Rosemount 24091-02 Flow Cell
Purchase options
$219.49
- Availability
- Call for availability Originalx
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Polycarbonate low-flow sample cell with a 3/4 in MNPT sensor connection for Rosemount 400 conductivity sensors, 1/4 in FNPT sample ports, 70 C and 90 psig.
Key features
- Polycarbonate low-flow sample cell for the Rosemount 400 conductivity sensor
- 3/4 in MNPT sensor connection (the -02 version; -00 and -01 take 1 in MNPT sensors)
- 1/4 in FNPT inlet and outlet, with 1/4 in MNPT to 1/4 in OD tubing fittings supplied
- Polycarbonate/polyester body and nut, 316 stainless fittings, silicone O-ring
- 158 °F (70 °C) and 90 psig maximum
- Clear body shows the sensor and any trapped bubbles
Key specifications
- Process temperature
- Maximum 158 °F (70 °C)
- Pressure rating
- Maximum 90 psig (621 kPa)
- Process connection
- 1/4 in FNPT inlet and outlet, supplied with optional 1/4 in MNPT to 1/4 in OD tubing fittings
- Wetted materials
- Polycarbonate/polyester blend body and nut, 316 stainless steel 1/4 in fittings, silicone O-ring
- Sensor connection
- 3/4 in MNPT threads on the sensor (the -00 and -01 versions of this cell take 1 in MNPT sensors instead)
- Application
- Low-flow side-stream sample cell for a Rosemount 400 conductivity sensor
A sample cell for a 400 sensor
A clear polycarbonate side-stream cell that holds a Rosemount 400 conductivity sensor on a ¾ in MNPT connection and passes a small sample through the chamber around it. It is what turns a screw-in sensor into a sample point on a skid, a bypass line or a drain.
Why a low-flow cell at all
Because the sample is often the expensive part. On a purification skid or a high-purity loop, diverting a large flow to a sensor costs product and treatment capacity; a low-flow cell keeps the sample small and lets it go to waste without upsetting the balance of the line. It also puts the sensor where it can be seen and serviced — the clear body shows the sensor and any bubbles in it, which matters on a line where a trapped bubble reads as a conductivity drop.
Where it sits in the family
The 24091 cells look alike and take different sensors: the -00 and -01 accept 1 in MNPT sensors and cover pH/ORP, dissolved oxygen, ozone and chlorine, while this -02 accepts the ¾ in MNPT connection used by the Rosemount 400 conductivity sensor. Ordering by appearance is the mistake here; the sensor thread is the discriminator.
Plumbing and limits
¼ in FNPT inlet and outlet, supplied with optional ¼ in MNPT to ¼ in OD tubing fittings for flexible tube. Body and nut are a polycarbonate/polyester blend with a silicone O-ring and 316 stainless fittings — fine for clean water service at up to 158 °F (70 °C) and 90 psig, but not for solvents that attack polycarbonate, and not for a sample carrying solids that would settle in the chamber.
Using it well
Feed it from a point that represents the process, not from a dead leg, and keep the flow steady: a low-flow cell's reading depends on the sample being exchanged at a consistent rate. If the application is dissolved oxygen, chlorine or ozone — which the -00 and -01 versions of this cell serve — flow control is not a refinement, it is part of the measurement.
Where it fits
Side-stream conductivity measurement with a Rosemount 400 sensor: high purity water, reverse osmosis and pharmaceutical water systems, and any point where the sample volume has to stay small.
Manufacturer reference: Emerson Liquid Analysis Accessory Guide, table 12 and figure 14 (polycarbonate low flow cell 24091-00, 24091-01, 24091-02).
Specifications
| Attribute | Value |
|---|---|
| Process temperature | Maximum 158 °F (70 °C) |
| Pressure rating | Maximum 90 psig (621 kPa) |
| Process connection | 1/4 in FNPT inlet and outlet, supplied with optional 1/4 in MNPT to 1/4 in OD tubing fittings |
| Wetted materials | Polycarbonate/polyester blend body and nut, 316 stainless steel 1/4 in fittings, silicone O-ring |
| Sensor connection | 3/4 in MNPT threads on the sensor (the -00 and -01 versions of this cell take 1 in MNPT sensors instead) |
| Application | Low-flow side-stream sample cell for a Rosemount 400 conductivity sensor |
Accessories and related items
Accessories and related items for this product are not available to order on Sproh yet. If the manufacturer lists them, an engineer can source them for you.
Ask an engineer