Rosemount Model 3500 pH/ORP Sensor
No product images are published for this product yet.
Purchase options
From$412.14
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
pH/ORP sensor in a Ryton body with 1 in MNPT threads, pH 0-14 and ORP +/-1500 mV, rated 100 psig and 120 C for water and light chemical duty.
Key features
- pH/ORP sensor in a Ryton plastic body with 1 in MNPT front and rear threads
- pH 0-14 and ORP -1500 to +1500 mV
- 5 to 120 C (41 to 248 F) and 100 psig maximum
- Double-junction reference with replaceable electrolyte and Teflon junction
- Pt-100 RTD for temperature compensation
- 25 ft or 15 ft integral cable, or the VP8 connector
Key specifications
- Process temperature
- 41 to 248 F (5 to 120 C)
- Pressure rating
- 100 psig (790 kPa abs) maximum; CRN rating 40 psig at room temperature
- Process connection
- 1 in MNPT threads, front and rear facing, for insertion through a tee or submersion mounting
- Wetted materials
- Titanium, Ryton, Teflon and glass, with EPDM, Viton or Kalrez O-rings to order
- Measurement range
- pH 0-14 and ORP -1500 to +1500 mV
- Reference
- Double junction with a replaceable process-side electrolyte and a Teflon junction
What the 3500 is
A pH/ORP sensor in a Ryton plastic body with 1 in MNPT threads at both ends, which makes it the general-purpose member of the PERpH-X family: screw it into a pipe tee or a tank wall, front thread or rear, and it measures. Where the titanium 3300HT and 3400HT are for hot pressurised lines, the 3500 is for the ordinary ones — and it is priced accordingly.
The range it covers
pH 0-14 and ORP -1500 to +1500 mV from 5 to 120 C (41 to 248 F), at up to 100 psig. The CRN rating is 40 psig at room temperature, so a Canadian registration review sees a lower ceiling than the datasheet headline — worth knowing before specifying it for a pressurised line there.
Reference and compensation
Double-junction reference with a replaceable process-side electrolyte and a Teflon junction: the same renewal path as the titanium models, so a poisoned reference is a service job rather than a scrapped sensor. A Pt-100 RTD handles temperature compensation.
Installation
Threads are 1 in MNPT front and rear, which covers insertion through a tee or submersion on a standpipe. Cable options are 25 ft integral (option -01) or 15 ft integral (option -02); the 3500VP uses the VP8 connector with the cable ordered separately.
Where it fits
Municipal and industrial water, wastewater, cooling towers, and light chemical duty where a plastic-bodied pH sensor is the right instrument and the process is not hot enough or pressurised enough to need titanium.
Manufacturer reference: Emerson PERpH-X 3300HT/3400HT/3500P product data sheet.
Specifications
| Attribute | Value |
|---|---|
| Process temperature | 41 to 248 F (5 to 120 C) |
| Pressure rating | 100 psig (790 kPa abs) maximum; CRN rating 40 psig at room temperature |
| Process connection | 1 in MNPT threads, front and rear facing, for insertion through a tee or submersion mounting |
| Wetted materials | Titanium, Ryton, Teflon and glass, with EPDM, Viton or Kalrez O-rings to order |
| Measurement range | pH 0-14 and ORP -1500 to +1500 mV |
| Reference | Double junction with a replaceable process-side electrolyte and a Teflon junction |
| Cable connection | 25 ft integral cable (option -01) or 15 ft integral cable (option -02); VP8 connector on the 3500VP with the cable sold separately |
| Application | Municipal and industrial water, wastewater, cooling towers and light chemical duty |
| Temperature sensor | Pt-100 resistance temperature device (RTD) for compensation |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer