Rosemount Model 142 Conductivity Sensor
Purchase options
From$363.37
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Contacting conductivity sensor for low-conductivity water in hot, pressurised lines, with 0.01/cm or 0.1/cm cell constants and a 3/4 in MPT connection.
Key features
- Contacting conductivity sensor for low-conductivity water
- 0.01/cm or 0.1/cm cell constants
- 316 SST electrodes and body, PEEK high-temperature body option
- 3/4 in MPT process connection
- Pt-100 temperature element available, cast aluminium junction box
- Built for hot, pressurised lines such as condensate and demineralised water
Key specifications
- Measurement principle
- Contacting (electrode) conductivity; the small cell constant suits water where the liquid itself is a poor conductor
- Process temperature
- Configuration dependent, plotted in the 140/141/142 pressure and temperature graph; the PEEK body extends the high-temperature limit
- Pressure rating
- Configuration dependent, from the same pressure and temperature graph
- Process connection
- 3/4 in MPT
- Wetted materials
- 316 SST electrodes and body; PEEK body with the high-temperature option; PCTFE insulator with the low-temperature option; Viton O-rings
- Measurement range
- Low-conductivity ranges through a 0.01/cm or 0.1/cm cell constant
What the 142 is
A contacting conductivity sensor for the other end of the scale: low-conductivity water, where the cell constants are small and the measurement is about purity rather than concentration. It is built for lines that are simultaneously hot and pressurised — demineralised water, condensate and boiler-quality streams — and it screws into the process with a 3/4 in MPT connection.
Why low conductivity needs a different cell
At very low conductivity the water itself is a poor conductor, so the cell has to present a large electrode area relative to its spacing: that is what a 0.01/cm cell constant does. Using a general-purpose cell in this range gives a reading dominated by cable capacitance and noise rather than by the water.
Construction for hot, pressurised lines
316 stainless electrodes and body as standard, with a PEEK body for the high-temperature option and a PCTFE insulator for the low-temperature option; Viton O-rings seal the assembly. The pressure and temperature limits are plotted per configuration in the data sheet rather than reduced to one figure, because the two options trade one limit against the other.
Installation
Screws into a 3/4 in MPT fitting, with a cast-aluminium junction box on the sensor and a Pt-100 temperature element available. The interconnecting cable between that box and the transmitter is ordered separately, at the length the installation needs.
Where it fits
Demineralised and deionised water, condensate monitoring, boiler feedwater quality and other low-conductivity streams in power and process plants.
Manufacturer reference: Emerson 140/141/142 contacting conductivity sensors product data sheet.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | Contacting (electrode) conductivity; the small cell constant suits water where the liquid itself is a poor conductor |
| Process temperature | Configuration dependent, plotted in the 140/141/142 pressure and temperature graph; the PEEK body extends the high-temperature limit |
| Pressure rating | Configuration dependent, from the same pressure and temperature graph |
| Process connection | 3/4 in MPT |
| Wetted materials | 316 SST electrodes and body; PEEK body with the high-temperature option; PCTFE insulator with the low-temperature option; Viton O-rings |
| Measurement range | Low-conductivity ranges through a 0.01/cm or 0.1/cm cell constant |
| Cable connection | Cast-aluminium junction box on the sensor, Pt-100 element available (code 54); interconnecting cable to the transmitter is ordered separately |
| Application | Demineralised and deionised water, condensate monitoring and boiler feedwater quality in power and process plants |
| Cell constant | 0.01/cm or 0.1/cm |
| Vacuum | At 1.6 in Hg (5.2 kPa) air leakage is below 0.005 SCFM (0.00014 m3/min) |