Calibration is evidence, not a ritual: another engineer should be able to repeat it and reach the same conclusion from the record.

A field calibration compares an instrument with a reference that is itself traceable. The comparison is only as good as that reference: a cylinder past its validity, or a regulator that adds humidity, produces a record that looks sound and is not.

Zero and span answer different questions: zero shows what the instrument reads with no gas present, span how it responds to a known concentration.

How a calibration is carried out

  1. Establish the reference

    Check the reference gas, its stated concentration and its validity, and record what was used.

  2. Zero

    Present clean air or zero gas and record the offset before any adjustment is made.

  3. Span

    Present the reference concentration, let the reading settle and record the response before and after.

  4. Record the conditions

    Note temperature, pressure and humidity, because a response is only comparable within them.

  5. Issue the record

    State the reference, the readings, the adjustment and the next due point, and flag any failure.