Rosemount 23550-00 Remote Junction Box
No product images are published for this product yet.
Purchase options
$132.43
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Weather-proof remote junction box with an internal extension board and no preamplifier, for extending Rosemount liquid analysis sensor cables to a distant transmitter.
Key features
- Weather-proof remote junction box with an internal extension board
- No preamplifier - for sensors whose signal does not need amplification at the sensor
- ABS plastic enclosure, 4 in D x 5 in H x 8.25 in W
- Wall or pipe mounting with bracket PN 2002565 (ordered separately)
- Works with the Rosemount 56, 1056, 1057, 1066 and 5081 transmitters
- Replacing a sensor never disturbs the cable run to the analyzer
Key specifications
- Process connection
- Cable entries 3/4 in NPT with a 3/4 in female connection to the analyzer
- Enclosure rating
- ABS plastic enclosure, weather-proof; not submersible
- Sensor compatibility
- Most Rosemount liquid analysis sensors other than pH/ORP sensors without a preamplifier
- Application
- Extending a Rosemount liquid analysis sensor cable to a transmitter that is installed far from the measurement point
- Mounting
- Wall or pipe of up to 2 in (52 mm) using mounting bracket PN 2002565, ordered separately
- Dimensions
- 4 in D x 5 in H x 8.25 in W (102 x 127 x 210 mm)
What the 23550-00 junction box does
A weather-proof enclosure that sits between a liquid analysis sensor and its transmitter when the two cannot be close together. The sensor cable lands on an extension board inside the box, and a separate cable carries the signal on to the analyzer.
The point of the extra connection
It is a serviceability decision. The sensor is the part that fouls, breaks and gets replaced; the cable run to the analyzer is the part that is hard to install, often in conduit or a tray. With the junction box, replacing a sensor means opening a box and disconnecting one cable — the long run is never touched, never re-terminated, and never re-validated.
Why there is no preamplifier in this one
The 23550-00 carries an extension board and nothing else. That is exactly right for a sensor whose signal can travel — a contacting conductivity or toroidal sensor, a dissolved oxygen sensor, a chlorine sensor. The pH and ORP analog sensors that need their high-impedance signal amplified at the sensor end use the 23555-00 instead, which has the preamplifier inside. Ordering the wrong one is the most common mistake in this family: an unamplified pH signal over a long cable is noise, and no amount of transmitter configuration recovers it.
Installing it
The enclosure is ABS, rated weather-proof, and mounts to a wall or to a pipe of up to 2 in with bracket 2002565, ordered separately. Cable entries are ¾ in NPT, with a ¾ in female connection to the analyzer. Keep the lid gasket clean and the entries pointing down — the box is weather-proof, not submersible.
Where it fits
Rosemount liquid analysis installations where the sensor is far from the transmitter: water and wastewater plants, cooling towers, chemical process lines and any site where the analyzer lives in a cabinet away from the measurement point.
Manufacturer reference: Emerson Liquid Analysis Accessory Guide, table 19 (junction boxes 23550-00, 23555-00, 23709-00).
Specifications
| Attribute | Value |
|---|---|
| Process connection | Cable entries 3/4 in NPT with a 3/4 in female connection to the analyzer |
| Enclosure rating | ABS plastic enclosure, weather-proof; not submersible |
| Sensor compatibility | Most Rosemount liquid analysis sensors other than pH/ORP sensors without a preamplifier |
| Application | Extending a Rosemount liquid analysis sensor cable to a transmitter that is installed far from the measurement point |
| Mounting | Wall or pipe of up to 2 in (52 mm) using mounting bracket PN 2002565, ordered separately |
| Dimensions | 4 in D x 5 in H x 8.25 in W (102 x 127 x 210 mm) |
| Preamplifier | None - the box carries an internal extension board only; pH/ORP sensors that need amplification at the sensor end use the 23555-00 |
| Transmitter compatibility | Rosemount 56, 1056, 1057, 1066 and 5081 |
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.