Rosemount 628 Universal Gas Sensor

Part number628ManufacturerEmerson

Purchase options

From$323.85

Availability
Call for availability Originalx
Select option

Configurable — choose options to build the exact part number.

Independent supplier

Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.

Overview

Hot-swappable gas sensor cartridge for the 928 monitor: EC electrochemical or MP combustible, with calibration stored in the sensor itself.

Key features

  • Hot-swappable sensor cartridge for the Rosemount 928 wireless gas monitor, changed in the field without tools
  • Technology codes EC electrochemical (H2S, CO, O2) and MP combustible MEMS - the gas decides the technology, so the two are not free combinations
  • Gas codes do not look like the chemicals: T02 hydrogen sulfide, A03 oxygen, T04 carbon monoxide, F41 hydrocarbon and hydrogen
  • Unit and range follow the gas: percent LEL for combustibles, ppm for H2S and CO, percent by volume for oxygen
  • The sensor stores its own calibration and moves between transmitters already calibrated, but must be connected to a 928 to be calibrated
  • End-of-life diagnostics exist on the H2S and CO sensors only; the oxygen and combustible cells depend on a scheduled replacement interval

Key specifications

Accuracy
By gas as published for the 928 with this sensor: H2S +/-3 ppm or 10 percent of reading, CO +/-6 ppm or 10 percent, O2 +/-0.5 percent of reading at or above 4 percent and +/-1.2 percent below, combustible +/-3 percent LEL methane and +/-5 percent LEL hydrogen
Output signal
Not applicable; the cartridge reports through the 928 transmitter over WirelessHART. The sensor is a smart device that carries its own calibration data to whichever transmitter it is installed in
Diagnostics
End-of-life diagnostics exist on the H2S and CO sensors only. The oxygen and combustible cells carry no life warning, so an O2 monitor depends on a scheduled replacement interval rather than on the instrument reporting its own depletion
Pressure range
Not applicable - an ambient air sensor cartridge that samples the atmosphere at the sensor head, with no process connection
Variants
Model 628 EC T02 2 F as a typical example. Technology EC (electrochemical) or MP (combustible MEMS); gas T02 hydrogen sulfide, A03 oxygen, T04 carbon monoxide, F41 hydrocarbon and hydrogen; unit 1 percent LEL, 2 ppm, 3 percent by volume; range F 0-100 for H2S, hydrocarbon and hydrogen, D 0-25 for oxygen
Application
Replacement and spare gas sensor cartridges for Rosemount 928 wireless gas monitors, and for building a gas mix across a site from one transmitter type