Rosemount 23555-00 Junction Box with Preamplifier
No product images are published for this product yet.
Purchase options
$390.24
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Weather-proof remote junction box with an internal pH/ORP preamplifier (PN 23557-00, FM/CSA/ATEX intrinsically safe) for long sensor cable runs to a distant transmitter.
Key features
- Remote junction box with an internal pH/ORP preamplifier
- Preamplifier PN 23557-00 with FM, CSA and ATEX intrinsically safe approvals
- For pH and ORP analog sensors only - conductivity and DO sensors use the 23550-00
- ABS weather-proof enclosure, 4 in D x 5 in H x 8.25 in W
- Wall or pipe mounting with bracket PN 2002565 (ordered separately)
- Works with the Rosemount 56, 1056, 1057, 1066 and 5081 transmitters
Key specifications
- Process connection
- Cable entries 3/4 in NPT with a 3/4 in female connection to the analyzer
- Enclosure rating
- ABS plastic, weather-proof; not submersible
- Sensor compatibility
- Any Rosemount pH/ORP sensor without a preamplifier (see table 20 of the accessory guide); conductivity and dissolved oxygen sensors use the 23550-00 instead
- Application
- Extending a pH or ORP analog sensor cable to a transmitter that is installed far from the measurement point
- Mounting
- Wall or pipe of up to 2 in (52 mm) using mounting bracket PN 2002565, ordered separately
- Dimensions
- 4 in D x 5 in H x 8.25 in W (102 x 127 x 210 mm)
What the 23555-00 junction box does
A weather-proof enclosure for a pH or ORP sensor that has to be installed far from its analyzer. Unlike the plain extension box, this one contains an amplifier: the preamplifier sits at the sensor end, where the signal is still strong, and the box then carries an ordinary low-impedance signal down the long cable to the transmitter.
Why pH needs a preamplifier and conductivity does not
A glass pH electrode presents an impedance measured in hundreds of megohms. That signal survives a couple of metres of cable and very little more — cable capacitance and electrical noise swamp it, and what arrives at the analyzer is a drifting, unstable number that looks like a failing electrode. Amplifying at the sensor end changes the problem entirely: after the preamplifier the signal is low-impedance and can travel the length of a plant. A contacting conductivity or dissolved oxygen sensor has no such problem, which is why the 23550-00 exists without an amplifier and this box is specifically for pH and ORP.
What is inside
Preamplifier PN 23557-00, mounted on the internal board, carrying FM, CSA and ATEX intrinsically safe approvals for hazardous area installation. The enclosure is the same ABS, weather-proof housing as the extension box — 4 in deep, 5 in high and 8.25 in wide — with ¾ in NPT cable entries and a ¾ in female connection to the analyzer.
Installing it
Mounts to a wall or to a pipe of up to 2 in using bracket 2002565, which is ordered separately. Keep the lid gasket clean, the entries pointing down, and the preamplifier board dry — a wet preamplifier reads as a failed sensor, and the box is weather-proof rather than submersible. Extension cable from the box to the transmitter is also a separate item.
Where it fits
pH and ORP measurement points in water treatment, chemical plants, cooling systems and pulp mills where the sensor is on the line and the analyzer is in a cabinet or control room.
Manufacturer reference: Emerson Liquid Analysis Accessory Guide, table 19 (junction boxes 23550-00, 23555-00, 23709-00) and table 20.
Specifications
| Attribute | Value |
|---|---|
| Process connection | Cable entries 3/4 in NPT with a 3/4 in female connection to the analyzer |
| Enclosure rating | ABS plastic, weather-proof; not submersible |
| Sensor compatibility | Any Rosemount pH/ORP sensor without a preamplifier (see table 20 of the accessory guide); conductivity and dissolved oxygen sensors use the 23550-00 instead |
| Application | Extending a pH or ORP analog sensor cable to a transmitter that is installed far from the measurement point |
| Mounting | Wall or pipe of up to 2 in (52 mm) using mounting bracket PN 2002565, ordered separately |
| Dimensions | 4 in D x 5 in H x 8.25 in W (102 x 127 x 210 mm) |
| Preamplifier | Includes preamplifier PN 23557-00 mounted on the internal board, with FM, CSA and ATEX intrinsically safe approvals |
| Transmitter compatibility | Rosemount 56, 1056, 1057, 1066 and 5081 |
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.