Rosemount 490A Optical Dissolved Oxygen Sensor
No product images are published for this product yet.
Purchase options
From$1,586.46
- Availability
- Call for availability Originalx
Configurable — choose options to build the exact part number.
Independent supplier
Sproh is an independent supplier and is not an authorized distributor or official representative of Rosemount or Emerson.
Overview
Optical dissolved oxygen sensor using EPA-approved luminescence quenching: 0-60 mg/L, 0.01 mg/L resolution, two-year cap, no flow dependence.
Key features
- Optical dissolved oxygen sensor using the EPA-approved luminescence-quenching method, 0 to 60 mg/L or 0 to 600 percent saturation, with 0.01 mg/L resolution
- It consumes no oxygen and has no electrolyte, so the sensing cap is replaced every two years rather than being recalibrated and rebuilt
- The measurement does not need flow past the sensor: an amperometric cell consumes oxygen at its own surface and reads low in still water, which the optical method does not do
- Measures accurately in both liquid and gas without separate calibrations, and reads accurately within 90 seconds of going from dry to wet, so a cleaned sensor is valid as soon as it is back in
- Accuracy is 0.1 mg/L from 0 to 20 mg/L and 2 percent from 20 to 60 mg/L, with response under 45 seconds to 90 percent of final value
- Temperature accuracy of 0.1 C is part of the oxygen measurement, because mg/L is derived from partial pressure using the solubility of oxygen, which depends on temperature
Key specifications
- Accuracy
- 0.1 mg/L from 0 to 20 mg/L, and 2 percent from 20 to 60 mg/L - so a sensor quoted as 0.1 mg/L is that good in the lower band, which is where aeration control actually operates. Resolution 0.01 mg/L. Temperature accuracy 0.1 C, which is part of the oxygen figure because mg/L is derived from partial pressure using the solubility of oxygen, and solubility depends on temperature
- Output signal
- Digital Modbus over RS-485, with an M12 connector and either 0.5 ft (0.15 m) or 32.8 ft (10 m) of integral cable
- Diagnostics
- The failure mode to watch is a temperature element, because a failed one does not stop the oxygen reading, it makes it plausible and wrong. Otherwise the maintenance drivers are physical: biofilm on the sensing face reads as falling oxygen, which is why an anti-fouling guard, an air-blast guard and a weighted guard are offered - choose from the water, not from the catalogue
- Pressure range
- 150 psi (10.3 bar) from +32 to +122 F (0 to +50 C). In an aeration basin at atmospheric pressure that is margin; in a pressurised line it is the difference between a reading and a leak
- Application
- Dissolved oxygen in wastewater aeration basins and ponds, surface water and aquaculture, where the sensing face can be cleaned on a two-year cycle rather than a weekly calibration routine. Clean, warm, low-fouling streams where someone is on site daily are still a fair application for an amperometric cell
- Operating temperature (°C)
- +23 to +122 F (-5 to +50 C)
It measures light, and everything else follows from that
The Rosemount 490A is an optical dissolved oxygen sensor using the EPA-approved luminescence-quenching method, with a range of 0 to 60 mg/L or 0 to 600 % saturation, ±0.1 mg/L accuracy below 20 mg/L, 0.01 mg/L resolution, Modbus/RS-485 communication over an M12 connector, and a sensing cap that is replaced every two years.
It is worth understanding what the sensor actually does, because the maintenance behaviour follows from it rather than from good design. A conventional amperometric DO sensor is an electrochemical cell: oxygen diffuses through a membrane and is reduced at an electrode, producing a small current. That current is the measurement, and every part of the cell is consumed or contaminated by doing its job.
The 490A does not consume oxygen. A luminophore in the sensing cap is excited by a light pulse, and oxygen quenches that luminescence — the more oxygen, the shorter the glow. The sensor times the decay. Nothing in the measurement is depleted, and nothing needs the sample to flow past it.
Three consequences, and each one removes a familiar chore
No electrolyte to run out. An amperometric cell has an electrolyte reservoir that depletes and a membrane that fouls; both change the reading slowly and in the same direction, so the operator compensates with frequent calibration and eventually a rebuild. The optical cap has no electrolyte, which is why its specified life is a two-year replacement interval rather than a maintenance schedule.
No flow dependence. An amperometric sensor consumes oxygen at its own surface, so it needs flow to keep the measurement honest — which is why those sensors carry a minimum-flow requirement, and why a probe in a quiet corner of a basin reads low while the probe beside the mixer reads correctly. The optical sensor consumes nothing, so it can sit in still water. The weighted guard in the accessory list exists to keep the sensor vertical in high-flow or mixed conditions, which is about mechanical position, not about feeding the sensor.
One calibration for liquid and gas. The datasheet states the sensor measures accurately in both liquid and gas without requiring separate calibrations, and that it reads accurately within 90 seconds of going from dry to wet. For a plant that pulls a sensor out for cleaning, that is the practical difference between a reading that is valid when it goes back in and one that needs an hour to polarise. The air calibration cup is supplied with the sensor, because air is the convenient reference.
Why a temperature specification sits inside a dissolved oxygen specification
The sensor measures the partial pressure of oxygen. The number an operator needs — mg/L — is that partial pressure converted using the solubility of oxygen, and solubility depends strongly on temperature and on salinity. The temperature specification is therefore part of the measurement, not a housekeeping detail: ±0.1 °C over −5 to +50 °C is what makes the mg/L figure trustworthy, and it is why a sensor with a failed temperature element reads plausibly and wrongly rather than failing outright.
Pressure matters for the same reason, and is rated to 150 psi (10.3 bar) from 0 to 50 °C. In an aeration basin at atmospheric pressure that is simply margin; in a pressurised line it is the difference between a reading and a leak.
Response time is a control-loop parameter
Under 45 seconds to 90 % of final value, and under 60 seconds to 95 %. That is fast enough to matter for aeration control: the blowers are the largest energy consumer on a wastewater plant, and dissolved oxygen is the variable being held. A slow sensor forces the control loop to be detuned, which shows up as an oxygen setpoint that wanders instead of holding — and the energy bill follows the wandering, not the setpoint.
The accuracy bands are worth quoting precisely because they are two: ±0.1 mg/L between 0 and 20 mg/L, and ±2 % between 20 and 60 mg/L. A sensor specified as "±0.1 mg/L" is only that good in the lower band, which is where aeration control actually operates.
The guards are the site's problem, not the sensor's
Three guards are offered, and each answers a specific nuisance rather than adding a feature. The anti-fouling guard reduces biological growth on the sensing face — the dominant maintenance driver in a nutrient-rich basin, where a biofilm forms in days and reads as falling oxygen. The air-blast guard uses an air supply to keep the face clear where fouling is severe. The weighted guard holds the sensor vertical in high flow.
Choose from the water, not from the catalogue: a clean final effluent needs none of them, a mixed liquor basin needs the anti-fouling guard, and a plant that already has instrument air at the basin can use the air blast.
Where it fits
Wastewater aeration basins and ponds — the application the datasheet names first — plus surface water, aquaculture and any dissolved oxygen measurement where the maintenance department is the constraint. If the measurement is in a clean, warm, low-fouling stream and someone is on site daily, an amperometric sensor is still a reasonable instrument. If nobody is, the two-year cap is the argument.
Manufacturer reference: Rosemount 490A Optical Dissolved Oxygen Sensor Product Data Sheet for ranges, accuracy, response time, wetted materials, ordering codes and guards.
Specifications
| Attribute | Value |
|---|---|
| Accuracy | 0.1 mg/L from 0 to 20 mg/L, and 2 percent from 20 to 60 mg/L - so a sensor quoted as 0.1 mg/L is that good in the lower band, which is where aeration control actually operates. Resolution 0.01 mg/L. Temperature accuracy 0.1 C, which is part of the oxygen figure because mg/L is derived from partial pressure using the solubility of oxygen, and solubility depends on temperature |
| Output signal | Digital Modbus over RS-485, with an M12 connector and either 0.5 ft (0.15 m) or 32.8 ft (10 m) of integral cable |
| Diagnostics | The failure mode to watch is a temperature element, because a failed one does not stop the oxygen reading, it makes it plausible and wrong. Otherwise the maintenance drivers are physical: biofilm on the sensing face reads as falling oxygen, which is why an anti-fouling guard, an air-blast guard and a weighted guard are offered - choose from the water, not from the catalogue |
| Pressure range | 150 psi (10.3 bar) from +32 to +122 F (0 to +50 C). In an aeration basin at atmospheric pressure that is margin; in a pressurised line it is the difference between a reading and a leak |
| Application | Dissolved oxygen in wastewater aeration basins and ponds, surface water and aquaculture, where the sensing face can be cleaned on a two-year cycle rather than a weekly calibration routine. Clean, warm, low-fouling streams where someone is on site daily are still a fair application for an amperometric cell |
| Operating temperature (°C) | +23 to +122 F (-5 to +50 C) |
| Response time | Under 45 seconds to 90 percent of final value, and under 60 seconds to 95 percent. Fast enough to matter for aeration control, where the blowers are the largest energy consumer on the plant and dissolved oxygen is the variable being held; a slow sensor forces the loop to be detuned and shows up as a wandering setpoint |
| Sensor | Optical, luminescence quenching, EPA-approved method. A luminophore in the sensing cap is excited by a light pulse and oxygen quenches the luminescence, so the sensor times a decay rather than measuring a current. Nothing in the measurement is depleted and the sample does not have to flow past the sensor. Sensing cap replaced every two years |