Overview
4000 NTU formazin turbidity standard (905-761854), the replacement solution for the premixed kit and the primary reference for turbidity calibration.
Key features
- 4000 NTU formazin turbidity standard, part number 905-761854
- Replacement solution for the 4000 NTU premixed kit
- Primary reference for turbidity calibration; lower points made by dilution with turbidity-free water
- Composition: 0.5% hydrazine sulfate, 5% hexamethylenetetramine, 94.5% water
- Mix before use; formazin settles and an old bottle reads low
- SDS carries an EPCRA Section 313 note for hydrazine sulfate - read before disposal
Key specifications
- Diagnostics
- Formazin is a reacting suspension: it settles and ages, so the bottle is mixed before use and an old settled bottle reads low regardless of instrument condition; the SDS carries an EPCRA Section 313 note for hydrazine sulfate
- Variants
- Supplied at 4000 NTU so lower calibration points are prepared by dilution with turbidity-free water rather than ordered separately
- Application
- Calibration and verification of Rosemount turbidity instruments and their 4000 NTU premixed kits, in drinking water, wastewater and industrial water treatment
- Kit contents
- One bottle of 4000 NTU formazin turbidity standard, part number 905-761854, as the replacement solution for the premixed kit
- Material of construction
- Aqueous solution of 0.5% hydrazine sulfate and 5% hexamethylenetetramine in 94.5% water; turbid appearance, specific gravity approximately 1.0
The reference the turbidity meter is actually judged against
The 905-761854 is a 4000 NTU formazin turbidity standard — the replacement solution for the premixed kit, and the primary reference a Rosemount turbidity instrument is calibrated against. Formazin is not a convenience: it is the only turbidity standard defined by how it is made rather than by what someone measured it to be, which is why instrument specifications are quoted against it.
Why 4000 NTU and not 10
The standard is supplied at the top of the working range so that lower points are made by dilution rather than shipped separately: 4000 NTU diluted with turbidity-free water covers the whole calibration curve from one bottle. The dilution has to be done with water that is genuinely at zero, because a 20 NTU point made with water containing 0.5 NTU is off by 2.5% before the instrument is even involved.
What is in it, and why that matters
The solution is water with 0.5% hydrazine sulfate and 5% hexamethylenetetramine, which react slowly to form the suspended polymer that scatters light. Because the reaction continues, formazin has a shelf life and a settling characteristic: the standard is mixed before use, it is not filtered, and an old bottle that has settled reads low no matter how well the meter is working. The SDS also carries an EPCRA Section 313 note for hydrazine sulfate — worth reading before disposal, since this is a chemical and not water with chalk in it.
Where it fits
Calibration and verification of Rosemount 4000 NTU premixed kits and the turbidity instruments they serve, in drinking water, wastewater and industrial water treatment — the applications where the turbidity reading is a regulatory number and the standard is part of the evidence chain.
Practical handling
Store sealed and out of direct sunlight, let the bottle reach room temperature before use so condensation does not form on the optics, and never return used standard to the bottle. The SDS lists freezing at 0 °C and boiling at 100 °C, which describes the water base rather than a recommendation for storing the standard.
Manufacturer reference: Material Safety Data Sheet for 905-761854, 4000 NTU Formazin Turbidity Standard, Rosemount Analytical, revised 01/26/11.
Specifications
| Attribute | Value |
|---|---|
| Diagnostics | Formazin is a reacting suspension: it settles and ages, so the bottle is mixed before use and an old settled bottle reads low regardless of instrument condition; the SDS carries an EPCRA Section 313 note for hydrazine sulfate |
| Variants | Supplied at 4000 NTU so lower calibration points are prepared by dilution with turbidity-free water rather than ordered separately |
| Application | Calibration and verification of Rosemount turbidity instruments and their 4000 NTU premixed kits, in drinking water, wastewater and industrial water treatment |
| Kit contents | One bottle of 4000 NTU formazin turbidity standard, part number 905-761854, as the replacement solution for the premixed kit |
| Material of construction | Aqueous solution of 0.5% hydrazine sulfate and 5% hexamethylenetetramine in 94.5% water; turbid appearance, specific gravity approximately 1.0 |