Controlled release testing
A controlled release is the closest thing the field has to a known answer: gas leaves a calibrated source at a measured rate, and the system must recover it.
- Section
- Research
- Language
- English
- Text
- Conceptual / Reference
A controlled release test replaces the leak with a source whose rate is known. A mass flow controller passes pure gas, usually methane, through a release point at a set rate, sited relative to the monitors and the prevailing wind. The system must detect it, locate it and estimate its rate.
How a release test runs
Choose the geometry
Site the release upwind so the plume stays in range at the expected wind speeds.
Set and record the rate
Log the setpoint, calibration, gas, release height and orifice.
Hold the release
Long enough for the plume to establish and the averaging period to fill; a short release tests response time, not quantification.
Recover the rate blind
Run the operational inversion without the setpoint, then compare estimate and released rate.
Repeat across conditions
Vary rate, distance, wind and stability. One release proves little beyond that day.
What it establishes
It validates the chain, not the sensor: a good sensor with a poor dispersion model fails, and so does a good model fed a biased reading. It is also a point-source idealisation — real sites have area sources, simultaneous leaks and buildings the model does not know.
Passing means the method works under the conditions tested. Presenting that as an accuracy figure for every site is a step the evidence does not support.