Rosemount Oxygen Test Gas Kit
No product images are published for this product yet.
Purchase options
$1,277.57
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Two-point oxygen calibration gas kit with 0.4% and 8% O2 cylinders for verifying in-situ oxygen analyzers on boilers and furnaces.
Key features
- Two oxygen calibration gases supplied as a kit: 0.4% and 8% O2
- Two-point calibration for in-situ oxygen analyzers, checking both offset and span
- Low point tests the bottom of the range where excess-air control is most sensitive
- Supplied in pressurised cylinders with a safety data sheet
- For commissioning and periodic calibration of combustion oxygen analyzers
Key specifications
- Certifications
- Supplied with a safety data sheet; cylinder handling per standard compressed gas practice with a back-flow prevention device and a regulator rated for cylinder pressure
- Measurement range
- Two calibration points, 0.4% and 8% oxygen, bracketing the working range of a combustion oxygen measurement
- Application
- Two-point calibration gas supply for in-situ oxygen analyzers on boilers, furnaces and process heaters
- Kit contents
- Two oxygen test gases: 0.4% O2 and 8% O2, supplied in pressurised cylinders
- Material of construction
- Pressurised gas cylinders with a safety data sheet supplied for the gas mixture
Two gas points for an oxygen analyzer
The Rosemount Oxygen Test Gas Kit supplies two calibration gases — 0.4% and 8% oxygen — for verifying an in-situ oxygen analyzer. It is the consumable half of a measurement that otherwise has no reference: an oxygen probe sits in a flue and reports a number, and the only way to know that number is right is to feed it a gas whose oxygen content is known.
Why two points, and why these two
A single-point check can tell you the analyzer responds to oxygen; it cannot tell you whether the response is linear. The pair is chosen to bracket the working range: the higher point sets the span near normal combustion oxygen levels, and the low point tests the bottom of the range where small absolute errors matter most. A probe that reads correctly at 8% and drifts at 0.4% is a probe that will hide a problem in exactly the region where excess-air control is most sensitive.
What a calibration actually proves
It does not prove the probe is undamaged — a cracked sensor can pass a two-point check and still fail in service. It proves the measurement chain, from probe to electronics to displayed value, is offset-free and has the right slope at the moment of the test. That is why the record of a calibration is worth keeping: the trend of the span correction across successive checks says more about probe health than any single result.
Handling
The gases ship in pressurised cylinders with a safety data sheet. Standard cylinder discipline applies: secure the cylinder, open the valve slowly, use a regulator rated for the cylinder pressure, use a back-flow prevention device, and close the valve after each use. Regulators, tubing and a calibration cup are supplied with the analyzer's calibration infrastructure rather than with the gas itself.
Where it fits
Commissioning and periodic calibration of in-situ oxygen analyzers on boilers, furnaces and process heaters, and any site where the combustion trim loop depends on an oxygen reading that has to be defensible at an audit.
Manufacturer reference: Rosemount Oxygen Test Gas Kit, supplied with safety data sheet.
Specifications
| Attribute | Value |
|---|---|
| Certifications | Supplied with a safety data sheet; cylinder handling per standard compressed gas practice with a back-flow prevention device and a regulator rated for cylinder pressure |
| Measurement range | Two calibration points, 0.4% and 8% oxygen, bracketing the working range of a combustion oxygen measurement |
| Application | Two-point calibration gas supply for in-situ oxygen analyzers on boilers, furnaces and process heaters |
| Kit contents | Two oxygen test gases: 0.4% O2 and 8% O2, supplied in pressurised cylinders |
| Material of construction | Pressurised gas cylinders with a safety data sheet supplied for the gas mixture |
Documents
No documents are published for this product yet.