Comparing continuous sensor technologies
A sensor chosen on selectivity alone disappoints in the field. Response time, cross-sensitivity and maintenance decide whether a continuous measurement survives its first year.
- Section
- Research
- Language
- English
- Text
- Conceptual / Reference
Choosing a principle for a handheld instrument is a different exercise from choosing one for a fixed continuous monitor. The fixed monitor samples unattended for years, in weather, and must still be believable when it reports a rise.
| Principle | Response time | Cross-sensitivity | Maintenance |
|---|---|---|---|
| Optical (NDIR, TDLAS) | Seconds; cell or path limited | Low on a resolved line; dirty optics bias it | Window and mirror cleaning, alignment |
| Electrochemical | Tens of seconds; slow recovery | Other redox gases; humidity shifts the baseline | Cell replacement on a known life |
| Catalytic (pellistor) | A few seconds to tens | Any combustible gas reads as the calibration gas | Bump tests; the beads are consumables |
| Photoionisation (PID) | Seconds; lamp and inlet limited | Sums every ionisable compound | Lamp cleaning or replacement |
Why response time is a system property
A diffusion housing, a filter, a sample line and a pump each add delay, and a long line smooths the peak: a short excursion arrives as a low, wide bump. Where the reading feeds an alarm, that distortion matters more than the datasheet.
Cross-sensitivity is the same problem in chemical form: a reading sums everything the element responds to, and the instrument must separate the gas of interest — by wavelength, by filtration, or with a second measurement. Maintenance is the cost specifications hide: optics want cleanliness but no consumable, cells deplete from manufacture, pellistors die from poisoning.