Resumen

Indicador de nivel magnetico con columna de banderas seca y flotador elegido por SG y presion: sin fluido de proceso en el vidrio, de -162 a 454 C y hasta 3400 psig.

Especificaciones principales

Exactitud
Resolution is set by the flag pitch, and the reading is the float position rather than a computed value. The selection constraint is the float: it must float in the process liquid and survive the process pressure, and those two requirements pull opposite ways
Señal de salida
Local flag column, red for liquid and silver for vapour. The same float drives externally mounted magnetostrictive transmitters and magnetic switches through the chamber wall, so a vessel can have local indication, a 4-20 mA signal and alarm contacts with no second process penetration
Diagnósticos
The float table is a trade curve and reading it correctly is the whole specification: as minimum SG rises from 0.58 to 0.86 the float's pressure limit rises from 94 to 948 psig, because a float that floats in a light liquid needs thin walls and large displacement while one that holds pressure needs thick walls and therefore weight. A float chosen for 0.86 SG in a liquid that actually sits at 0.7 does not float, and the failure is silent - the flags simply stop where they stopped
Rango de presión
Float pressure rating up to 3400 psig (234.4 bar) at 100 F, with the float's own limit set by its selection: 94 psig for the 0.58 SG float up to 948 psig for the 0.86 SG float. The chamber is designed to ASME B31.3 and can be hydrostatically tested to 1.5 times the MAWP of its flange or pressure rating before shipment, with certificates available
Aplicación
Turbulent tanks, heat exchangers, boilers, separators, acid storage and any vessel where a sight glass is being replaced because it is a leak point, a fouling surface or a hazard. A local level that reads without power is the quiet advantage of a mechanical indication
Montaje
Chamber with process connections, vent and drain, mounted beside the vessel. A dual chamber version is built for use with the Rosemount 5300, 3308 and 3300 guided wave radars, so the radar measures inside a calm stilling column instead of the open vessel - which is how a radar gets a stable surface on a turbulent, boiling or agitated tank