Rosemount Oxymitter 4000 Hazardous Area Oxygen Transmitter
Purchase options
Request a quote
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
In-situ zirconia oxygen transmitter for flue gas in hazardous areas: ATEX Exd IIB + H2 T4 Gb, 0-10% to 0-40% O2, process to 800 C with Xi electronics.
Key features
- In-situ zirconia oxygen transmitter for flue gas in hazardous areas, model OXT4C
- ATEX II 2G Exd IIB + H2 T4 Gb flameproof housing with through-glass infrared buttons
- Net O2 range variable 0-10% to 0-40% (0-50% with Xi electronics); accuracy ±0.75% of reading or 0.05% O2
- Process temperature to 705 C (1300 F), or 800 C with Xi electronics; probe lengths 18 in to 18 ft
- Response to calibration gas under 3 seconds with T90 inside 8 seconds
- HART standard, with HART wireless through a THUM adaptor; electronics integral to the probe or remote
Key specifications
- Accuracy
- +/-0.75% of reading or 0.05% O2, whichever is greater, in oxidizing conditions; +/-0.10% of reading or 0.1% O2 in reducing conditions; process temperature effect below 0.05% O2 from 100 to 700 C
- Output signal
- 4-20 mA with HART digital communication; HART wireless available through a THUM adaptor; optional Xi electronics add a large backlit LCD and advanced software features
- Diagnostics
- Heater and thermocouple hold the sensing cell at a 736 C setpoint within +/-1 C; the probe is field repairable with access to internal components, and the transmitter has no potentiometers or jumpers to set
- Certifications
- ATEX II 2G Exd IIB + H2 T4 Gb for the hazardous area version (model OXT4C), in a Type 4X (IP65/IP66) enclosure
- Measurement range
- Net O2 range variable from 0-10% to 0-40%; the optional Xi electronics extend it to 0-50% O2; lowest detectable limit 0.02% O2
- Application
- In-situ oxygen measurement in flue gas of boilers, process heaters, furnaces and kilns installed in hazardous (classified) areas, where the O2 reading drives combustion trim
The same zirconia probe, built for the area classification
The Oxymitter 4000 hazardous area version is the in-situ oxygen transmitter for flue gas in classified areas: a zirconia sensing cell at the tip, a heated probe that holds the cell at 736 °C, and electronics mounted either on the probe or remotely — all inside a housing certified ATEX II 2G Exd IIB + H2 T4 Gb, with through-glass infrared buttons so nobody has to open it in a hazardous area.
What the certification actually changes
The measurement is unchanged from a standard Oxymitter: net oxygen range variable from 0-10% to 0-40%, accuracy ±0.75% of reading or 0.05% O2 (whichever is greater), and a lowest detectable limit of 0.02% O2. What the Exd classification changes is the housing and the interface — a flameproof enclosure, a reference air exhaust port piped to a clean area, and controls operated through the glass rather than by opening a cover. Those are the details an area classification review asks about, and they are why this version exists as a separate model.
Why in-situ, in one paragraph
A sampling system pulls flue gas through a line to an analyser, and the line is where the problems live: it plugs, it leaks, it cools and condenses, and the reading lags the furnace by the time the gas arrives. An in-situ probe puts the zirconia cell directly in the duct, so there is no sample line to maintain and the response to calibration gas is under 3 seconds with a T90 inside 8 seconds. Emerson's own numbers show the cost of the alternative: response in reducing conditions is 120 seconds going in and 30 seconds coming out — the cell's chemistry, not the electronics.
Temperature is the operating envelope
Process temperature runs to 705 °C (1300 °F) with Oxymitter electronics and 800 °C (1472 °F) with the optional Xi electronics, which also extends the range to 0-50% O2 and adds a backlit display and a wireless path through a THUM adaptor. Probe lengths run from 18 inches to 18 feet, which is what lets the tip sit in the gas stream rather than in a dead pocket at the duct wall. Electronics ambient is -40 to 80 °C, and the process temperature effect on the reading is under 0.05% O2 across 100 to 700 °C.
Repairability
The probe is field-repairable: the internal components are accessible without returning the assembly, the heater and thermocouple hold the cell setpoint within ±1 °C, and the transmitter has no potentiometers or jumpers to set. On a boiler that is the difference between a maintenance task and an outage.
Where it fits
Boilers, process heaters, furnaces and kilns in classified areas — refinery and petrochemical fired equipment, and any flue where the oxygen reading drives combustion trim and the area is Zone 1.
Manufacturer reference: Rosemount Oxymitter 4000 Hazardous Area In-Situ Oxygen Transmitter Product Data Sheet, September 2023.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | +/-0.75% of reading or 0.05% O2, whichever is greater, in oxidizing conditions; +/-0.10% of reading or 0.1% O2 in reducing conditions; process temperature effect below 0.05% O2 from 100 to 700 C |
| Output signal | 4-20 mA with HART digital communication; HART wireless available through a THUM adaptor; optional Xi electronics add a large backlit LCD and advanced software features |
| Diagnostics | Heater and thermocouple hold the sensing cell at a 736 C setpoint within +/-1 C; the probe is field repairable with access to internal components, and the transmitter has no potentiometers or jumpers to set |
| Certifications | ATEX II 2G Exd IIB + H2 T4 Gb for the hazardous area version (model OXT4C), in a Type 4X (IP65/IP66) enclosure |
| Measurement range | Net O2 range variable from 0-10% to 0-40%; the optional Xi electronics extend it to 0-50% O2; lowest detectable limit 0.02% O2 |
| Application | In-situ oxygen measurement in flue gas of boilers, process heaters, furnaces and kilns installed in hazardous (classified) areas, where the O2 reading drives combustion trim |
| Mounting | Probe inserted through the duct wall on its mounting flange, with a heat sink between the process and the electronics; reference air exhaust port piped to a clean area; through-glass infrared push buttons suit hazardous areas |
| Material of construction | Process wetted materials 316L or 304 stainless steel; low copper aluminum transmitter housing rated Type 4X (IP65/IP66); probe lengths from 18 in. to 18 ft, with the electronics integral to the probe or remote mounted |
| Operating temperature (°C) | Process temperature 0 to 705 C (32 to 1300 F) with Oxymitter electronics and 0 to 800 C (32 to 1472 F) with Xi electronics; electronics ambient -40 to 80 C (-40 to 176 F), or -20 to 50 C for the Xi housing |