Rosemount Dissolved Oxygen Sensor Fill Solution
No product images are published for this product yet.
Purchase options
$48.28
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Fill solution #1 for amperometric dissolved oxygen sensors, 4 oz (125 ml): 7.5% KCl with a silver chloride trace that fixes the reference potential.
Key features
- Amperometric dissolved oxygen sensor fill solution #1, 4 oz (125 ml) bottle
- Composition: 92.5% water, 7.5% potassium chloride, 0.0005% silver chloride
- The silver chloride trace sets the equilibrium at the sensor's silver anode and fixes the reference potential
- Replace with the membrane and anode service, and before calibration on a sensor stored dry, not after it fails
- Not classified as a health hazard under GHS and not flammable, but carries H402 harmful to aquatic life for disposal
- Keep closed, upright and unfrozen - a bottle that has frozen and thawed has separated
Key specifications
- Accuracy
- Not a measured property: the solution sets the sensor's reference potential, and its chloride activity and conductivity decide whether a calibration will hold rather than any accuracy figure of its own
- Output signal
- Not applicable; the solution is the electrolyte between the membrane and the electrodes, and it carries the ionic current the sensor measures
- Diagnostics
- Not classified as a health hazard under GHS and not flammable, but carries H402 harmful to aquatic life: spent solution is a waste stream and does not go down a drain
- Pressure range
- Not applicable - electrolyte supplied in a 4 oz (125 ml) bottle for sensor service, with no process connection and no pressure rating
- Variants
- Fill solution #1, 4 oz (125 ml) bottle. Composition 92.4995% water, 7.5% potassium chloride and 0.0005% silver chloride
- Application
- Fill solution for Rosemount amperometric dissolved oxygen sensors in aeration basins, drinking water plants and industrial effluent monitoring, where the DO reading drives blower control
An electrolyte, not a consumable to substitute
This is the fill solution for Rosemount amperometric dissolved oxygen sensors: a 4 oz (125 ml) bottle of solution #1. It reads like a trivial item on a purchase order, and it is the part of the measurement that most often gets treated carelessly — refilled with whatever is on the shelf, topped up with distilled water, or left in a sensor that has been dry for a year.
What is actually in the bottle
The published composition is 92.5 percent water, 7.5 percent potassium chloride and 0.0005 percent silver chloride. That last figure is the one that matters and the one that looks like a rounding error. The silver chloride is there because the sensor's anode is silver: the chloride content of the electrolyte sets the equilibrium at that electrode, and therefore the reference potential the whole measurement is built on. A solution made up with plain salt water will run — for a while — and then drift, because the electrode chemistry it is sitting against has changed.
Why the concentration matters more than the volume
An amperometric oxygen sensor works by reducing oxygen at a cathode and balancing the reaction at a silver anode. The current is proportional to oxygen flux through the membrane, and the reference potential fixes the driving voltage. Change the chloride activity and the reference shifts; change the water content and the conductivity changes with it. Both show up as a calibration that will not hold rather than as an obvious failure, which is why a sensor with a mystery drift is often a sensor that was refilled with the wrong thing.
When to replace it
The electrolyte is consumed slowly and contaminated faster. Replacing it belongs with the membrane and the anode service — a sensor that has been opened for a membrane change should get fresh solution in the same visit, because the old solution has been in contact with the sample side and has had its chance to absorb what the process contains. Sensors that have been stored dry, or stored with the old solution evaporating, need it before they are calibrated, not after they fail the calibration.
Handling and disposal
The solution is not classified as a health hazard under GHS: the MSDS notes it is not expected to present a significant hazard under normal use, and it is not flammable. It does carry H402, harmful to aquatic life, which is the classification that decides how the used solution is disposed of — it should not go down a drain, and the spent solution from a sensor service is a small but real waste stream. Eye contact calls for rinsing for at least 15 minutes; skin contact for drenching the area.
Storage
Keep the bottle closed and upright, away from freezing — the solution is water based, and a bottle that has frozen and thawed has separated and should not be trusted. A stock of small bottles rotated on use is better practice than one large bottle opened repeatedly, since every opening is an opportunity to dilute or contaminate what is left.
Where it fits
Routine maintenance of Rosemount amperometric dissolved oxygen sensors in water and wastewater aeration basins, drinking water plants and industrial effluent monitoring — the same installations where the DO reading is used for blower control, and where a drifting sensor costs energy rather than just a bad number on a screen.
Manufacturer reference: Rosemount Analytical amperometric fill solution #1 safety data sheet (062116) for composition and handling, and product data sheet 00813-0100-3491 for the sensor family it serves.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Not a measured property: the solution sets the sensor's reference potential, and its chloride activity and conductivity decide whether a calibration will hold rather than any accuracy figure of its own |
| Output signal | Not applicable; the solution is the electrolyte between the membrane and the electrodes, and it carries the ionic current the sensor measures |
| Diagnostics | Not classified as a health hazard under GHS and not flammable, but carries H402 harmful to aquatic life: spent solution is a waste stream and does not go down a drain |
| Pressure range | Not applicable - electrolyte supplied in a 4 oz (125 ml) bottle for sensor service, with no process connection and no pressure rating |
| Variants | Fill solution #1, 4 oz (125 ml) bottle. Composition 92.4995% water, 7.5% potassium chloride and 0.0005% silver chloride |
| Application | Fill solution for Rosemount amperometric dissolved oxygen sensors in aeration basins, drinking water plants and industrial effluent monitoring, where the DO reading drives blower control |
| Mounting | Poured into the sensor body during membrane and anode service; a sensor opened for a membrane change should receive fresh solution in the same visit |
| Operating temperature (°C) | Water based and must not be frozen: a bottle that has frozen and thawed has separated and should not be trusted. Store closed and upright |