Rosemount R508 ORP Standard
No product images are published for this product yet.
Purchase options
$36.82
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
475 mV ORP test solution in an 8 oz bottle (Rosemount R-508), used to verify an ORP sensor and its reference junction, with NIST-traceable certification.
Key features
- 475 mV ORP (redox) test solution
- 8 oz bottle, Rosemount part number R-508
- Used to verify an ORP sensor and its reference junction
- NIST-traceable calibration per ANSI/NCSL Z540-1-1994
- GHS-US: skin and eye irritant, category 1A carcinogen - gloves and eye protection required
- Supplied with the safety data sheet and certification of compliance
Key specifications
- Certifications
- NIST-traceable calibration meeting ANSI/NCSL Z540-1-1994, under an ISO 9001:2008 quality system certified by DNV-GL
- Measurement range
- 475 mV ORP test solution, the value certified by the manufacturer for this standard
- Sensor compatibility
- ORP sensors and their transmitters; pH sensors are verified against pH buffers, not against this redox standard
- Application
- Verifying the response of an ORP (redox) sensor and the condition of its reference junction
- Volume
- 8 oz bottle, Rosemount part number R-508
- Storage
- Store locked up, out of reach of children and away from food, per the safety data sheet supplied
What this standard is
A 475 mV ORP (redox) test solution in an 8 oz bottle, part R-508. It is the reference against which an ORP sensor is checked: dip the electrode pair, let the reading settle, and the number on the transmitter is compared with the 475 mV the solution is certified to produce.
Why ORP needs a standard at all
An ORP sensor is a noble-metal electrode read against a reference, and its reading is always relative to that reference — which is why an ORP value means nothing without naming the reference electrode it was measured against. What a standard does is remove the guesswork from the sensor's condition: if the electrode reads 475 mV in this solution, the metal surface and the reference junction are both working. If it reads low and drifts, the problem is usually the junction, not the transmitter.
It is not a pH buffer
The two are checked with different liquids and mean different things. A pH buffer fixes hydrogen ion activity; this solution fixes redox potential. Using one to check the other produces a confident, wrong calibration, which is why the two standards are stocked and labelled separately.
Handling
The safety data sheet attached classifies the solution under GHS-US as a skin irritant (H315), a serious eye irritant (H319) and a category 1A carcinogen (H350), with the signal word Danger. Wear gloves and eye protection, keep it locked up and away from food, and dispose of it as the SDS requires — this is one of the few liquids in the range that genuinely warrants the warning on the label.
Traceability
Supplied with Rosemount's certification of compliance: the calibration system meets ANSI/NCSL Z540-1-1994 and is traceable to NIST, under an ISO 9001:2008 quality system certified by DNV-GL. That certificate is in the same file, which matters when an auditor asks what the 475 mV was checked against.
Manufacturer reference: Rosemount Analytical ORP Test Solution, part number R-508, with safety data sheet and certification of compliance.
Specifications
| Attribute | Value |
|---|---|
| Certifications | NIST-traceable calibration meeting ANSI/NCSL Z540-1-1994, under an ISO 9001:2008 quality system certified by DNV-GL |
| Measurement range | 475 mV ORP test solution, the value certified by the manufacturer for this standard |
| Sensor compatibility | ORP sensors and their transmitters; pH sensors are verified against pH buffers, not against this redox standard |
| Application | Verifying the response of an ORP (redox) sensor and the condition of its reference junction |
| Volume | 8 oz bottle, Rosemount part number R-508 |
| Storage | Store locked up, out of reach of children and away from food, per the safety data sheet supplied |
| Hazard classification | GHS-US: Skin Irrit. 2 (H315), Eye Irrit. 2A (H319), Carc. 1A (H350); signal word Danger - wear gloves and eye protection and follow the safety data sheet |
Documents