Overview
Submersion or insertion toroidal conductivity sensor for dirty, corrosive and high-conductivity liquids, measuring up to 2 S/cm in a PEEK or Tefzel body.
Key features
- Toroidal (non-contacting) conductivity sensor for dirty, corrosive and high-conductivity liquids
- Up to 2 S/cm (2,000,000 uS/cm)
- Submersion or insertion mounting
- 5/8 in 11 UNC or 3/4 in MNPT process connection
- PEEK or Tefzel body options, rated 120 C or 200 C at 295 psig
- 20 ft cable standard, 200 ft maximum
Key specifications
- Measurement principle
- Toroidal (inductive), non-contacting; reads through fouling that would blind an electrode cell
- Process temperature
- 248 F (120 C) with the -02 body; 392 F (200 C) with the -03 body
- Pressure rating
- 295 psig (2,135 kPa abs); CRN registration limits it to 220 psig (1,618 kPa abs)
- Process connection
- 5/8 in 11 UNC (option -20) or 3/4 in MNPT (option -21)
- Wetted materials
- Glass-filled PEEK, glass-filled Tefzel or unfilled Tefzel body; EPDM gasket with option -20
- Measurement range
- Up to 2 S/cm (2,000,000 uS/cm)
What the 228 is
A submersion or insertion toroidal conductivity sensor for the streams that destroy electrode cells: dirty liquids, corrosive chemistry and conductivities so high that a contacting sensor polarises. The measuring coils sit inside a moulded body with no exposed metal, so the sensor reads through a layer of fouling that would blind a contacting cell and does not care whether the liquid is moving.
What it measures and where
Conductivity up to 2 S/cm (2,000,000 µS/cm) — the range where acid, base and salt concentrations are read directly rather than inferred from a dilute sample. That is why the usual applications are concentration control on chemical streams, caustic and acid dilution, and brine strength, in lines that also happen to be dirty or corrosive.
Mounting
Submersion into a tank or channel, or insertion through a fitting with the 5/8 in 11 UNC or 3/4 in MNPT connection. It also works with the insertion mounting adapters and with retraction assemblies when a sensor has to come out without shutting the line down.
Materials and limits
Body in glass-filled PEEK, glass-filled Tefzel or unfilled Tefzel depending on the chemical duty, with an EPDM gasket on option -20. The -02 body is rated 248 F (120 C) and the -03 body 392 F (200 C), both at 295 psig; CRN registration limits that to 220 psig. Cable is 20 ft as standard, extendable to 200 ft.
Where it fits
Chemical concentration control, caustic and acid service, brine, pulp and any high-conductivity stream where the sensor has to survive the process rather than be protected from it.
Manufacturer reference: Emerson product data sheet LIQ-PDS-228.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | Toroidal (inductive), non-contacting; reads through fouling that would blind an electrode cell |
| Process temperature | 248 F (120 C) with the -02 body; 392 F (200 C) with the -03 body |
| Pressure rating | 295 psig (2,135 kPa abs); CRN registration limits it to 220 psig (1,618 kPa abs) |
| Process connection | 5/8 in 11 UNC (option -20) or 3/4 in MNPT (option -21) |
| Wetted materials | Glass-filled PEEK, glass-filled Tefzel or unfilled Tefzel body; EPDM gasket with option -20 |
| Measurement range | Up to 2 S/cm (2,000,000 uS/cm) |
| Cable connection | 20 ft (6.1 m) standard; 200 ft (61 m) maximum |
| Variants | Body options -02 (120 C) and -03 (200 C); option -54 valve insertion variant |
| Application | Acid, base and salt solution concentration, brine, caustic and pulp streams |
| Mounting | Submersion into a tank or channel, or insertion through a fitting; compatible with insertion mounting adapters and retraction assemblies |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.