1-1/2 in or 2 in stainless steel sanitary flange connection
- Cell constants of 0.01, 0.1 or 1.0/cm
- Titanium electrodes with 316 SST, PCTFE and EP wetted materials, 21 CFR 177 compliant
- 32 to 221 °F (0 to 105 °C) operating, steam sterilizable to 275 °F (135 °C)
- 250 psig; certificate of conformance with every sensor
Key specifications
- Measurement principle
- Contacting (electrolytic) conductivity in a sanitary flange body - titanium electrodes with a PCTFE insulator
- Process temperature
- 32 to 221 °F (0 to 105 °C); steam sterilizable to 275 °F (135 °C)
- Pressure rating
- 250 psig (1825 kPa abs) maximum
- Process connection
- 1-1/2 in or 2 in stainless steel sanitary flange (tri-clamp)
- Wetted materials
- Titanium electrodes; 316 stainless steel; PCTFE (Neoflon) compliant with 21 CFR 177.1380; EP compliant with 21 CFR 177.2600 and USP Class VI; 16 micro-in. (0.4 um) Ra finish
- Certifications
- Wetted polymers 21 CFR 177 compliant and O-rings USP Class VI; certificate of conformance supplied with every sensor; option -99Q8 adds supplier material traceability certificates
The sanitary sensor for water for injection
A 1½ in or 2 in sanitary flange sensor with titanium electrodes, built for the water systems a pharmaceutical plant validates: water for injection and purified water. It mounts on a tri-clamp ferrule like any other hygienic instrument, steam-sterilises in place, and reads the low conductivity those systems run at.
Why WFI needs its own sensor
Water for injection is specified by conductivity as well as by chemistry, and the measurement is made against a tight limit at a low value — which means the cell constant has to match the band, not just the pipe. The 0.01/cm cell is what covers the WFI range; the 0.1 and 1.0 cells cover purified water and the cleaning solutions around it. All three use the same body, so a skid can carry one spare and change cells to suit the loop.
Materials that carry the compliance
The wetted polymers are 21 CFR 177 compliant and the O-rings meet USP Class VI, which is the documentation a validation department asks for — not a marketing claim. Every sensor ships with a certificate of conformance, and option -99Q8 adds supplier material traceability certificates if the quality system requires the paper trail down to the polymer batch. Wetted surfaces are finished to 16 micro-inches Ra so the sensor can be cleaned in place along with the rest of the line.
Temperature and sterilisation
Operating range is 32 to 221 °F (0 to 105 °C), and the sensor is steam sterilizable to 275 °F (135 °C). That second number is the one that matters in a WFI loop: the sensor has to survive SIP cycles that run hotter than any normal operating condition, and a cell that drifts after its first sterilisation is a sensor that fails its next calibration.
Installation and versions
Fits a standard sanitary ferrule with a clamp and gasket. The 403 has 10 ft of integral cable (50 ft optional); the 403VP ends in a Variopol connector so the cable is separate and the sensor can be replaced without re-pulling the run. An extended insertion length option adds 6 in from the inside face of the flange to the end of the sensor, for lines where the standard length would not reach the flow.
Where it fits
Pharmaceutical water systems — WFI and purified water loops, stills, storage and distribution — and food and beverage water purification where a sanitary connection and FDA-compliant wetted materials are required.
Manufacturer reference: Rosemount 403/403VP Contacting Conductivity Sensors Product Data Sheet, LIQ-PDS-403, and Quick Start Guide 00825-0100-3403.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | Contacting (electrolytic) conductivity in a sanitary flange body - titanium electrodes with a PCTFE insulator |
| Process temperature | 32 to 221 °F (0 to 105 °C); steam sterilizable to 275 °F (135 °C) |
| Pressure rating | 250 psig (1825 kPa abs) maximum |
| Process connection | 1-1/2 in or 2 in stainless steel sanitary flange (tri-clamp) |
| Wetted materials | Titanium electrodes; 316 stainless steel; PCTFE (Neoflon) compliant with 21 CFR 177.1380; EP compliant with 21 CFR 177.2600 and USP Class VI; 16 micro-in. (0.4 um) Ra finish |
| Certifications | Wetted polymers 21 CFR 177 compliant and O-rings USP Class VI; certificate of conformance supplied with every sensor; option -99Q8 adds supplier material traceability certificates |
| Measurement range | Low-conductivity clean water service: water for injection, purified water and the cleaning solutions around them |
| Cable connection | 10 ft (3.1 m) integral cable standard, 50 ft (15.2 m) optional; the 403VP uses a Variopol connector with the interconnecting cable sold separately |
| Cell constant | 0.01, 0.1 or 1.0/cm |
| Dimensions | 2 lb (1.0 kg) net, 3 lb (1.5 kg) shipping weight with the 10 ft cable; extended insertion length option adds 6 in from the flange face |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer