Rosemount Model 396P / 396PVP TUpH pH/ORP Sensors
Purchase options
From$262.37
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Titanium-body TUpH pH/ORP sensor with 1 in MNPT threads front and rear, so it screws directly into a tee or fitting without a separate process connector.
Key features
- Titanium-body TUpH pH/ORP sensor with 1 in MNPT threads front and rear
- pH 0-14 or ORP -1500 to +1500 mV, with a platinum electrode on the ORP version
- Screws directly into a tee or fitting - no separate process connector needed
- Hemi or flat ACCUGLASS GPLR bulb; the flat bulb resists coating
- 0 to 100 °C; 150 psig with a hemi bulb, 100 psig with a flat bulb
- 25 ft 11-conductor or 15 ft coaxial integral cable, remote or integral SMART preamp
Key specifications
- Measurement principle
- TUpH glass electrode with an ACCUGLASS GPLR bulb, hemi or flat; platinum electrode on the ORP version
- Process temperature
- 0 to 100 °C (32 to 212 °F)
- Pressure rating
- 100 to 1135 kPa abs (0 to 150 psig) with a hemi bulb; 100 to 790 kPa abs (0 to 100 psig) with a flat bulb
- Process connection
- 1 in MNPT threads on both the front and the rear of the sensor, so it mounts without a separate process connector
- Wetted materials
- Titanium, polypropylene, EPDM and glass; platinum on the ORP version
- Measurement range
- pH 0-14, or ORP -1500 to +1500 mV; a stable pH reading needs at least 100 uS/cm in the sample
The titanium TUpH sensor that threads itself in
The 396P takes the TUpH electrode and puts it in a titanium body with 1 in MNPT threads on both the front and the rear. That is the difference that matters in the field: the stainless 396 needs a separate process connector to hold it, while the 396P screws straight into a tee, a submersion fitting or a standpipe — and the second thread at the back gives you a way to mount it from either end.
When titanium is the right body
Titanium is specified where stainless would pit or where the product must not see iron: chlorinated water, hypochlorite, many chemical streams, and food or pharmaceutical contact where a corroding fitting is a contamination risk rather than a maintenance item. The trade is cost. Where the process is kind to 316 stainless, the 396 does the same measurement for less, and the electrode chemistry is identical.
The bulb choice
ACCUGLASS GPLR in hemi or flat. The flat bulb is the coating answer: it sits nearly flush so the flow wipes it, where the hemispherical bulb would collect a deposit in a slurry and drift as the reference junction plugs behind it. In a clean stream the hemi bulb is the more accurate of the two.
Signal path
A pH electrode is a high-impedance source, so the preamplifier has to be somewhere: the family supports remote or integral SMART preamplifiers, and the cable terminates in either a 25 ft 11-conductor lead (code 01) or a 15 ft coaxial lead (code 02) depending on which arrangement the transmitter expects. Getting that choice wrong is a re-order, not a field fix.
Limits
0 to 100 °C, 150 psig with a hemi bulb and 100 psig with a flat bulb, and at least 100 µS/cm in the sample for a stable pH reading — in ultrapure water the measurement is unstable whatever the sensor. The ORP version uses a platinum electrode and covers -1500 to +1500 mV.
Where it fits
Chlorinated water and wastewater, chemical process lines, food and pharmaceutical contact points and any pH/ORP service where a titanium body and a direct 1 in thread are required — with 396PVP where a Variopol connector suits the installation.
Manufacturer reference: Rosemount 396/396VP/396P/396PVP/396R/396RVP TUpH pH/ORP Sensors Product Data Sheet, LIQ-PDS-396, table 7.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | TUpH glass electrode with an ACCUGLASS GPLR bulb, hemi or flat; platinum electrode on the ORP version |
| Process temperature | 0 to 100 °C (32 to 212 °F) |
| Pressure rating | 100 to 1135 kPa abs (0 to 150 psig) with a hemi bulb; 100 to 790 kPa abs (0 to 100 psig) with a flat bulb |
| Process connection | 1 in MNPT threads on both the front and the rear of the sensor, so it mounts without a separate process connector |
| Wetted materials | Titanium, polypropylene, EPDM and glass; platinum on the ORP version |
| Measurement range | pH 0-14, or ORP -1500 to +1500 mV; a stable pH reading needs at least 100 uS/cm in the sample |
| Cable connection | Integral cable: code 01 is 25 ft of 11-conductor cable, code 02 is 15 ft coaxial; the 396PVP uses a Variopol connector |
| Dimensions | 1 lb (0.45 kg) net, 2 lb (0.9 kg) shipping weight |