Skip to content
MetIQGas intelligence

Research

Measurement uncertainty

A reading is an estimate with a range. Publishing the estimate and hiding the range is the most common way a monitoring system overstates what it knows.

Section
Research
Language
English
Text
Conceptual / Reference

Uncertainty is not error. Error is the difference from a true value nobody knows; uncertainty is the range where that difference is expected to lie. One can be estimated and reported; the other cannot be observed.

Two quantities are confused. The detection limit is where the instrument can decide gas is present; it says nothing about how well a concentration above it is measured. The limit of quantification is where a numerical result becomes defensible, and always higher.

Where the uncertainty enters
SourceTypeHandling
Calibration gas certificateBPropagated from a traceable standard
Zero and span driftBControl charts and a drift budget
RepeatabilityAStandard deviation of replicates
Cross-sensitivityB, often the largest termCorrection or stated bound
Sampling and response timeBCharacterised by a step change
Meteorology and modelBEvaluated against release tests
A is statistical; B comes from certificates or specifications. Both combine in quadrature, usually with a factor of two.

Traceability

Traceability is an unbroken chain of calibrations, each with a stated uncertainty, to a national metrology institute. A cylinder with a traceable certificate carries part of that chain; an instrument calibrated against a colleague's does not.

Why one figure misleads

An accuracy of "±x%" describes one concentration, one temperature, one mixture. Uncertainty grows near the detection limit, and again when concentration becomes a mass rate, because meteorology and model structure enter. The honest output is a value with an interval, and the conditions it applies to.

Continue reading