Rosemount Model 3400HT pH / ORP Sensor
Purchase options
From$401.82
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Retractable titanium-body pH/ORP sensor for high temperature and pressure service, pH 0-14 and ORP +/-1500 mV, insertable through a ball valve.
Key features
- Retractable pH/ORP sensor in a titanium tube, inserts through a ball valve
- pH 0-14 and ORP -1500 to +1500 mV
- 5 to 155 C without preamplifier, up to 145 C with it
- 400 psig in service; retraction limited to 64 psig (21 in) or 35 psig (36 in)
- Double-junction reference with replaceable electrolyte and Teflon junction
- 15 ft cable, 9.5 in junction-box lead, or VP8 connector
Key specifications
- Process temperature
- 41 to 311 F (5 to 155 C) without a preamplifier; up to 293 F (145 C) with the preamplifier fitted
- Pressure rating
- 400 psig (2,859 kPa abs) maximum in service; CRN rating 200 psig at room temperature
- Process connection
- Requires a 1 in compression process connector (23166-00 or 23166-01); can be inserted through a ball valve
- Wetted materials
- Titanium body, Ryton, Teflon and glass, with EPDM, Viton or Kalrez O-rings to order
- Measurement range
- pH 0-14 and ORP -1500 to +1500 mV
- Reference
- Double junction with a replaceable process-side electrolyte and a Teflon junction
What the 3400HT is
The retractable member of the PERpH-X titanium family: a pH/ORP sensor long enough to be inserted through a ball valve and withdrawn again without depressurising the line. Where the 3300HT is the sensor you install and leave, the 3400HT is the one you can pull out for cleaning while the process keeps running.
Retraction is a pressure question, not just a length
The body is rated 400 psig in service, but the pressure you may retract or insert it against is lower and depends on the length: 64 psig with the 21 in body and 35 psig with the 36 in body. Those two numbers are the ones to check before planning a retraction under pressure — quoting the 400 psig as if it applied to the movement is the mistake this page exists to prevent.
Temperature, with and without the preamplifier
5 to 155 C (41 to 311 F) without a preamplifier, and up to 145 C with one fitted — a much smaller penalty than the 3300HT's 100 C ceiling, which is a real reason to choose the 3400 for a hot line that also needs SMART electronics.
What it measures with
pH 0-14 and ORP -1500 to +1500 mV, a double-junction reference with replaceable process-side electrolyte and Teflon junction, and a Pt-100 RTD for compensation. Wetted parts are titanium, Ryton, Teflon and glass, with EPDM, Viton or Kalrez O-rings to order.
Installation
Seats in a 1 in compression process connector (23166-00 or -01) and can be inserted through a ball valve. Cable is 15 ft integral, or the 9.5 in integral lead for a sensor-head junction box; the 3400HTVP uses the VP8 connector with the cable ordered separately.
Where it fits
Hot, pressurised pH/ORP points that must stay serviceable: chemical reactors, pulp and paper lines, and power plant water chemistry.
Manufacturer reference: Emerson PERpH-X 3300HT/3400HT/3500P product data sheet.
Specifications
| Attribute | Value |
|---|---|
| Process temperature | 41 to 311 F (5 to 155 C) without a preamplifier; up to 293 F (145 C) with the preamplifier fitted |
| Pressure rating | 400 psig (2,859 kPa abs) maximum in service; CRN rating 200 psig at room temperature |
| Process connection | Requires a 1 in compression process connector (23166-00 or 23166-01); can be inserted through a ball valve |
| Wetted materials | Titanium body, Ryton, Teflon and glass, with EPDM, Viton or Kalrez O-rings to order |
| Measurement range | pH 0-14 and ORP -1500 to +1500 mV |
| Reference | Double junction with a replaceable process-side electrolyte and a Teflon junction |
| Cable connection | 15 ft integral cable, or the 9.5 in integral lead for a sensor-head junction box on the 3400HT; VP8 connector on the 3400HTVP with the cable sold separately |
| Application | Hot, pressurised pH/ORP points that must stay serviceable: chemical reactors, pulp and paper lines and power plant water chemistry |
| Mounting | Retractable through a ball valve; maximum pressure at retraction or insertion is 64 psig (542 kPa) at 21 in length and 35 psig (343 kPa) at 36 in length |
| Temperature sensor | Pt-100 resistance temperature device (RTD) for compensation |