Rosemount 225 Toroidal Conductivity Sensor
Purchase options
From$612.02
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Toroidal (electrodeless) sanitary conductivity sensor with a 2 inch tri-clamp connection, 200 uS/cm to 2 S/cm range and a nominal 2.7/cm cell constant.
Key features
- Toroidal (electrodeless) conductivity sensor for sanitary service
- 200 uS/cm to 2 S/cm, down to 15 uS/cm with Rosemount 1056 or 56 transmitters
- Nominal cell constant 2.7/cm
- 2 inch tri-clamp process connection, 200 psig rating
- Body options in glass-filled PEEK, unfilled PEEK (3-A / FDA or USP Class VI) and Tefzel
- 20 ft integral cable, extendable to 200 ft with a remote junction box
Key specifications
- Measurement principle
- Toroidal (electrodeless) - two coils sealed in a polymer body, with the process liquid completing the magnetic circuit through the bore
- Process temperature
- 230 °F (110 °C) for options 03 and 09; 266 °F (130 °C) for options 07 and 08
- Pressure rating
- 200 psig (1480 kPa abs)
- Process connection
- 2 inch tri-clamp
- Wetted materials
- Option 03 glass-filled PEEK, option 07 unfilled PEEK (21 CFR 177.2415 and 3-A 74-06), option 08 unfilled PEEK (USP Class VI), option 09 unfilled Tefzel
- Certifications
- 3-A Sanitary Standard 74-06 and FDA food contact 21 CFR 177.2415 with option -07; USP Class VI with option -08
A conductivity sensor with no electrodes in the fluid
The Rosemount 225 measures conductivity with two coils sealed inside a polymer toroid. The liquid passes through the hole in the middle and completes the magnetic circuit between them, so there is nothing metal in contact with the product and nothing to polarise, foul or erode. It is the design that made in-line conductivity practical in sanitary service.
Why toroidal, and why sanitary
A contacting cell is more sensitive at low conductivity, but it puts two electrodes in the process: they polarise in strong solutions, they foul, and in a hygienic line they are a crevice to clean around. The 225 gives that up in exchange for a range that starts at 200 µS/cm and runs to 2 S/cm — concentrated CIP caustic and acid, brine, product interfaces — with a body that meets 3-A 74-06 (option -07) or USP Class VI (option -08) depending on which moulding you specify.
The cell constant is fixed, the range is not
A nominal cell constant of 2.7/cm is what the transmitter is calibrated against, and it does not change with the process. What does change is the low end: the sensor reads down to 15 µS/cm when paired with a Rosemount 1056 or 56 transmitter, and 200 µS/cm with other electronics. If a line is expected to run near the bottom of its range, the transmitter choice is part of the sensor decision, not a separate purchase.
Body options decide the temperature
Four mouldings cover the range: glass-filled PEEK (option 03) and unfilled Tefzel (option 09) are rated to 230 °F (110 °C), while the unfilled PEEK of options 07 and 08 is rated to 266 °F (130 °C). Options 07 and 08 differ in the compliance they carry rather than the material — 21 CFR 177.2415 and 3-A for 07, USP Class VI for 08. All four share the 200 psig pressure rating and the 2 in tri-clamp connection.
Installing it in a hygienic line
The 2 in tri-clamp fits the standard sanitary ferrule, so the sensor goes in where a valve or a spool would. Inside the line it is insensitive to flow rate and direction, which is what makes it usable in a CIP return header where the flow reverses. Keep 20 ft of integral cable as supplied, or extend to 200 ft (61 m) through a 23550-00 remote junction box — the box is the point where the sensor can be changed without pulling the long cable run.
Where it fits
Pharmaceutical and food and beverage lines: CIP concentration control, product/water interface detection, product quality checks and chromatographic separation monitoring.
Manufacturer reference: Rosemount 225 Toroidal Conductivity Sensors Product Data Sheet, LIQ-PDS-225, and Quick Start Guide 00825-0100-3225.
Specifications
| Attribute | Value |
|---|---|
| Measurement principle | Toroidal (electrodeless) - two coils sealed in a polymer body, with the process liquid completing the magnetic circuit through the bore |
| Process temperature | 230 °F (110 °C) for options 03 and 09; 266 °F (130 °C) for options 07 and 08 |
| Pressure rating | 200 psig (1480 kPa abs) |
| Process connection | 2 inch tri-clamp |
| Wetted materials | Option 03 glass-filled PEEK, option 07 unfilled PEEK (21 CFR 177.2415 and 3-A 74-06), option 08 unfilled PEEK (USP Class VI), option 09 unfilled Tefzel |
| Certifications | 3-A Sanitary Standard 74-06 and FDA food contact 21 CFR 177.2415 with option -07; USP Class VI with option -08 |
| Measurement range | 200 uS/cm to 2 S/cm; down to 15 uS/cm when paired with a Rosemount 1056 or 56 transmitter |
| Cable connection | 20 ft (6.1 m) integral cable; up to 200 ft (61 m) total with a remote junction box |
| Cell constant | 2.7/cm nominal |
| Dimensions | 2 lb (1.0 kg) net, 3 lb (1.5 kg) shipping weight |
Documents
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer