Retractable TUpH pH/ORP sensor with a titanium body that is withdrawn through a ball valve for cleaning without shutting down the line.
Key features
- Retractable TUpH pH/ORP sensor in a titanium body
- pH 0-14 or ORP -1500 to +1500 mV
- Withdraws through a 1-1/4 or 1-1/2 in ball valve kit (sold separately) for cleaning at line pressure
- Hemi or flat ACCUGLASS bulb for abrasive service
- 0 to 100 °C; 150 psig static with a hemi bulb, 64 psig while retracting
- 15 ft coaxial cable standard, short 9.5 in pigtail options
Key specifications
- Measurement principle
- TUpH glass electrode with an ACCUGLASS GPLR bulb, hemi or flat; platinum electrode on the ORP version
- Process temperature
- 0 to 100 °C (32 to 212 °F)
- Pressure rating
- 100 to 1135 kPa abs (0 to 150 psig) with a hemi bulb; 100 to 790 kPa abs (0 to 100 psig) with a flat bulb; 64 psig (code 21) or 35 psig (code 25) while retracting or inserting
- Process connection
- The sensor has no process thread - it uses a 1 in. process connector or a 1-1/4 / 1-1/2 in. ball valve kit, both sold separately
- Wetted materials
- Titanium body, polypropylene, EPDM and glass; platinum (ORP only)
- Measurement range
- pH 0-14, or ORP -1500 to +1500 mV; a stable pH measurement needs at least 100 uS/cm conductance in the sample
A pH sensor built to be pulled out and cleaned
The 396R is the retractable member of the TUpH family: a glass pH or ORP electrode in a titanium tube that slides through a packing gland and a ball valve, so the sensor can be withdrawn from a live line for cleaning or calibration. It is the version to specify where the process coats the glass — pulp stock, slurries, dirty cooling water — and where a fixed sensor would be blind within a week.
What TUpH is for
The reference junction is the part of a pH sensor that fails first in a dirty process: it plugs, the reference potential drifts, and the reading drifts with it. TUpH sensors are built around a junction designed to resist that fouling, which is why they are specified for coating and high-solids service rather than clean water. The glass bulb comes as hemi or flat; the flat bulb is there for abrasive streams where a protruding hemisphere would be worn away.
Retraction, and what it costs
The sensor has no process thread of its own — it relies on a ball valve kit (1¼ or 1½ in, sold separately) and a retraction chamber. That is an extra purchase and an extra installation step, and it buys the ability to service the sensor at line pressure without a shutdown. The pressure limits tell you how much that costs: 150 psig with a hemi bulb and 100 psig with a flat bulb as a static sensor, but only 64 or 35 psig respectively while the sensor is moving through the gland. Retraction is a maintenance action, not an operating position.
Signal and cable
A pH electrode's signal is high impedance, so the cable is part of the measurement: coaxial, 15 ft standard, with short 9.5 in pigtail options (codes 60/61) for installations where the preamplifier sits close by. The minimum conductivity of 100 µS/cm is the other limit worth knowing — in ultrapure water the pH measurement becomes unstable regardless of the sensor.
Materials and service
Titanium body, polypropylene, EPDM and glass wetted parts, with platinum on the ORP version. Rated 0 to 100 °C (32 to 212 °F). Titanium is what makes the retractable body survive chemical and abrasive service where a stainless tube would pit.
Where it fits
Pulp and paper stock lines, chemical reactors with solids, dirty cooling water and any pH or ORP point where the electrode has to be cleaned on a schedule — with 396VP when a Variopol connection suits the installation.
Manufacturer reference: Rosemount 396/396VP/396P/396PVP/396R/396RVP TUpH pH/ORP Sensors Product Data Sheet, LIQ-PDS-396.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | TUpH glass electrode with an ACCUGLASS GPLR bulb, hemi or flat; platinum electrode on the ORP version |
| Process temperature | 0 to 100 °C (32 to 212 °F) |
| Pressure rating | 100 to 1135 kPa abs (0 to 150 psig) with a hemi bulb; 100 to 790 kPa abs (0 to 100 psig) with a flat bulb; 64 psig (code 21) or 35 psig (code 25) while retracting or inserting |
| Process connection | The sensor has no process thread - it uses a 1 in. process connector or a 1-1/4 / 1-1/2 in. ball valve kit, both sold separately |
| Wetted materials | Titanium body, polypropylene, EPDM and glass; platinum (ORP only) |
| Measurement range | pH 0-14, or ORP -1500 to +1500 mV; a stable pH measurement needs at least 100 uS/cm conductance in the sample |
| Cable connection | 15 ft coaxial cable standard; code 60 gives a 9.5 in. coaxial pigtail with BNC, code 61 a 9.5 in. coaxial pigtail |
| Dimensions | Sensor weight 2 lb (0.9 kg) for code 21 and 3 lb (1.4 kg) for code 25; shipping 3 lb / 4 lb |