Overview
External sensor O-ring for the Rosemount Clarity II turbidimeter, sealing the sensor where it enters the molded debubbler.
Key features
- External sensor O-ring for the Rosemount Clarity II turbidimeter
- Fits the molded debubbler assembly
- EPDM, matching the debubbler's specified wetted materials
- Replace whenever the sensor is removed for cleaning
- A bypassing seal shows up as low or drifting turbidity, not as a leak
- Rosemount part number 9550145
Key specifications
- Wetted materials
- EPDM, matching the wetted materials Emerson specifies for the debubbler assembly (ABS, EPDM, Delrin, polypropylene and nylon)
- Sensor compatibility
- Rosemount Clarity II turbidity sensors with the molded debubbler assembly
- Variants
- Rosemount part number 9550145, supplied as the O-ring alone
- Application
- Sealing a Clarity II turbidity sensor where it enters the molded debubbler and flow chamber
- Seal type
- External O-ring on the sensor body
The seal between sensor and debubbler
An external O-ring for a Clarity II turbidity sensor, sized for the molded debubbler. It sits on the outside of the sensor where it enters the measuring chamber and keeps the sample where the optics expect it.
Why this O-ring matters more than its price
A turbidimeter measures the light scattered by particles in a fixed volume. If the seal at the sensor lets sample bypass the chamber, or lets air creep in, the instrument does not report a leak — it reports less turbidity, which looks like clean water. A hardened O-ring that has been compressed through a few services is the usual cause, and it is invisible from the outside until the reading is compared with a grab sample.
When to replace it
Every time the sensor is removed from the debubbler for cleaning. An O-ring that has taken a set will re-seal at first and weep later; on a turbidity measurement that shows up as drift, not as a puddle. Keeping spares is the cheap part of maintaining a turbidimeter, because the alternative is chasing an optical fault that is actually a seal.
Material
EPDM, matching the wetted materials Emerson specifies for the debubbler and flow chamber assembly (ABS, EPDM, Delrin, polypropylene and nylon). That is the right elastomer for potable and wastewater service, and the wrong one for hydrocarbon streams — which is not what a Clarity II is for.
Where it fits
Rosemount Clarity II turbidimeter installations with the molded debubbler assembly, in drinking water, wastewater effluent and process water monitoring.
Manufacturer reference: Rosemount T1056 Clarity II Turbidimeter Product Data Sheet, debubbler and flow chamber specifications; O-ring part number 9550145.
Specifications
| Attribute | Value |
|---|---|
| Wetted materials | EPDM, matching the wetted materials Emerson specifies for the debubbler assembly (ABS, EPDM, Delrin, polypropylene and nylon) |
| Sensor compatibility | Rosemount Clarity II turbidity sensors with the molded debubbler assembly |
| Variants | Rosemount part number 9550145, supplied as the O-ring alone |
| Application | Sealing a Clarity II turbidity sensor where it enters the molded debubbler and flow chamber |
| Seal type | External O-ring on the sensor body |