Rosemount 9310100 Ferrule
No product images are published for this product yet.
Purchase options
$56.37
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
1 in compression ferrule for Rosemount process connectors, the ring that deforms onto the sensor tube to grip it and seal the joint.
Key features
- 1 in compression ferrule, Rosemount part 9310100
- Deforms onto the sensor tube to grip it and make the joint seal
- A single-use part: replace it whenever the connector is broken
- Fits the 1 in compression process connectors 23166-00, 23166-01 and 9510066
- A scored ferrule seeps rather than fails, which is why a spare matters
- Connector assemblies are ordered separately
Key specifications
- Sensor connection
- Fits the 1 in compression process connectors 23166-00 (316 SST), 23166-01 (titanium) and 9510066 (nylon)
- Variants
- The connector assemblies are separate items; this page covers the ferrule alone, Rosemount part 9310100
- Application
- Gripping the 1 in sensor tube inside a compression process connector, so the joint holds pressure and stays sealed
- Seal type
- Compression ferrule - the ring deforms onto the tube on first make-up, so it is replaced rather than re-used
- Dimensions
- 1 in compression ferrule
What a ferrule does
A 1 in compression ferrule — the metal ring that a compression connector deforms onto the sensor tube. When the nut is tightened the ferrule collapses slightly and grips the tube, and that grip is what holds the sensor against pressure and keeps the seal at the joint.
Why it is replaced, not re-used
A ferrule takes its shape once. After the first make-up it is matched to that tube at that depth; loosening the nut and re-tightening it deforms the ring a second time and the grip is never as good. On a pH or ORP installation, where the sensor comes out for cleaning, that is the difference between a dry connector and a slow weep that reaches the cable entry. Replace the ferrule whenever the joint is broken.
The failure it causes
A scored or over-compressed ferrule does not usually fail outright — it seeps. The symptom appears elsewhere: a wet connector body, a drifting pH reading, or a reference that fails its calibration. By the time it is diagnosed, the ferrule has been re-used three times. Keeping a spare is cheaper than the diagnostic time.
Where it fits
The 1 in compression process connectors used with Rosemount pH/ORP sensors — 23166-00 in 316 stainless steel, 23166-01 in titanium and the nylon 9510066 — which are ordered as complete assemblies and this ferrule separately.
Manufacturer reference: Rosemount ferrule 9310100, 1 in, for compression process connectors.
Specifications
| Attribute | Value |
|---|---|
| Sensor connection | Fits the 1 in compression process connectors 23166-00 (316 SST), 23166-01 (titanium) and 9510066 (nylon) |
| Variants | The connector assemblies are separate items; this page covers the ferrule alone, Rosemount part 9310100 |
| Application | Gripping the 1 in sensor tube inside a compression process connector, so the joint holds pressure and stays sealed |
| Seal type | Compression ferrule - the ring deforms onto the tube on first make-up, so it is replaced rather than re-used |
| Dimensions | 1 in compression ferrule |
Documents
No documents are published for this product yet.
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer