Gas blends and mixed streams

A blend is not two gases measured separately. It is a stream whose physical properties — density, buoyancy, speed of sound, infrared absorption — depend on the ratio.

As networks begin to carry methane-hydrogen blends, the composition stops being a constant. A sensor calibrated for pure methane reads a blend as a lower concentration even when the leak rate is unchanged, because the fraction of the mixture that absorbs at the chosen wavelength has fallen.

The same shift changes buoyancy and therefore where a leak accumulates, and it changes the lower flammability limit that an alarm threshold was derived from.

What composition changes
PropertyEffect of adding hydrogenConsequence for monitoring
DensityDecreasesLeaking gas rises faster; high-level accumulation
Infrared absorptionDilutes the methane signalA methane-only calibration under-reads
Flammability rangeWidens considerablyAlarm philosophy must be reconsidered
Speed of soundIncreasesAffects ultrasonic and flow-based inference

Where composition varies, a monitoring system needs either a composition input or a measurement that is insensitive to it. Assuming a fixed blend ratio is a defensible simplification only while the operator can state that the ratio really is fixed.