Service kit (9160487) for Rosemount Hx438 dissolved oxygen sensors: three FDA-approved membrane modules, O-rings, polishing tool and 20 ml electrolyte.
Key features
- Service kit for Rosemount Hx438 dissolved oxygen sensors, part number 9160487
- Three FDA-approved membrane modules, O-rings, polishing tool and 20 ml electrolyte
- Renews membrane, electrolyte and electrode surface in one service, as the procedure intends
- FDA-approved membranes for fermentation, pharmaceutical and food-contact applications
- Separate parts for the consumables: 9160489 30 ml electrolyte and 24262-00 cathode cartridge
- Manufacturer reference 00825-0100-3494 documents the replacement procedure
Key specifications
- Diagnostics
- Renew the membrane and electrolyte when the sensor no longer reaches its expected current in air or responds sluggishly to saturated air
- Application
- Scheduled service kit for the Rosemount Hx438 dissolved oxygen sensor, renewing the membrane, electrolyte and electrode surface in one procedure
- Mounting
- Membrane cartridge unscrewed with the sensor held vertically and the membrane pointing down, glass tip polished with the cloth, fresh electrolyte added and a new membrane module fitted
- Transmitter compatibility
- Rosemount Hx438 dissolved oxygen sensor
- Kit contents
- Three FDA-approved membrane modules, cartridge O-rings, one polishing tool or polishing paper, and 20 ml of electrolyte solution
- Material of construction
- FDA-approved membrane modules with the O-rings for the membrane cartridge; electrolyte solution in a 20 ml bottle
Everything a dissolved oxygen sensor needs at its next service
The 9160487 sensor service kit is the consumable set for the Rosemount Hx438 dissolved oxygen sensor: three FDA-approved membrane modules, O-rings, a polishing tool and 20 ml of electrolyte. It is the kit that goes on the maintenance order, not the one that sits in a drawer — a membrane has a working life, and when it is spent the sensor reads low or slow regardless of how well the electronics are working.
Why one kit rather than four parts
The membrane, the electrolyte and the electrode surface are one measurement chain, and each disturbs the others. Renewing the membrane without polishing the glass tip leaves the old film under the new membrane; adding electrolyte to a stretched membrane changes the reference without fixing the diffusion path. The kit exists so the three steps happen together, in the order the procedure specifies: unscrew the cartridge, polish the tip with the cloth, fill with fresh electrolyte, fit a new membrane module.
The FDA detail
The three membrane modules are FDA-approved. That matters only in applications where it matters — fermentation, pharmaceutical water and food-contact processes — but in those applications it is the difference between a usable spare and one that cannot be installed. Buying the kit by part number rather than by appearance is the only way to be sure which membrane is in the box.
What is in it
Three membrane modules, the O-rings for the cartridge, one polishing tool or paper, and a 20 ml bottle of electrolyte. Larger electrolyte is available separately (9160489, 30 ml) for sites that go through membranes faster than kits, and the cathode cartridge (24262-00) is a separate part for when the electrode itself has aged out.
Where it fits
Rosemount Hx438 installations on fermentation, aeration basins and industrial water loops, on the normal service interval or whenever a calibration check shows a sensor that no longer reaches its expected current in air.
Manufacturer reference: Rosemount Hx438 Dissolved Oxygen Sensor Quick Start Guide, 00825-0100-3494, section 4.2 and the spares table.
Specifications
| Attribute | Value |
|---|---|
| Diagnostics | Renew the membrane and electrolyte when the sensor no longer reaches its expected current in air or responds sluggishly to saturated air |
| Application | Scheduled service kit for the Rosemount Hx438 dissolved oxygen sensor, renewing the membrane, electrolyte and electrode surface in one procedure |
| Mounting | Membrane cartridge unscrewed with the sensor held vertically and the membrane pointing down, glass tip polished with the cloth, fresh electrolyte added and a new membrane module fitted |
| Transmitter compatibility | Rosemount Hx438 dissolved oxygen sensor |
| Kit contents | Three FDA-approved membrane modules, cartridge O-rings, one polishing tool or polishing paper, and 20 ml of electrolyte solution |
| Material of construction | FDA-approved membrane modules with the O-rings for the membrane cartridge; electrolyte solution in a 20 ml bottle |