Rosemount Refill Kits
No product images are published for this product yet.
Purchase options
From$57.61
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
30 cc syringe of reference electrolyte for refillable pH/ORP sensors, enough for four to five refills, supplied with the electrolyte safety data sheets.
Key features
- 30 cc syringe of reference electrolyte
- Four to five refills per syringe
- For refillable reference electrodes on PERpH-X sensors
- Standard and poisoning-resistant gel options
- Safety data sheets supplied for both electrolytes
- Restores a drifting reference instead of replacing the sensor
Key specifications
- Wetted materials
- Reference electrolyte supplied as a standard gel or a poisoning-resistant gel; the safety data sheet for each is attached
- Sensor compatibility
- Refillable reference electrodes on the PERpH-X families (3300HT, 3400HT, 3500)
- Variants
- Each syringe carries enough electrolyte for four to five refills
- Application
- Refilling the reference electrolyte of refillable pH/ORP sensors during routine maintenance
- Volume
- 30 cc syringe
- Storage
- Per the electrolyte safety data sheet supplied with the kit
What the refill kit is
A 30 cc syringe of reference electrolyte for refillable pH/ORP sensors — the consumable that keeps a replaceable-reference sensor alive. Each syringe carries enough electrolyte for four to five refills, which is roughly the working life of a reference between services in normal duty.
Why the reference needs refilling at all
A refillable reference electrode works by slowly bleeding electrolyte through a junction into the process; that is what keeps the junction clean and the potential stable. The electrolyte is consumed, so the reservoir empties and the reference drifts — not because the sensor has failed, but because it has been doing its job. Refilling restores the reference, and a sensor that reads erratically after months of service is often just dry.
Matching the electrolyte to the process
The electrolyte type has to suit the chemistry: a standard gel for ordinary service, and a poisoning-resistant gel where sulfides, mercaptans or cyanides are present. The safety data sheets for both gels are attached to this product, because what you are handling is a chemical, not just a spare part.
Where it fits
Routine maintenance on refillable PERpH-X sensors — the 3300HT, 3400HT and 3500 families — where the reference is designed to be renewed rather than replaced.
Manufacturer reference: Rosemount liquid analysis reference electrolyte refill kits and their safety data sheets.
Specifications
| Attribute | Value |
|---|---|
| Wetted materials | Reference electrolyte supplied as a standard gel or a poisoning-resistant gel; the safety data sheet for each is attached |
| Sensor compatibility | Refillable reference electrodes on the PERpH-X families (3300HT, 3400HT, 3500) |
| Variants | Each syringe carries enough electrolyte for four to five refills |
| Application | Refilling the reference electrolyte of refillable pH/ORP sensors during routine maintenance |
| Volume | 30 cc syringe |
| Storage | Per the electrolyte safety data sheet supplied with the kit |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.