Ir al contenido principal
Vendido y enviado por Sproh

Transmisor de caudal Rosemount 8782

Opciones de compra

Desde2044,16 US$

Disponibilidad
Consultar disponibilidad Original
Seleccionar opción

Configurable: elija las opciones para construir el número de pieza exacto.

Proveedor independiente

Sproh es un proveedor independiente y no es un distribuidor autorizado ni representante oficial de Rosemount o Emerson.

Resumen

Transmisor de caudal magnético de montaje en pared con Smart Meter Verification: mide resistencia de bobina, inductancia y resistencia de electrodo contra una línea base guardada.

Especificaciones principales

Exactitud
The transmitter carries the meter's accuracy class: 0.25 percent of rate standard and 0.15 percent of rate optional with a compatible sensor. Smart Meter Verification reports sensor calibration deviation against the stored baseline
Señal de salida
4-20 mA with HART, pulse output and discrete input/outputs, with intrinsically safe output options. Coil excitation circuit rated 2.0 A at up to 85 V and 80 W
Diagnósticos
Smart Meter Verification measures coil resistance, coil inductance and electrode resistance and compares them to the values taken during the sensor baseline process, reporting sensor calibration deviation plus coil circuit health and electrode circuit health. It compares electrical properties to a baseline, not to a flow standard - the 8785 Calibration Standard is the separate independent check - and a baseline taken on an already-coated sensor verifies the fault rather than catching it
Rango de presión
The transmitter is not a pressure boundary - it is mounted remotely and connected to the sensor by coil and electrode cables, so the process pressure rating belongs to the sensor
Variantes
Wall-mount enclosure with hinged door, mounting ears for wall or pipe mounting, optional backlit display and a 15-button tactile keypad. Diagnostic suites DS1 (High Process Noise, Coated Electrode Detection) and MV (commanded and continuous Smart Meter Verification, 4-20 mA loop verification)
Aplicación
Magnetic flow installations where the electronics should not live on the pipe, and specifically slurry lines where High Process Noise, Coated Electrode Detection and Smart Meter Verification earn their place