Rosemount 9210438 Fill Solution

Part number9210438ManufacturerEmerson

No product images are published for this product yet.

Purchase options

$48.33

Availability
Call for availability Originalx

Independent supplier

Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.

Overview

Fill solution for total chlorine sensors: 7.1 percent KCl with 1.6 percent KI, the iodide that lets the sensor read chloramines as well as free chlorine.

Key features

  • Fill solution 9210438 for Rosemount total chlorine sensors: 91.3 percent water, 7.1 percent KCl, 1.6 percent KI, 0.0005 percent AgI
  • The iodide is the reagent, not just the electrolyte: it is oxidised by both free chlorine and chloramines, so the sensor reports the total rather than the free fraction
  • The electrolyte decides the parameter - a total chlorine sensor filled with the free chlorine solution stops reporting chloramines and silently becomes a free chlorine meter
  • Silver iodide fixes the reference potential in the same role silver chloride plays in the chloride solutions; the halide has to match the electrolyte
  • Hazard profile is the most serious of the family: KI at 1.6 percent carries H360 reproductive toxicity and H372 thyroid target organ warning
  • For chloraminated drinking water distribution, where the compliance measurement is total chlorine and a free chlorine reading understates the residual

Key specifications

Accuracy
Not a measured property: the iodide in the electrolyte is what makes the sensor respond to combined chlorine as well as free, so the solution defines the parameter rather than carrying an accuracy figure
Output signal
Not applicable; the solution is the electrolyte between the membrane and the electrodes and carries the ionic current the sensor measures
Diagnostics
The most serious hazard profile of the family: potassium iodide at 1.6 percent carries H360 (may damage fertility or the unborn child) and H372 (damages the thyroid through prolonged or repeated exposure), with H402 aquatic. Gloves, no eating or drinking at the bench, and no handling by anyone pregnant without an assessment
Pressure range
Not applicable - electrolyte supplied in a bottle for sensor service, with no process connection and no pressure rating
Variants
Part 9210438. Composition 91.3 percent water, 7.1 percent KCl, 1.6 percent potassium iodide, 0.0005 percent silver iodide. Distinct from #1 (7.5 percent KCl, dissolved oxygen), #4 (15-40 percent KCl, free chlorine) and #3 (11 percent KBr, dissolved ozone)
Application
Fill solution for Rosemount amperometric total chlorine sensors in chloraminated drinking water distribution, where monochloramine is the residual disinfectant and the compliance measurement is total chlorine
See all specifications