Ir al contenido principal
Vendido y enviado por Sproh

Cable plano Rosemount 24224-00

Número de pieza24224-00FabricanteEmerson

Todavía no hay imágenes publicadas para este producto.

Opciones de compra

48,55 US$

Disponibilidad
Consultar disponibilidad Original

Proveedor independiente

Sproh es un proveedor independiente y no es un distribuidor autorizado ni representante oficial de Rosemount o Emerson.

Resumen

Subconjunto de cable plano de 14 conductores que une la placa de CPU con la placa de fuente de alimentacion en un transmisor de analisis de liquidos Rosemount.

Especificaciones principales

Exactitud
No electrical specification of its own in the published data. What it does is carry power and signals a few centimetres between two boards inside the enclosure, so a degraded cable shows up as an intermittent fault rather than as a measurement error
Señal de salida
Not applicable. 14 conductors with an insulation-displacement connector at each end; the pin-1 marking must be observed at both ends, because a 14-way connector inverted into its header will seat and will not work
Diagnósticos
Three failures account for a ribbon cable replacement and none is visible on inspection: a connector latch that cracks after a few removals and leaves the cable held by nothing, an insulation-displacement contact that goes resistive after the cable is flexed at the wrong angle, and conductors pulled at the edge when the cable is tugged instead of the connector. One conductor of fourteen failing makes a single function misbehave rather than killing the instrument - the intermittent fault that responds to a tap on the enclosure. Reseller listings for this family note that retired products cannot be ordered, so a critical analyser deserves one of these on the shelf
Rango de presión
Not applicable - an internal wiring sub-assembly with no process connection and no wetted parts
Aplicación
Internal spare for a Rosemount liquid analysis transmitter, bought when the CPU-to-power-supply interconnect has been disturbed during board service or when an intermittent fault has been traced to it. Not part of a new analyser specification
Montaje
Internal to the transmitter, connecting the CPU board to the power supply board. Fold it along the crease the original had: ribbon cable is flexible in one axis and stiff in the other, and routing a new cable against that crease puts a permanent shear load on the outermost conductors
Ver todas las especificaciones