Ir al contenido principal
Vendido y enviado por Sproh

Sensor optico de oxigeno disuelto Rosemount 490A

Número de pieza490AFabricanteEmerson

Todavía no hay imágenes publicadas para este producto.

Opciones de compra

Desde1586,46 US$

Disponibilidad
Consultar disponibilidad Original
Seleccionar opción

Configurable: elija las opciones para construir el número de pieza exacto.

Proveedor independiente

Sproh es un proveedor independiente y no es un distribuidor autorizado ni representante oficial de Rosemount o Emerson.

Resumen

Sensor optico de oxigeno disuelto por luminiscencia aprobada por EPA: 0-60 mg/L, resolucion 0.01 mg/L, capuchon de dos anos y sin dependencia de flujo.

Especificaciones principales

Exactitud
0.1 mg/L from 0 to 20 mg/L, and 2 percent from 20 to 60 mg/L - so a sensor quoted as 0.1 mg/L is that good in the lower band, which is where aeration control actually operates. Resolution 0.01 mg/L. Temperature accuracy 0.1 C, which is part of the oxygen figure because mg/L is derived from partial pressure using the solubility of oxygen, and solubility depends on temperature
Señal de salida
Digital Modbus over RS-485, with an M12 connector and either 0.5 ft (0.15 m) or 32.8 ft (10 m) of integral cable
Diagnósticos
The failure mode to watch is a temperature element, because a failed one does not stop the oxygen reading, it makes it plausible and wrong. Otherwise the maintenance drivers are physical: biofilm on the sensing face reads as falling oxygen, which is why an anti-fouling guard, an air-blast guard and a weighted guard are offered - choose from the water, not from the catalogue
Rango de presión
150 psi (10.3 bar) from +32 to +122 F (0 to +50 C). In an aeration basin at atmospheric pressure that is margin; in a pressurised line it is the difference between a reading and a leak
Aplicación
Dissolved oxygen in wastewater aeration basins and ponds, surface water and aquaculture, where the sensing face can be cleaned on a two-year cycle rather than a weekly calibration routine. Clean, warm, low-fouling streams where someone is on site daily are still a fair application for an amperometric cell
Temperatura de operación (°C)
+23 to +122 F (-5 to +50 C)
Ver todas las especificaciones