Overview
Amperometric steam-sterilizable dissolved oxygen sensor, 0 to 20 ppm, 0 to 130 C, holding calibration for up to 50 sterilization cycles.
Key features
- Amperometric steam-sterilizable dissolved oxygen sensor, model Hx438-01, 120 mm
- Drift-free signal with calibration held for up to 50 steam sterilization cycles
- Range 0 to 20 ppm or 0 to 250% saturation depending on the transmitter
- Operating range 0 to 130 C (32 to 266 F) at up to 43 psig (4 bar absolute)
- Stainless steel and EPDM wetted materials, PG 13.5 process thread, 4-pin cable connector
- Compatible with Rosemount transmitters 56, 1056, 5081 and 1066; insertion assembly 9160478 for SIP mounting
Key specifications
- Accuracy
- Drift-free signal that maintains its calibration value for up to 50 steam sterilization cycles
- Diagnostics
- Serviceable membrane cartridge, electrolyte and cathode: 9160487 service kit (three FDA-approved membranes, O-rings, polishing tool, 20 ml electrolyte) and 24262-00 cathode cartridge
- Pressure range
- 43 psig (400 kPa abs, 4 bar) maximum
- Measurement range
- 0 to 20 ppm or 0 to 250% saturation, depending on the transmitter; 22K NTC temperature compensation built into the sensor
- Sensor connection
- PG 13.5 process thread and a standard 4-pin cable connector; cable sold separately
- Application
- Amperometric dissolved oxygen measurement in steam-in-place and autoclave batch processes: biotechnology, pharmaceutical, food and beverage, and chemical vessels requiring sterilization between batches
A dissolved oxygen sensor built to survive sterilization
The Rosemount Hx438 is an amperometric dissolved oxygen sensor designed around steam-in-place and autoclave cycles — the conditions that destroy an ordinary DO probe. Its spec sheet is unusually specific about what that means: a drift-free signal, with the calibration value held for up to 50 steam sterilization cycles.
Why sterilization is the hard part
A DO sensor measures a tiny current produced by oxygen diffusing through a membrane onto a cathode. Heat changes the membrane's permeability, the electrolyte's conductivity and the electrode's surface, all at once. After a 130 °C cycle the sensor either returns to its previous slope or it does not, and the only useful question is how many cycles it survives before the calibration drifts out. Fifty cycles is the number Emerson publishes; it is also the number a validation protocol will be written against, so it belongs on the datasheet rather than in a footnote.
The measurement itself
Range is 0 to 20 ppm or 0 to 250% saturation depending on the transmitter, with 22K NTC temperature compensation built into the sensor. Wetted materials are stainless steel and EPDM, connected through a PG 13.5 thread, and the electrical interface is a standard 4-pin plug so the cable disconnects without disturbing the process connection. Maximum pressure is 43 psig (4 bar absolute) and the operating range runs from 0 to 130 °C.
Mounting two ways
The 120 mm sensor threads directly into a top plate for small batch vessels, or into the insertion mounting assembly (9160478, sold separately) for large-scale SIP installations where the sensor must be retractable — the assembly gives a G 1¼ in. mechanical connection, 70 mm insertion depth and a 16 µin Ra surface finish. The transmitter is whichever Rosemount unit the loop already uses: 56, 1056, 5081 or 1066.
Consumables are part of the design
The membrane cartridge, electrolyte and cathode are serviceable parts, and the quick start guide documents replacing each in place rather than returning the sensor: the 9160487 service kit carries three FDA-approved membranes with electrolyte and polishing materials, and 24262-00 is the cathode cartridge for when the electrode itself has aged. A sensor with a documented service path outlives one that has to be replaced at every drift event.
Where it fits
Biotechnology and pharmaceutical fermentation, food and beverage batch vessels, and chemical processes where the DO probe is cleaned and sterilized in place between batches. For clean-in-place duty Emerson directs you to the factory, which is a real limitation worth knowing before specifying.
Manufacturer reference: Rosemount Hx438 Amperometric Steam Sterilizable Dissolved Oxygen Sensors Product Data Sheet, LIQ-PDS-Hx438, June 2017, and Quick Start Guide 00825-0100-3494 Rev AA.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | Drift-free signal that maintains its calibration value for up to 50 steam sterilization cycles |
| Diagnostics | Serviceable membrane cartridge, electrolyte and cathode: 9160487 service kit (three FDA-approved membranes, O-rings, polishing tool, 20 ml electrolyte) and 24262-00 cathode cartridge |
| Pressure range | 43 psig (400 kPa abs, 4 bar) maximum |
| Measurement range | 0 to 20 ppm or 0 to 250% saturation, depending on the transmitter; 22K NTC temperature compensation built into the sensor |
| Sensor connection | PG 13.5 process thread and a standard 4-pin cable connector; cable sold separately |
| Application | Amperometric dissolved oxygen measurement in steam-in-place and autoclave batch processes: biotechnology, pharmaceutical, food and beverage, and chemical vessels requiring sterilization between batches |
| Mounting | Threaded directly into a top plate for small batch vessels, or fitted in the insertion mounting assembly for G 1¼ in. sockets with 70 mm insertion depth; sensor length 120 mm (Hx438-01) |
| Transmitter compatibility | Rosemount transmitters 56, 1056, 5081 and 1066 |
| Material of construction | Wetted materials stainless steel and EPDM; insertion mounting assembly (9160478) in 316L SS/DIN 1.4435 with EPDM and a surface finish below 16 micro inch Ra |
| Operating temperature (°C) | 0 to 130 °C (32 to 266 °F) |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.